Defibrillation coil reversal: A rare cause of abnormal noise and inappropriate shocks

被引:3
作者
Jeevanantham, Vinodh [2 ]
Levine, Ethan [1 ]
Budzikowski, Adam S. [1 ]
Shah, Abrar H. [1 ,2 ]
Daubert, James P. [1 ]
机构
[1] Univ Rochester, Med Ctr, Div Cardiol, Dept Med, Rochester, NY 14642 USA
[2] Unity Hosp, New York, NY USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2008年 / 31卷 / 03期
关键词
defibrillation; -; ICD; coil reversal; inappropriate shocks;
D O I
10.1111/j.1540-8159.2008.01000.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a rare case of inadvertent reversal of the defibrillator shock coils during generator change which then resulted in both intermittent, recurrent "noise" on both the rate sensing and the shock electrogram leads and multiple inappropriate shocks.
引用
收藏
页码:375 / 377
页数:3
相关论文
共 6 条
[1]   Malfunction of endocardial defibrillator leads and lead extraction: where do they meet? [J].
Bracke, FALE ;
Meijer, A ;
van Gelder, LM .
EUROPACE, 2002, 4 (01) :19-24
[2]   Implantable cardioverter-defibrillators [J].
DiMarco, JP .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (19) :1836-1847
[3]   Inappropriate sensing of pectoral myopotentials by an implantable cardioverter defibrillator system [J].
Karanam, S ;
John, RM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (05) :688-689
[4]   SUBCLAVIAN CRUSH SYNDROME COMPLICATING TRANSVENOUS CARDIOVERTER-DEFIBRILLATOR SYSTEMS [J].
ROELKE, M ;
ONUNAIN, SS ;
OSSWALD, S ;
GARAN, H ;
HARTHORNE, JW ;
RUSKIN, JN .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :973-979
[5]  
SEIFERT T, 1995, PACE, V19, P1592
[6]   High incidence of device-related and lead-related complications in the dual-chamber implantable cardioverter defibrillator compared with the single-chamber version [J].
Takahashi, T ;
Bhandari, AK ;
Watanuki, M ;
Cannom, DS ;
Sakurada, H ;
Hiraoka, M .
CIRCULATION JOURNAL, 2002, 66 (08) :746-750