Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation

被引:20
|
作者
Chandola, R. M. [1 ,2 ]
Tiwari, S. [3 ]
Kowar, M. K. [4 ]
Choudhary, V. [1 ,2 ]
机构
[1] Pt JNM Med Coll, Dept Radiotherapy, Raipur, Chhattisgarh, India
[2] Dr BRAM Hosp, Raipur, Chhattisgarh, India
[3] Bhilai Inst Technol, Dept Appl Phys, Durg, Chhattisgarh, India
[4] Bhilai Inst Technol, Dept Elect, Durg, Chhattisgarh, India
关键词
Brachytherapy; inhomogeneity; Monte Carlo; radiation-tissue interaction; DOSIMETRY; TLD;
D O I
10.4103/0973-1482.63567
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The presence of least dense dry air and highly dense cortical bone in the path of radiation and the position of source, near or far from the surface of patient, affects the exact dose delivery like in breast brachytherapy. Aim: This study aims to find out the dose difference in the presence of inhomogenieties like cortical bone and dry air as well as to find out difference of dose due to position of source in water phantom of high dose rate (HDR) (192) Ir nucletron microselectron v2 (mHDRv2) brachytherapy source using Monte Carlo (MC) simulation EGSnrc code, so that the results could be used in Treatment Planning System (TPS) for more precise brachytherapy treatment. Settings and Design: The settings and design are done using different software of the computer. Methods and Materials: For this study, the said source, water phantom of volume 30 x 30 x 30 cm (3) , inhomogeneities each of volume 1 x 2 x 2 cm (3) with their position, water of water phantom and position of source are modeled using three-dimensional MC EGSnrc code. Statistical Analysis Used: Mean and probability are used for results and discussion. Results : The relative dose difference is calculated here as 5.5 to 6.5 higher and 4.5 to 5 lower in the presence of air and cortical bone respectively at transverse axis of the source, which may be due to difference of linear attenuation coefficients of the inhomogeneities. However, when the source was positioned at 1 cm distance from the surface of water phantom, the near points between 1 to 2 cm and 3 to 8 cm. from the source, at its transverse axis, were 2 to 3.5 and 4 to 16 underdose to the dose when the source was positioned at mid-point of water phantom. This may be due to lack of back scatter material when the source was positioned very near to the surface of said water phantom and overlap of the additional cause of missing scatter component with the primary dose for near points from the source. These results were found in good agreement with literature data. Conclusion: The results can be used in TPS.
引用
收藏
页码:54 / 57
页数:4
相关论文
共 50 条
  • [1] A Monte Carlo study on dose distribution evaluation of Flexisource Ir-192 brachytherapy source
    Alizadeh, Majid
    Ghorbani, Mandi
    Haghparast, Abbas
    Zare, Naser
    Moghaddas, Toktam Ahmadi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2015, 20 (03) : 204 - 209
  • [2] Dosimetric investigation of high dose rate Ir-192 source with Monte Carlo method
    Acun, H.
    Bozkurt, A.
    Kemikler, G.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2017, 15 (03): : 241 - 249
  • [3] Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    MEDICAL PHYSICS, 2000, 27 (11) : 2521 - 2527
  • [4] Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Sarabiasl, Akbar
    Ayoobian, Navid
    Jabbari, Iraj
    Poorbaygi, Hossein
    Javanshir, Mohammad Reza
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 413 - 422
  • [5] Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    Kouwenhoven, E
    Baltas, D
    MEDICAL PHYSICS, 2003, 30 (01) : 9 - 16
  • [6] EVALUATION OF THE BUILDUP EFFECT OF AN IR-192 HIGH DOSE-RATE BRACHYTHERAPY SOURCE
    PARK, HC
    ALMOND, PR
    MEDICAL PHYSICS, 1992, 19 (05) : 1293 - 1297
  • [7] Dosimetry of indigenously developed Ir-192 high-dose rate brachytherapy source: An EGSnrc Monte Carlo study
    Sahoo, Sridhar
    Selvam, T. Palani
    Sharma, S. D.
    Das, Trupti
    Dey, A. C.
    Patil, B. N.
    Sastri, K. V. S.
    JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 115 - 122
  • [8] A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Sandilos, P
    Dardoufas, K
    Vlachos, L
    MEDICAL PHYSICS, 1999, 26 (08) : 1498 - 1502
  • [9] Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Sakelliou, L
    Sandilos, P
    Antypas, C
    Vlachos, L
    Koutsouveli, E
    MEDICAL PHYSICS, 1998, 25 (10) : 1975 - 1984
  • [10] EXPERIMENTAL VALIDATION OF MONTE-CARLO DOSE CALCULATIONS ABOUT A HIGH-INTENSITY IR-192 SOURCE FOR PULSED DOSE-RATE BRACHYTHERAPY
    VALICENTI, RK
    KIROV, AS
    MEIGOONI, AS
    MISHRA, V
    DAS, RK
    WILLIAMSON, JF
    MEDICAL PHYSICS, 1995, 22 (06) : 821 - 829