Improvement of Electromagnetic Field Distributions Using High Dielectric Constant (HDC) Materials for CTL-Spine MRI: Numerical Simulations and Experiments

被引:4
作者
Park, Bu S. [1 ]
McCright, Brent [1 ]
Angelone, Leonardo M. [2 ]
Razjouyan, Amir [2 ]
Rajan, Sunder S. [2 ]
机构
[1] US FDA, CBER, OTAT Div Cellular & Gene Therapies, Silver Spring, MD 20993 USA
[2] US FDA, CDRH, OSEL, Div Biomed Phys, Silver Spring, MD 20993 USA
关键词
Finite-difference time domain (FDTD); high dielectric materials; robotic measurements; SAR; signal-to-noise ratio (SNR); HIGH-PERMITTIVITY PADS; ABSORPTION RATE; 3; T; MAGNETIC-RESONANCE; BODY COIL; HOMOGENEITY; MODELS; BRAIN;
D O I
10.1109/TEMC.2017.2677581
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This study investigates the use of pads with high dielectric constant (HDC) materials to alter electromagnetic field distributions in patients during magnetic resonance imaging (MRI). The study was performed with numerical simulations and phantom measurements. An initial proof-of-concept and validation was performed using a phantom at 64 MHz, showing increases of up to 10% in electromagnetic field when using distilled water as the high dielectric material. Additionally, numerical simulations with computational-models of human anatomy were performed at 128 MHz. Results of these simulations using barium titanate (BaTiO3) beads showed a 61% increase of vertical bar B-1(+)vertical bar with a quadrature driven RF coil and a 64% increase with a dual-transmit array. The presence of the HDC material also allowed for a decrease of SAR up to twofold (e.g., peak 10 g-averaged SAR from 54 to 22 W/kg with a quadrature driven RF coil and from 27 to 22 W/kg with a dual-transmit array using CaTiO3 powder at 128 MHz). The results of this study show that the use of HDC pads at 128 MHz for MRI spine applications could result in improved magnetic fields within the region of interest, while decreasing SAR outside the region.
引用
收藏
页码:1382 / 1389
页数:8
相关论文
共 31 条
  • [11] The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz
    Gabriel, S
    Lau, RW
    Gabriel, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) : 2251 - 2269
  • [12] New high dielectric constant materials for tailoring the B1+ distribution at high magnetic fields
    Haines, K.
    Smith, N. B.
    Webb, A. G.
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2010, 203 (02) : 323 - 327
  • [13] Heer P., 2012, MAGN RESON MED, V68, P1317
  • [14] IT IS foundations, 2015, DAT BAS TISS PROP
  • [15] Johnk C.T. A., 1988, Engineering Electromagnetic Fields and Waves, V2
  • [16] Assessing the Electromagnetic Fields Generated By a Radiofrequency MRI Body Coil at 64 MHz: Defeaturing Versus Accuracy
    Lucano, Elena
    Liberti, Micaela
    Mendoza, Gonzalo G.
    Lloyd, Tom
    Iacono, Maria Ida
    Apollonio, Francesca
    Wedan, Steve
    Kainz, Wolfgang
    Angelone, Leonardo M.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (08) : 1591 - 1601
  • [17] Permittivity and performance of dielectric pads with sintered ceramic beads in MRI: early experiments and simulations at 3 T
    Luo, Wei
    Lanagan, Michael T.
    Sica, Christopher T.
    Ryu, YeunChul
    Oh, Sukhoon
    Ketterman, Matthew
    Yang, Qing X.
    Collins, Christopher M.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) : 269 - 275
  • [18] Medical Equipment, 2010, 60601233 IEC
  • [19] Whole-Body and Local RF Absorption in Human Models as a Function of Anatomy and Position within 1.5T MR Body Coil
    Murbach, Manuel
    Neufeld, Esra
    Kainz, Wolfgang
    Pruessmann, Klaas P.
    Kuster, Niels
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (02) : 839 - 845
  • [20] Park B., 2013, Journal of Electromagnetic Analysis and Applications, V05, P205