Low variation and high reproducibility in plaque volume with intravascular ultrasound

被引:11
作者
Jensen, LO
Thayssen, P
Pedersen, KE
Stender, S
Haghfelt, T
机构
[1] Odense Univ Hosp, Dept Cardiol, Catheterizat Lab, DK-5000 Odense, Denmark
[2] Gentofte Univ Hosp, Dept Clin Biochem, Gentofte, Denmark
关键词
atherosclerotic regression and progression; 3D IVUS; reproducibility;
D O I
10.1016/j.ijcard.2003.10.036
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Intravascular ultrasound (IVUS) has several advantages compared to angiography when evaluating coronary atherosclerosis in the vessel wall. Methods; The accuracy, reproducibility, and short-time spontaneous variation in volume of vessel, plaque and lumen were studied by electrocardiographic-gated three-dimensional (3D) IVUS in 20 male patients with ischaemic heart disease (IHD). Results: The study lesions were angiographically insignificant, with a length of the analysed segment on 11.4 +/- 5.9 mm. At baseline the mean minimal lumen diameter was 2.41 +/- 0.59 mm, minimal lumen area 4.82 +/- 2.38 mm(2), and maximal plaque burden 65.61 +/- 9.57%. Mean reference diameter was 3.1 +/- 0.6 mm. No significant changes were observed in volumes of total vessel, lumen or plaque. The coefficient of variation (CV) for two volume measurements at baseline was: vessel 0.8%, plaque 1.3%, and lumen 1.4%. For measurements recorded at baseline and after 12.6 +/- 1.5 weeks, CV was respectively 3.5%, 3.3% and 6.6%. Reproducibility and interobserver and intraobserver variation showed very high correlations. A linear correlation was present in percent changes over 12.6 1.5 weeks between vessel volume and lumen volume (r= 0.804; p < 0.001) and between percent changes in plaque volume and vessel volume (r= 0.581; p= 0.007). No correlation was found between changes in plaque volume and lumen volume (r=0.015; p=0.950). Conclusion: ECG-gated 3D IVUS is a highly reproducible method when applied on coronary artery atherosclerosis. CV for lumen volume over 12.6 +/- 1.5 weeks is twice that of plaque volume indicating the superiority of the 3D IVUS compared to coronary angiography (CAG). (c) 2004 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 24 条
  • [1] Aengevaeren WRM, 1997, CIRCULATION, V96, P429
  • [2] Bruining N, 1998, CATHETER CARDIO DIAG, V43, P254, DOI 10.1002/(SICI)1097-0304(199803)43:3<254::AID-CCD3>3.0.CO
  • [3] 2-8
  • [4] PROGRESSION AND REGRESSION OF THE ATHEROSCLEROTIC PLAQUE
    DEFEYTER, PJ
    VOS, J
    DECKERS, JW
    [J]. EUROPEAN HEART JOURNAL, 1995, 16 : 26 - 30
  • [5] CORONARY-ARTERY REMODELING IN ATHEROSCLEROTIC DISEASE - AN INTRAVASCULAR ULTRASONIC STUDY IN-VIVO
    GE, J
    ERBEL, R
    ZAMORANO, J
    KOCH, L
    KEARNEY, P
    GORGE, G
    GERBER, T
    MEYER, J
    [J]. CORONARY ARTERY DISEASE, 1993, 4 (11) : 981 - 986
  • [6] GIBBONS GH, 1994, NEW ENGL J MED, V330, P1431
  • [7] COMPENSATORY ENLARGEMENT OF HUMAN ATHEROSCLEROTIC CORONARY-ARTERIES
    GLAGOV, S
    WEISENBERG, E
    ZARINS, CK
    STANKUNAVICIUS, R
    KOLETTIS, GJ
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1987, 316 (22) : 1371 - 1375
  • [8] INVIVO VALIDATION OF COMPENSATORY ENLARGEMENT OF ATHEROSCLEROTIC CORONARY-ARTERIES
    HERMILLER, JB
    TENAGLIA, AN
    KISSLO, KB
    PHILLIPS, HR
    BASHORE, TM
    STACK, RS
    DAVIDSON, CJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) : 665 - 668
  • [9] LIMITATIONS OF ANGIOGRAPHY FOR ANALYZING CORONARY ATHEROSCLEROSIS PROGRESSION OR REGRESSION
    HONG, MK
    MINTZ, GS
    POPMA, JJ
    KENT, KM
    PICHARD, AD
    SATLER, LF
    LEON, MB
    [J]. ANNALS OF INTERNAL MEDICINE, 1994, 121 (05) : 348 - 354
  • [10] FOCAL COMPENSATORY ENLARGEMENT OF HUMAN ARTERIES IN RESPONSE TO PROGRESSIVE ATHEROSCLEROSIS - IN-VIVO DOCUMENTATION USING INTRAVASCULAR ULTRASOUND
    LOSORDO, DW
    ROSENFIELD, K
    KAUFMAN, J
    PIECZEK, A
    ISNER, JM