Digitally recycled incubators: better economic alternatives to modern systems in low-income countries

被引:6
作者
Amadi, H. O. [1 ]
Mokuolu, O. A.
Adimora, G. N.
Pam, S. D.
Etawo, U. S.
Ohadugha, C. O.
Adesiyun, O. O.
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Bioengn, London SW7 2BX, England
[2] Univ Ilorin, Teaching Hosp, Dept Paediat, Ilorin, Nigeria
[3] Univ Nigeria, Teaching Hosp, Enugu, Nigeria
[4] Univ Jos, Teaching Hosp, Jos, Nigeria
[5] Univ Port Harcourt, Teaching Hosp, Port Harcourt, Nigeria
[6] Fed Med Ctr, Owerri, Nigeria
来源
ANNALS OF TROPICAL PAEDIATRICS | 2007年 / 27卷 / 03期
关键词
D O I
10.1179/146532807X220325
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The need to maintain a neutral thermal environment is critical to newborn care. Aim: To investigate reasons for the insufficiency of functional incubators and develop a cost-effective technique for using electronic digital components to recycle obsolete incubators in Nigeria. Methods: Following interview of 84 clinicians and administrators in Nigerian hospitals, it was identified that inadequate funding was the main reason for lack of functional incubators. Two groups of incubator units were then created and their performance compared. Sixteen units of modern (group A) and 19 units of obsolete (group B) incubators were obtained from six hospitals. An assembly design applying independent generic components for recycling systems was specified and produced. These were sourced through the internet at competitive cost and fitted into the reconstructed panels of the obsolete systems. The functional performance of each recycled system was rigorously monitored for 6 months and graded using ten performance indices. The same indices were used to quantify group A systems. Results: The performance of the recycled incubators (group B) was found to be similar to those of modern incubators. Group B's cost index was found to be 25% of that of group A's. Conclusion: Appropriate incubator recycling is a cost-effective method of re-equipping hospitals in low-income countries.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 23 条
  • [1] BACH V, 1999, RBM NEWS, V21, P6
  • [2] BACH V, 1995, RBM NEWS, V17, P7
  • [3] Skin-to-skin (kangaroo) care, respiratory control, and thermoregulation
    Bohnborst, B
    Heyne, T
    Peter, CS
    Poets, CF
    [J]. JOURNAL OF PEDIATRICS, 2001, 138 (02) : 193 - 197
  • [4] Boo N. Y., 2006, SMJ Singapore Medical Journal, V47, P757
  • [5] Bosque E M, 1995, J Obstet Gynecol Neonatal Nurs, V24, P219, DOI 10.1111/j.1552-6909.1995.tb02466.x
  • [6] Critical dependence of acetate thermal mattress on gel activation temperature
    Carmichael, A.
    McCullough, S.
    Kempley, S. T.
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2007, 92 (01): : F44 - F45
  • [7] Kangaroo mother care for low birthweight infants: a randomized controlled trial in different settings
    Cattaneo, A
    Davanzo, R
    Worku, B
    Surjono, A
    Echeverria, M
    Bedri, A
    Haksari, E
    Osorno, L
    Gudetta, B
    Setyowireni, D
    Quintero, S
    Tamburlini, G
    [J]. ACTA PAEDIATRICA, 1998, 87 (09) : 976 - 985
  • [8] Kangaroo mother care
    Doyle, LW
    [J]. LANCET, 1997, 350 (9093) : 1721 - 1722
  • [9] Dry heat loss in incubator: comparison of two premature newborn sized manikins
    Elabbassi, EB
    Beighazi, K
    Delanaud, S
    Libert, JP
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2004, 92 (06) : 679 - 682
  • [10] JI XC, 1993, CHINESE MED J-PEKING, V106, P428