The development of a novel dry powder inhaler

被引:19
作者
Zhang, Xi [1 ,2 ]
Ma, Yingliang [1 ]
Zhang, Liqiang [1 ]
Zhu, Jingxu [1 ]
Jin, Fang [2 ]
机构
[1] Univ Western Ontario, London, ON N6A 5B9, Canada
[2] China State Inst Pharmaceut Ind, Shanghai Inst Pharmaceut Ind, Shanghai Engn Res Ctr Resp Med, Shanghai 200437, Peoples R China
关键词
Dry powder inhaler; Active inhaler; In vitro evaluation; Animal test; Fluorescent image; PULMONARY DRUG-DELIVERY; PHARMACOKINETICS; PERFORMANCE; FORMULATION; MICROSCOPY; SALBUTAMOL; EXCIPIENTS; LUNGS;
D O I
10.1016/j.ijpharm.2012.04.019
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
A novel active and multi-dose dry powder inhaler (DPI) was developed and evaluated to deliver a small quantity (100-500 mu g) of pure drug without any excipient. This dry powder inhaler utilized two compressed air flows to dispense and deliver drug powder: the primary flow aerosolizes the drug powder from its pocket and the secondary flow further disperses the aerosol. In vitro tests by Anderson Cascade Impactor (ACI) indicated that the fine particle fraction (FPF) (<4.7 mu m) of drug delivery could reach over a range of 50-70% (w/w). Emitted dose tests showed that delivery efficiency was above 85% and its relative standard deviation (RSD) was under 10%. Confocal microscopy was used to confirm the deposition of fluorescently labeled spray-dried powder in rabbit lungs. Also, a chromatographic method was used to quantify drug deposition. The results of animal tests showed that 57% of aerosol deposited in the rabbit lung and 24% deposited in its trachea. All the results implied that this novel active dry powder inhaler could efficiently deliver a small quantity of fine drug particles into the lung with quite high fine particle fraction. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:45 / 52
页数:8
相关论文
共 27 条
[1]   Influence of formulation excipients and physical characteristics of inhalation dry powders on their aerosolization performance [J].
Bosquillon, C ;
Lombry, C ;
Préat, V ;
Vanbever, R .
JOURNAL OF CONTROLLED RELEASE, 2001, 70 (03) :329-339
[2]   Predicting the physical state of spray dried composites: salbutamol sulphate/lactose and salbutamol sulphate/polyethylene glycol co-spray dried systems [J].
Corrigan, DO ;
Corrigan, OI ;
Healy, AM .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2004, 273 (1-2) :171-182
[3]   Dry powder platform for pulmonary drug delivery [J].
Daniher, Derek Ivan ;
Zhu, Jesse .
PARTICUOLOGY, 2008, 6 (04) :225-238
[4]   Recent advances in pulmonary drug delivery using large, porous inhaled particles [J].
Edwards, DA ;
Ben-Jebria, A ;
Langer, R .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 85 (02) :379-385
[5]   PHARMACOKINETICS AND PHARMACODYNAMICS OF NITRENDIPINE IN HEALTHY-SUBJECTS AND PATIENTS WITH KIDNEY AND LIVER-DISEASE [J].
EICHELBAUM, M ;
MIKUS, G ;
MAST, V ;
FISCHER, C ;
KUHLMANN, U ;
MACHLEIDT, C .
JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1988, 12 :S6-S10
[6]  
Frijlink H W, 2004, Expert Opin Drug Deliv, V1, P67, DOI 10.1517/17425247.1.1.67
[7]   PHARMACOKINETICS AND ABSOLUTE BIOAVAILABILITY OF SALBUTAMOL IN HEALTHY ADULT VOLUNTEERS [J].
GOLDSTEIN, DA ;
TAN, YK ;
SOLDIN, SJ .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 1987, 32 (06) :631-634
[8]   Effect of Particle Shape on Dry Particle Inhalation: Study of Flowability, Aerosolization, and Deposition Properties [J].
Hassan, Meer Saiful ;
Lau, Raymond Wai Man .
AAPS PHARMSCITECH, 2009, 10 (04) :1252-1262
[9]  
HASSAN MS, 2009, IFMBE PROC, P1434
[10]   Deep tissue two-photon microscopy [J].
Helmchen, F ;
Denk, W .
NATURE METHODS, 2005, 2 (12) :932-940