Microwave-based digestion method for extraction of 127I and 129I from solid material for measurements by AMS and ICP-MS

被引:21
作者
Gomez-Guzman, J. M. [1 ,2 ]
Enamorado-Baez, S. M. [1 ,3 ]
Pinto-Gomez, A. R. [1 ]
Abril-Hernandez, J. M. [3 ]
机构
[1] CNA, Seville 41092, Spain
[2] Univ Seville, Dpto Fis Atom Mol & Nuclear, Seville, Spain
[3] Univ Seville, Dpto Fis Aplicada 1, EU Ingn Tecn Agr, Seville, Spain
关键词
Iodine-127; Iodine-129; Certified reference materials; Microwave digestion; Inductively coupled plasma mass spectrometry; Accelerator mass spectrometry; PLASMA-MASS-SPECTROMETRY; NEUTRON-ACTIVATION ANALYSIS; IODINE DETERMINATION; BIOLOGICAL SAMPLES; NATURAL-WATERS; MILK POWDER; FOOD; CHROMATOGRAPHY; COMBUSTION; CHLORINE;
D O I
10.1016/j.ijms.2011.01.006
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
This work presents a microwave-based digestion method followed by a radiochemical extraction procedure to extract iodine from environmental matrices that provides iodine in a form suitable for the measurement of I-129 by AMS, with shorter preparation times, small sample sizes and higher automation than previous methods. Samples were digested by a microwave digestion method in closed vessels using HNO3 as oxidizing agent. Following chemical iodine extraction consisted in an organic compound extraction followed by an aqueous solution extraction and iodine precipitation. Prepared samples were used to measure I-127 by ICP-MS and I-129 by AMS. The method was validated by ICP-MS measuring I-127 content in standard reference materials covering a wide variety of biological, soil and sediment matrices: 1547 Peach Leaves, 1537a Tomato Leaves, 1549 Non-Fat Milk Powder, 2704 Buffalo River Sediment, 2711 Montana Soil and 1648 Urban Particulate Matter from the National Institute of Standards and Technology, IAEA-375 Soil and IAEA SL-1 Lake Sediment from the International Atomic Energy Agency and 186 Pig Kidney from Community Bureau of Reference. The recoveries with respect to the reference values were about 90%. Iodine losses during chemical extraction could be due to its volatilization as HI or 12 in acid means. Accurate results for determination in certified materials and good recoveries. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:103 / 108
页数:6
相关论文
共 44 条
[1]   Analytical techniques and applications of 129I in natural water [J].
Buraglio, N ;
Aldahan, A ;
Possnert, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2000, 172 :518-523
[2]   APPLICATION OF ION-EXCHANGE CHROMATOGRAPHY WITH AN ION-SELECTIVE ELECTRODE DETECTOR TO IODINE DETERMINATION IN NATURAL-WATERS [J].
BUTLER, ECV ;
GERSHEY, RM .
ANALYTICA CHIMICA ACTA, 1984, 164 (OCT) :153-161
[3]   Status of the compact 1 MV AMS facility at the Centro Nacional de Aceleradores (Spain) [J].
Chamizo, E. ;
Lopez-Gutierrez, J. M. ;
Ruiz-Gomez, A. ;
Santos, F. J. ;
Garcia-Leon, M. ;
Maden, C. ;
Alfimov, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (10) :2217-2220
[4]  
CREED JT, 1994, METHOD 200 8 DETERMI
[5]   Vapor Generation of Inorganic Anionic Species After Aqueous phase Alkylation with Trialkyloxonium Tetrafluoroborates [J].
D'Ulivo, Alessandro ;
Pagliano, Enea ;
Onor, Massimo ;
Pitzalis, Emanuela ;
Zamboni, Roberto .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :6399-6406
[6]   APPLICATION OF INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY TO THE SIMULTANEOUS DETERMINATION OF CHLORINE, BROMINE AND IODINE IN THE NATIONAL-BUREAU-OF-STANDARDS STANDARD REFERENCE MATERIAL-1648 URBAN PARTICULATE [J].
DATE, AR ;
STUART, ME .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :659-665
[7]   SIMULTANEOUS DETERMINATION OF IODINE AND SELENIUM IN BIOLOGICAL SAMPLES BY RADIOCHEMICAL NEUTRON-ACTIVATION ANALYSIS [J].
DERMELJ, M ;
STIBILJ, V ;
STEKAR, J ;
BYRNE, AR .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1991, 340 (04) :258-261
[8]   ON THE BIOPHILIC NATURE OF IODINE IN SEAWATER [J].
ELDERFIELD, H ;
TRUESDALE, VW .
EARTH AND PLANETARY SCIENCE LETTERS, 1980, 50 (01) :105-114
[9]   A routine preparation method for AMS measurement of 129I in solid material [J].
Englund, E. ;
Aldahan, A. ;
Possnert, G. ;
Alfimov, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 259 (01) :365-369
[10]   Determination of iodine in food samples by inductively coupled plasma mass spectrometry after alkaline extraction [J].
Fecher, PA ;
Goldmann, I ;
Nagengast, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (09) :977-982