Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation

被引:20
作者
Chandola, R. M. [1 ,2 ]
Tiwari, S. [3 ]
Kowar, M. K. [4 ]
Choudhary, V. [1 ,2 ]
机构
[1] Pt JNM Med Coll, Dept Radiotherapy, Raipur, Chhattisgarh, India
[2] Dr BRAM Hosp, Raipur, Chhattisgarh, India
[3] Bhilai Inst Technol, Dept Appl Phys, Durg, Chhattisgarh, India
[4] Bhilai Inst Technol, Dept Elect, Durg, Chhattisgarh, India
关键词
Brachytherapy; inhomogeneity; Monte Carlo; radiation-tissue interaction; DOSIMETRY; TLD;
D O I
10.4103/0973-1482.63567
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The presence of least dense dry air and highly dense cortical bone in the path of radiation and the position of source, near or far from the surface of patient, affects the exact dose delivery like in breast brachytherapy. Aim: This study aims to find out the dose difference in the presence of inhomogenieties like cortical bone and dry air as well as to find out difference of dose due to position of source in water phantom of high dose rate (HDR) (192) Ir nucletron microselectron v2 (mHDRv2) brachytherapy source using Monte Carlo (MC) simulation EGSnrc code, so that the results could be used in Treatment Planning System (TPS) for more precise brachytherapy treatment. Settings and Design: The settings and design are done using different software of the computer. Methods and Materials: For this study, the said source, water phantom of volume 30 x 30 x 30 cm (3) , inhomogeneities each of volume 1 x 2 x 2 cm (3) with their position, water of water phantom and position of source are modeled using three-dimensional MC EGSnrc code. Statistical Analysis Used: Mean and probability are used for results and discussion. Results : The relative dose difference is calculated here as 5.5 to 6.5 higher and 4.5 to 5 lower in the presence of air and cortical bone respectively at transverse axis of the source, which may be due to difference of linear attenuation coefficients of the inhomogeneities. However, when the source was positioned at 1 cm distance from the surface of water phantom, the near points between 1 to 2 cm and 3 to 8 cm. from the source, at its transverse axis, were 2 to 3.5 and 4 to 16 underdose to the dose when the source was positioned at mid-point of water phantom. This may be due to lack of back scatter material when the source was positioned very near to the surface of said water phantom and overlap of the additional cause of missing scatter component with the primary dose for near points from the source. These results were found in good agreement with literature data. Conclusion: The results can be used in TPS.
引用
收藏
页码:54 / 57
页数:4
相关论文
共 13 条
[1]   Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source [J].
Angelopoulos, A ;
Baras, P ;
Sakelliou, L ;
Karaiskos, P ;
Sandilos, P .
MEDICAL PHYSICS, 2000, 27 (11) :2521-2527
[2]   Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources [J].
Baltas, D ;
Geramani, K ;
Ioannidis, GT ;
Hierholz, K ;
Rogge, B ;
Kolotas, C ;
Müller-Sievers, K ;
Milickovic, N ;
Kober, B ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03) :653-661
[3]   Beta versus gamma dosimetry close to Ir-192 brachytherapy sources [J].
Baltas, D ;
Karaiskos, P ;
Papagiannis, P ;
Sakelliou, L ;
Loeffler, E ;
Zamboglou, N .
MEDICAL PHYSICS, 2001, 28 (09) :1875-1882
[4]  
CULLEN D, 1997, 50400 UCRL LOW LIV N
[5]   Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source [J].
Daskalov, GM ;
Loffler, E ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (11) :2200-2208
[6]   Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source [J].
Karaiskos, P ;
Angelopoulos, A ;
Sakelliou, L ;
Sandilos, P ;
Antypas, C ;
Vlachos, L ;
Koutsouveli, E .
MEDICAL PHYSICS, 1998, 25 (10) :1975-1984
[7]   TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy [J].
Kirov, AS ;
Williamson, JF ;
Meigooni, AS ;
Zhu, Y .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) :2015-2036
[8]  
Lux I., 1991, Monte Carlo Particle Transport Methods: Neutron and Photon Calculations
[9]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[10]   Phantom size in brachytherapy source dosimetric studies [J].
Pérez-Calatayud, J ;
Granero, D ;
Ballester, F .
MEDICAL PHYSICS, 2004, 31 (07) :2075-2081