Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead

被引:5
作者
Lam, Anna [1 ,2 ,3 ]
Buehler, Stefan [1 ]
Goulouti, Eleni [1 ]
Sweda, Romy [1 ,4 ]
Haeberlin, Andreas [1 ,4 ]
Medeiros-Domingo, Argelia [1 ]
Servatius, Helge [1 ]
Seiler, Jens [1 ]
Baldinger, Samuel [1 ]
Noti, Fabian [1 ]
Tanner, Hildegard [1 ]
Roten, Laurent [1 ]
机构
[1] Univ Bern, Dept Cardiol, Inselspital, Univ Hosp Bern, Freiburgstr, CH-3010 Bern, Switzerland
[2] Bordeaux Univ Hosp CHU, Electrophysiol & Ablat Unit, Bordeaux, France
[3] Fdn Bordeaux Univ, Electrophysiol & Heart Modeling Inst, IHU Liryc, Bordeaux, France
[4] Univ Bern, ARTORG Ctr Biomed Engn, CH-3008 Bern, Switzerland
关键词
Defibrillator lead; Lead failure; Linox; Riata; Sprint fidelis; EXPERT CONSENSUS STATEMENT; INAPPROPRIATE SHOCKS; INTEGRITY ALERT; DEFIBRILLATOR; ALGORITHM;
D O I
10.1007/s10840-018-0486-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeTo compare lead failure manifestation and lead performance of the Biotronik Linox/Sorin Vigila defibrillator lead (Linox group) with the St. Jude Medical Riata/Riata ST (Riata group) and Medtronic Sprint Fidelis defibrillator leads (Fidelis group).MethodsWe assessed the performance of all aforementioned leads implanted at our center and investigated the manifestation of lead failures.ResultsOf 93 Linox, 86 Riata, and 81 Fidelis leads implanted at our center, 11 (12%), 22 (26%), and 25 (31%) leads failed during a median follow-up of 46, 61, and 84months, respectively. Inappropriate shocks were delivered in 64% (Linox), 5% (Riata), and 32% (Fidelis) of lead failures; a device alert was noted in none (Linox), 5% (Riata), and 52% (Fidelis); and lead failure was a coincidental finding in 36% (Linox), 91% (Riata), and 16% (Fidelis) of cases (p<0.001). Non-physiological high rate signals were observed in 73% (Linox), 27% (Riata), and 80% (Fidelis) of lead failures (p=0.001) and damaged lead integrity was found in 36% (Linox), 73% (Riata), and 24% (Fidelis) of cases (p=0.064). Lead survival at 5years was 88%, 92%, and 71% for Linox, Riata, and Fidelis group, respectively.ConclusionsThe most frequent clinical manifestation of lead failure was inappropriate shocks for Linox, coincidental finding for Riata and device alert for Fidelis leads. Non-physiological high rate signals were frequently observed in Linox and Fidelis lead failures whereas in Riata lead failures, a damaged lead integrity was the predominant finding.
引用
收藏
页码:161 / 170
页数:10
相关论文
共 21 条
[1]   Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads [J].
Abdelhadi, Raed H. ;
Saba, Samir F. ;
Ellis, Christopher R. ;
Mason, Pamela K. ;
Kramer, Daniel B. ;
Friedman, Paul A. ;
Gura, Melanie T. ;
DiMarco, John P. ;
Mugglin, Andrew S. ;
Reynolds, Matthew R. ;
Bazaz, Raveen R. ;
Retel, Linda Kat Linen ;
Hayes, David L. ;
Hauser, Robert G. .
HEART RHYTHM, 2013, 10 (03) :361-365
[2]   Comparison of Sprint Fidelis and Riata defibrillator lead failure rates [J].
Fazal, Iftikhar A. ;
Shepherd, Ewen J. ;
Tynan, Margaret ;
Plummer, Christopher J. ;
McComb, Janet M. .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (02) :848-852
[3]   Long-Term Evaluation of Biotronik Linox and Linoxsmart Implantable Cardioverter Defibrillator Leads [J].
Good, Eric D. ;
Cakulev, Ivan ;
Orlov, Michael V. ;
Hirsh, David ;
Simeles, John ;
Mohr, Kelly ;
Moll, Phil ;
Bloom, Heather .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (06) :735-742
[4]   Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead [J].
Hauser, Robert G. ;
Kallinen, Linda M. ;
Almquist, Adrian K. ;
Gornick, Charles C. ;
Katsiyiannis, William T. .
HEART RHYTHM, 2007, 4 (07) :892-896
[5]   Longevity of Sprint Fidelis Implantable Cardioverter-Defibrillator Leads and Risk Factors for Failure Implications for Patient Management [J].
Hauser, Robert G. ;
Maisel, William H. ;
Friedman, Paul A. ;
Kallinen, Linda M. ;
Mugglin, Andrew S. ;
Kumar, Kapil ;
Hodge, David O. ;
Morrison, Thomas B. ;
Hayes, David L. .
CIRCULATION, 2011, 123 (04) :358-U247
[6]   Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures [J].
Kallinen, Linda M. ;
Hauser, Robert G. ;
Tang, Chuen ;
Melby, Daniel P. ;
Almquist, Adrian K. ;
Katsiyiannis, William T. ;
Gornick, Charles C. .
HEART RHYTHM, 2010, 7 (08) :1048-1055
[7]   Comparison of longevity and clinical outcomes of implantable cardioverter-defibrillator leads among manufacturers [J].
Kawada, Satoshi ;
Nishii, Nobuhiro ;
Morimoto, Yoshimasa ;
Miyoshi, Akihito ;
Tachibana, Motomi ;
Sugiyama, Hiroyasu ;
Nakagawa, Koji ;
Watanabe, Atsuyuki ;
Morita, Hiroshi ;
Ito, Hiroshi .
HEART RHYTHM, 2017, 14 (10) :1496-1503
[8]   Spectrum of issues detected by an ICD diagnostic alert that utilizes far-field electrograms: Clinical implications [J].
Koneru, Jayanthi N. ;
Kaszala, Karoly ;
Bordachar, Pierre ;
Shehata, Michael ;
Swerdlow, Charles ;
Ellenbogen, Kenneth A. .
HEART RHYTHM, 2015, 12 (05) :957-967
[9]   Fluoroscopic Screening of Asymptomatic Patients Implanted With the Recalled Riata Lead Family [J].
Liu, Jeffrey ;
Rattan, Rohit ;
Adelstein, Evan ;
Barrington, William ;
Bazaz, Raveen ;
Brode, Susan ;
Jain, Sandeep ;
Mendenhall, G. Stuart ;
Nemec, Jan ;
Razak, Eathar ;
Shalaby, Alaa ;
Schwartzman, David ;
Voigt, Andrew ;
Wang, Norman C. ;
Saba, Samir .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (04) :809-814
[10]   Failure rate and conductor externalization in the Biotronik Linox/Sorin Vigila implantable cardioverter-defibrillator lead [J].
Noti, Fabian ;
Lam, Anna ;
Klossner, Nicole ;
Seiler, Jens ;
Servatius, Helge ;
Medeiros-Domingo, Argelia ;
Van Nam Tran ;
Haeberlin, Andreas ;
Fuhrer, Juerg ;
Tanner, Hildegard ;
Roten, Laurent .
HEART RHYTHM, 2016, 13 (05) :1075-1082