Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010

被引:4
作者
Fania, Claudio [1 ]
Albertini, Federica [1 ]
Palatini, Paolo [1 ]
机构
[1] Univ Padua, Dept Med, Via Giustiniani 2, I-35128 Padua, Italy
关键词
cuff; device; office blood pressure; self-measurement; validation;
D O I
10.1097/MBP.0000000000000266
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The aim of this study was to define the accuracy of UM-211, an automated oscillometric device for office use coupled to several cuffs for different arm sizes, according to the International Protocol of the European Society of Hypertension. The validation was performed in 33 individuals. Their mean age was 59.6 +/- 12.9 years, systolic blood pressure (BP) was 144.3 +/- 21.5mmHg (range: 96-184mmHg), diastolic BP was 86.8 +/- 18.5mmHg (range: 48-124 mmHg), and arm circumference was 30.2 +/- 4.3 cm (range: 23-39 cm). Four sequential readings were taken by observers 1 and 2 using a double-headed stethoscope and a mercury sphygmomanometer, whereas three BP readings were taken by the supervisor using the test instrument. The differences between the readings provided by the device and the mean observer measurements were calculated. Therefore, each device measurement was compared with the previous and the next mean observer measurement. The validation results fulfilled all the 2010 European Society of Hypertension revision Protocol criteria for the general population and passed all validation grades. On average, the device overestimated systolic BP by 1.7 +/- 2.4mmHg and diastolic BP by 1.7 +/- 2.5mmHg. These data show that the UM-211 device coupled to several cuffs for different ranges of arm circumference met the requirements for validation according to the International Protocol and can be recommended for clinical use in the adult population. However, these results mainly apply to the use of the 22-32 and the 31-45 cm cuffs. Copyright (C) 2017 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:302 / 305
页数:4
相关论文
共 50 条
[31]   Validation of the Pangao PG-800B5 for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010 [J].
Chen, Wan ;
Zeng, Zhaolin ;
Li, Lizhi ;
Wan, Xiaofen ;
Wan, Yi .
BLOOD PRESSURE MONITORING, 2014, 19 (05) :302-305
[32]   'Legacy publication of a 2009 validation of the Riester Big Ben Square Desk aneroid device for blood pressure measurement according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults (2002)' [J].
McNally, Ryan J. ;
Dunkerley, Janette ;
Holland, Maureen ;
Eatough, Ruth ;
Lacy, Peter ;
McManus, Richard J. ;
Chapman, Neil ;
Chowienczyk, Philip J. ;
Lewis, Philip ;
Clark, Christopher E. ;
Denver, Elizabeth ;
Neary, Annette ;
McDonagh, Sinead T. J. ;
Sheppard, James P. .
BLOOD PRESSURE MONITORING, 2024, 29 (04) :203-206
[33]   Validation of the Omron® 705 IT device for blood pressure measurement according to the International validation Protocol [J].
El Assaad, MA ;
Topouchian, JA ;
Labaki, GS ;
Asmar, RG .
AMERICAN JOURNAL OF HYPERTENSION, 2003, 16 (05) :47A-47A
[34]   Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010 [J].
Guo, Wan-Gang ;
Li, Bing-Ling ;
He, Yong ;
Xue, Yu-Sheng ;
Wang, Hai-Yan ;
Zheng, Qiang-Sun ;
Xiang, Ding-Cheng .
BLOOD PRESSURE MONITORING, 2014, 19 (04) :242-245
[35]   Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010 [J].
Chen, Can ;
Shang, Fujun ;
Wang, Jiepin ;
Chen, Jianghong ;
Ji, Na ;
Wan, Yi .
BLOOD PRESSURE MONITORING, 2012, 17 (06) :253-256
[36]   Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol [J].
Coleman, A ;
Freeman, P ;
Steel, S ;
Shennan, A .
BLOOD PRESSURE MONITORING, 2005, 10 (03) :165-168
[37]   Validation of the Omron® HEM-637 IT device for blood pressure measurement according to the International validation Protocol? [J].
El Assaad, MA ;
Topouchian, JA ;
Asmar, RG .
AMERICAN JOURNAL OF HYPERTENSION, 2004, 17 (05) :38A-39A
[38]   Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [J].
Masiero, Serena ;
Fania, Claudio ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2011, 16 (04) :208-210
[39]   Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometric mode, for clinic use and self-measurement in a general population according to the European Society of Hypertension International Protocol, revision 2010 [J].
Masiero, Serena ;
Fania, Claudio ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2011, 16 (05) :262-264
[40]   Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International Protocol [J].
Viera, Anthony J. ;
Hinderliter, Alan L. .
BLOOD PRESSURE MONITORING, 2007, 12 (05) :335-338