Two-dimensional mapping of underdosed areas using radiochromic film for patients undergoing total skin electron beam radiotherapy

被引:17
作者
Gamble, LM
Farrell, TJ
Jones, GW
Hayward, JE
机构
[1] Juravinski Canc Ctr, Dept Med Phys, Hamilton, ON L8V 5C2, Canada
[2] Peel Reg Canc Program, Dept Radiat Oncol, Mississauga, ON, Canada
[3] McMaster Univ, Dept Med Phys & Appl Radiat Sci, Hamilton, ON, Canada
[4] McMaster Univ, Dept Radiol, Hamilton, ON, Canada
[5] McMaster Univ, Dept Med, Hamilton, ON, Canada
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2005年 / 62卷 / 03期
关键词
breast cancer; mycosis fungoides; total skin electron radiotherapy; in vivo dosimetry; radiochromic film;
D O I
10.1016/j.ijrobp.2005.02.057
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To demonstrate the viability of radiochromic film as an in vivo, two-dimensional dosimeter for the measurement of underdosed areas in patients undergoing total skin electron beam (TSEB) radiotherapy. The results were compared with thermoluminescent dosimeter measurements. Methods and Materials: Dosimetry results are reported for an inframammary fold of 2 patients treated using a modified version of the Stanford six-position (i.e., six-field and dual-beam) TSEB technique. The results are presented as contour plots of film optical density and percentage of dose. A linear dose profile measured from film was compared with the thermoluminescent dosimeter measurements. Results: The results showed that the percentage doses as measured by film are in good agreement with those measured by the thermoluminescent dosimeters. The isodose contour plots provided by film can be used as a two-dimensional dose map for a patient when determining the size of the supplemental patch fields. Conclusion: Radiochromic film is a viable dosimetry tool that the radiation oncologist can use to understand the surface dose heterogeneity better across complex concave regions of skin to help establish more appropriate margins to patch underdosed areas. Film could be used for patients undergoing TSEB for disorders such as mycosis fungoides or undergoing TSEB or regional skin electron beam for widespread skin metastases from breast cancer and other malignancies. (c) 2005 Elsevier Inc.
引用
收藏
页码:920 / 924
页数:5
相关论文
共 11 条
[1]  
Anacak Yavuz, 2003, Med Dosim, V28, P31, DOI 10.1016/S0958-3947(02)00235-2
[2]   Spatial resolution of a stacked radiochromic film dosimeter [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
RADIOTHERAPY AND ONCOLOGY, 2001, 61 (02) :211-213
[3]   Composite depth dose measurement for total skin electron (TSE) treatments using radiochromic film [J].
Gamble, LM ;
Farrell, TJ ;
Jones, GW ;
Hayward, JE .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (07) :891-898
[4]   Total skin electron beam radiotherapy for patients who have mycosis fungoides [J].
Jones, G ;
Wilson, LD ;
Fox-Goguen, L .
HEMATOLOGY-ONCOLOGY CLINICS OF NORTH AMERICA, 2003, 17 (06) :1421-+
[5]   Total skin electron radiation in the management of mycosis fungoides: Consensus of the European Organization for Research and Treatment of Cancer (EORTC) Cutaneous Lymphoma Project Group [J].
Jones, GW ;
Kacinski, BM ;
Wilson, LD ;
Willemze, R ;
Spittle, M ;
Hohenberg, G ;
Handl-Zeller, L ;
Trautinger, F ;
Knobler, R .
JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2002, 47 (03) :364-370
[6]   ELECTRON-BEAM TREATMENT FOR CUTANEOUS T-CELL LYMPHOMA [J].
JONES, GW ;
HOPPE, RT ;
GLATSTEIN, E .
HEMATOLOGY-ONCOLOGY CLINICS OF NORTH AMERICA, 1995, 9 (05) :1057-&
[7]  
JONES GW, 2003, INT J RAD ONCOL B S, V57, pS29
[8]  
Karzmark CJAJ, 1987, 23 AAPM STANF U SCH
[9]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115
[10]  
Peters V G, 1995, Med Dosim, V20, P99, DOI 10.1016/0958-3947(95)00013-M