Assessment of Test Performance and Potential for Environmental Contamination Associated with a Point-of-Care Molecular Assay for Group A Streptococcus in an End User Setting

被引:12
作者
Donato, Leslie J. [1 ]
Myhre, Nikki K. [1 ]
Murray, Martha A. [2 ]
McDonah, Margaret R. [2 ]
Myers, Jane F. [2 ]
Maxson, Julie A. [2 ]
Meek, Alicia M. [2 ]
Espy, Mark J. [1 ]
Furst, Joseph W. [2 ]
Karon, Brad S. [1 ]
Binnicker, Matthew J. [1 ]
机构
[1] Mayo Clin, Dept Lab Med & Pathol, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Family Med, Rochester, MN USA
关键词
Group A Streptococcus; Liat; point of care; ESWAB TRANSPORT-SYSTEM; LIAT(R) INFLUENZA A/B; PCR; CHEMOPROPHYLAXIS;
D O I
10.1128/JCM.01629-18
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Although U.S. Food and Drug Administration-approved and CLIA-waived point-of-care (POC) molecular systems are being implemented in routine clinical practice, instrument reliability, test performance in the hands of end users, and the potential for environmental contamination resulting from use of POC molecular systems have not been extensively evaluated. We performed a prospective evaluation of the Roche cobas Liat group A streptococcus (GAS) assay compared to routine real-time PCR. We evaluated test accuracy, instrument failure rate, and monitored for environmental contamination when testing was performed by minimally trained end users in an Express Care Clinic environment. The overall concordance of the Liat GAS assay with routine testing was 97.2% (455/468). The average Liat failure rate across three analyzers was 6.6% (33/501) (range, 3.7 to 11.6%), and no environmental contamination was detected during the course of the study. The cobas Liat platform and GAS assay demonstrated reliable performance in the end user setting and may serve as a rapid, POC option for routine diagnostic testing for certain infectious diseases, including GAS.
引用
收藏
页数:5
相关论文
共 11 条
[1]   Direct Detection of Influenza A and B Viruses in Less Than 20 Minutes Using a Commercially Available Rapid PCR Assay [J].
Binnicker, Matthew J. ;
Espy, Mark J. ;
Irish, Cole L. ;
Vetter, Emily A. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (07) :2353-2354
[2]  
Centor R M, 1981, Med Decis Making, V1, P239, DOI 10.1177/0272989X8100100304
[3]  
Fiore Anthony E., 2011, Morbidity and Mortality Weekly Report, V60, P1
[4]   Multi-center evaluation of the cobas® Liat® Influenza A/B & RSV assay for rapid point of care diagnosis [J].
Gibson, Jane ;
Schechter-Perkins, Elissa M. ;
Mitchell, Patricia ;
Mace, Sharon ;
Tian, Yu ;
Williams, Kemi ;
Luo, Robert ;
Yen-Lieberman, Belinda .
JOURNAL OF CLINICAL VIROLOGY, 2017, 95 :5-9
[5]   Seasonal Influenza in Adults and Children-Diagnosis, Treatment, Chemoprophylaxis, and Institutional Outbreak Management: Clinical Practice Guidelines of the Infectious Diseases Society of America [J].
Harper, Scott A. ;
Bradley, John S. ;
Englund, Janet A. ;
File, Thomas M. ;
Gravenstein, Stefan ;
Hayden, Frederick G. ;
McGeer, Allison J. ;
Neuzil, Kathleen M. ;
Pavia, Andrew T. ;
Tapper, Michael L. ;
Uyeki, Timothy M. ;
Zimmerman, Richard K. .
CLINICAL INFECTIOUS DISEASES, 2009, 48 (08) :1003-1032
[6]   Laboratory evaluation of the ESwab transport system for the recovery of carbapenem-resistant Acinetobacter baumannii [J].
Moran-Gilad, J. ;
Schwartz, D. ;
Navon-Venezia, S. ;
Carmeli, Y. .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2012, 31 (07) :1429-1433
[7]   Evaluation of the Copan ESwab transport system for the detection of methicillin-resistant Staphylococcus aureus: a laboratory and clinical study [J].
Smismans, Annick ;
Verhaegen, Jan ;
Schuermans, Annette ;
Frans, Johan .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2009, 65 (02) :108-111
[8]   Comparison of LightCycler PCR, rapid antigen immunoassay, and culture for detection of group A streptococci from throat swabs [J].
Uhl, JR ;
Adamson, SC ;
Vetter, EA ;
Schleck, CD ;
Harmsen, WS ;
Iverson, LK ;
Santrach, PJ ;
Henry, NK ;
Cockerill, FR .
JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (01) :242-249
[9]   Comparison of the Copan ESwab system with two Amies agar swab transport systems for maintenance of microorganism viability [J].
Van Horn, Kenneth G. ;
Audette, Carol D. ;
Sebeck, Denise ;
Tuckert, Kelly A. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (05) :1655-1658
[10]   Accurate Detection of Streptococcus pyogenes at the Point of Care Using the cobas Liat Strep A Nucleic Acid Test [J].
Wang, Fangnian ;
Tian, Yu ;
Chen, Lingjun ;
Luo, Robert ;
Sickler, Joanna ;
Liesenfeld, Oliver ;
Chen, Shuqi .
CLINICAL PEDIATRICS, 2017, 56 (12) :1128-1134