Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010

被引:4
作者
Fania, Claudio [1 ]
Albertini, Federica [1 ]
Palatini, Paolo [1 ]
机构
[1] Univ Padua, Dept Med, Via Giustiniani 2, I-35128 Padua, Italy
关键词
cuff; device; office blood pressure; self-measurement; validation;
D O I
10.1097/MBP.0000000000000266
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The aim of this study was to define the accuracy of UM-211, an automated oscillometric device for office use coupled to several cuffs for different arm sizes, according to the International Protocol of the European Society of Hypertension. The validation was performed in 33 individuals. Their mean age was 59.6 +/- 12.9 years, systolic blood pressure (BP) was 144.3 +/- 21.5mmHg (range: 96-184mmHg), diastolic BP was 86.8 +/- 18.5mmHg (range: 48-124 mmHg), and arm circumference was 30.2 +/- 4.3 cm (range: 23-39 cm). Four sequential readings were taken by observers 1 and 2 using a double-headed stethoscope and a mercury sphygmomanometer, whereas three BP readings were taken by the supervisor using the test instrument. The differences between the readings provided by the device and the mean observer measurements were calculated. Therefore, each device measurement was compared with the previous and the next mean observer measurement. The validation results fulfilled all the 2010 European Society of Hypertension revision Protocol criteria for the general population and passed all validation grades. On average, the device overestimated systolic BP by 1.7 +/- 2.4mmHg and diastolic BP by 1.7 +/- 2.5mmHg. These data show that the UM-211 device coupled to several cuffs for different ranges of arm circumference met the requirements for validation according to the International Protocol and can be recommended for clinical use in the adult population. However, these results mainly apply to the use of the 22-32 and the 31-45 cm cuffs. Copyright (C) 2017 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:302 / 305
页数:4
相关论文
共 50 条
[41]   Validation of A&D UA-85X device for blood pressure measurement [J].
Bonso, Elisa ;
Ragazzo, Fabio ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2008, 13 (06) :339-341
[42]   Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [J].
Chahine, Mirna N. ;
Bou Harb, Souad ;
Saad, Abdel Rahman ;
Sarkis, Patrick ;
Azaki, Alaa ;
Harb, Aya ;
Allouch, Ali ;
Asmar, Roland .
BLOOD PRESSURE MONITORING, 2021, 26 (03) :237-241
[43]   Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension [J].
Coleman, Andrew ;
Steel, Stephen ;
DeGreef, Annemarie ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (03) :163-166
[44]   Validation of the Beneware model ABP-021 ambulatory blood pressure monitor according to the revised 2010 European Society of hypertension international protocol [J].
Pereira, Telmo ;
Guimaraes, Joao .
BLOOD PRESSURE MONITORING, 2018, 23 (04) :210-213
[45]   Validation of the Omron HBP-1320 for professional use according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the 2010 revision of the European Society of Hypertension International Protocol [J].
Saito, Kanako ;
Hishiki, Yukiko ;
Takahashi, Hakuo .
BLOOD PRESSURE MONITORING, 2020, 25 (03) :162-166
[46]   European Society of Hypertension International Protocol for the validation of blood pressure monitors: a critical review of its application and rationale for revision [J].
Stergiou, George S. ;
Karpettas, Nikos ;
Atkins, Neil ;
O'Brien, Eoin .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :39-48
[47]   Validation of the visomat comfort eco blood pressure measuring device according to the International Protocol [J].
Perkovic, Davor ;
Guarnieri, Chiara ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2009, 14 (04) :178-180
[48]   Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002 [J].
Lueders, Stephan ;
Krueger, Ralf ;
Zemmrich, Claudia ;
Forstner, Klaus ;
Sturm, Claus-Dieter ;
Bramlage, Peter .
BLOOD PRESSURE MONITORING, 2012, 17 (06) :248-252
[49]   Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension [J].
Pini, Claudio ;
Natalizi, Anna ;
Gerosa, Pietro Francesco ;
Frigerioa, Mauro ;
Omboni, Stefano ;
Parati, Gianfranco .
BLOOD PRESSURE MONITORING, 2008, 13 (03) :177-182
[50]   Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension [J].
Pini, Claudio ;
Pastori, Marco ;
Baccheschi, Jordan ;
Omboni, Stefano ;
Parati, Gianfranco .
BLOOD PRESSURE MONITORING, 2007, 12 (03) :179-184