EP radiofrequency generators: Significant offsets between selected and delivered power?

被引:3
作者
Bourier, Felix [1 ]
Schwarz, Bernhard [2 ]
Brkic, Amir [1 ]
Wolff, Lara [1 ]
Semmler, Verena [1 ]
Kottmaier, Marc [1 ]
Risse, Elena [1 ]
Telishevska, Marta [1 ]
Brooks, Stephanie [1 ]
Lengauer, Sarah [1 ]
Kornmayer, Marielouise [1 ]
Koch-Buettner, Katharina [1 ]
Berger, Florian [1 ]
Hessling, Gabriele [1 ]
Deisenhofer, Isabel [1 ]
Reents, Tilko [1 ]
机构
[1] Tech Univ Munich, Dept Electrophysiol, German Heart Ctr Munich, Lazarettstr 36, D-80636 Munich, Germany
[2] Tech Univ Munich, Dept Med Technol, German Heart Ctr Munich, Munich, Germany
关键词
accuracy; catheter ablation; generator; lesion; radiofrequency; FORCE-SENSING CATHETERS; SURROUND FLOW CATHETER; CONTACT FORCE; ATRIAL-FIBRILLATION; ATRIOESOPHAGEAL FISTULA; IMPEDANCE DECREASE; ABLATION CATHETER; LESION SIZE; INSIGHTS; SAFETY;
D O I
10.1111/jce.13386
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: For radiofrequency (RF) ablation, the EP Shuttle (R) (Stockert GmbH, Freiburg, Germany), Ampere (R) (St. Jude Medical, St. Paul, MN, USA), and SmartAblate (R) (Biosense Webster, Diamond Bar, CA, USA) generator models are most frequently used in clinical practice. The aim of this study was to assess the correlation between selected and delivered RF power for different generators. Methods and results: In an experimental setup, ablation catheters were connected to the EP Shuttle (R), Ampere (R), and SmartAblate (R) generators. The power delivered by the generators was measured using a current converter and an oscilloscope. The selected power displayed on the generator was compared to the actually delivered power measured by the experimental setup (n = 800 measurements). The offsets between selected and delivered power increased significantly with impedance (EP Shuttle (R)). For example, at a selected power of 30 W, the delivered power was 40.3 W (EP Shuttle (R)), 30.1 W (Ampere (R)), and 28.1 W (SmartAblate (R)) at an impedance of 200 Omega. In addition, ablation lesions (n = 80) were created in ex vivo porcine cardiac muscle preparations. The resulting ablation lesion size was calculated in caliper measurements. When the EP Shuttle (R) generator was operated at 200 Omega, the resulting lesion size was significantly larger than at 100 Omega. There were no significant offsets between power delivery and lesion size when using the Ampere (R) or SmartAblate (R) generators. Conclusions: The Ampere (R) and SmartAblate (R) generator models deliver accurate power as selected by the user. The power delivered by the EP Shuttle (R) generator exceeds the selected power by up to 40% depending on impedance. The findings were confirmed in ex vivo porcine heart experiments and should be considered in clinical practice.
引用
收藏
页码:330 / 334
页数:5
相关论文
共 14 条
[1]   Risk of atrioesophageal fistula formation with contact force-sensing catheters [J].
Black-Maier, Eric ;
Pokorney, Sean D. ;
Barnett, Adam S. ;
Zeitler, Emily P. ;
Sun, Albert Y. ;
Jackson, Kevin P. ;
Bahnson, Tristram D. ;
Daubert, James P. ;
Piccini, Jonathan P. .
HEART RHYTHM, 2017, 14 (09) :1328-1333
[2]   A nationwide survey on the prevalence of atrioesophageal fistula after left atrial radiofrequency catheter ablation [J].
Ghia, Kasturi K. ;
Chugh, Aman ;
Good, Eric ;
Pelosi, Frank ;
Jongnarangsin, Krit ;
Bogun, Frank ;
Morady, Fred ;
Oral, Hakan .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2009, 24 (01) :33-36
[3]   THE BIOPHYSICS OF RADIOFREQUENCY CATHETER ABLATION IN THE HEART - THE IMPORTANCE OF TEMPERATURE MONITORING [J].
HAINES, DE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :586-591
[4]   Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter [J].
Knecht, Sven ;
Reichlin, Tobias ;
Pavlovic, Nikola ;
Schaer, Beat ;
Osswald, Stefan ;
Sticherling, Christian ;
Kuehne, Michael .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2015, 43 (03) :297-306
[5]   In Vivo Contact Force Analysis and Correlation With Tissue Impedance During Left Atrial Mapping and Catheter Ablation of Atrial Fibrillation [J].
Makimoto, Hisaki ;
Lin, Tina ;
Rillig, Andreas ;
Metzner, Andreas ;
Wohlmuth, Peter ;
Arya, Anita ;
Antz, Matthias ;
Mathew, Shibu ;
Deiss, Sebastian ;
Wissner, Erik ;
Rausch, Peter ;
Bardyszewski, Aleksander ;
Kamioka, Masashi ;
Li, Xuping ;
Kuck, Karl-Heinz ;
Ouyang, Feifan ;
Tilz, Roland Richard .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (01) :46-54
[6]   Pulmonary Vein Isolation Using a Second-Generation Cryoballoon in Patients With Paroxysmal Atrial Fibrillation: One-Year Outcome Using a Single Big-Balloon 3-Minute Freeze Technique [J].
Miyazaki, Shinsuke ;
Hachiya, Hitoshi ;
Nakamura, Hiroaki ;
Taniguchi, Hiroshi ;
Takagi, Takamitsu ;
Hirao, Kenzo ;
Iesaka, Yoshito .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 (12) :1375-1380
[7]   COMPARISON OF IN-VIVO TISSUE TEMPERATURE PROFILE AND LESION GEOMETRY FOR RADIOFREQUENCY ABLATION WITH A SALINE-IRRIGATED ELECTRODE VERSUS TEMPERATURE CONTROL IN A CANINE THIGH MUSCLE PREPARATION [J].
NAKAGAWA, H ;
YAMANASHI, WS ;
PITHA, JV ;
ARRUDA, M ;
WANG, XZ ;
OHTOMO, K ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1995, 91 (08) :2264-2273
[8]   Initial impedance decrease as an indicator of good catheter contact: Insights from radiofrequency ablation with force sensing catheters [J].
Reichlin, Tobias ;
Knecht, Sven ;
Lane, Christopher ;
Kuehne, Michael ;
Nof, Eyal ;
Chopra, Nagesh ;
Tadros, Thomas M. ;
Reddy, Vivek Y. ;
Schaer, Beat ;
John, Roy M. ;
Osswald, Stefan ;
Stevenson, William G. ;
Sticherling, Christian ;
Michaud, Gregory F. .
HEART RHYTHM, 2014, 11 (02) :194-201
[9]   Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: A multicenter registry [J].
Stabile, Giuseppe ;
Di Donna, Paolo ;
Schillaci, Vincenzo ;
Di Monaco, Antonio ;
Iuliano, Assunta ;
Caponi, Domenico ;
Urraro, Francesco ;
Solimene, Francesco ;
Grimaldi, Massimo ;
Scaglione, Marco .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (07) :762-767
[10]   The Incidence of Audible Steam Pops Is Increased and Unpredictable With the ThermoCool® Surround Flow Catheter During Left Atrial Catheter Ablation: A Prospective Observational Study [J].
Theis, Cathrin ;
Rostock, Thomas ;
Mollnau, Hanke ;
Sonnenschein, Sebastian ;
Himmrich, Ewald ;
Kaempfner, Denise ;
Ocete, Blanca Quesada ;
Bock, Karsten ;
Muenzel, Thomas ;
Konrad, Torsten .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (09) :956-962