Scatter in an uncollimated x-ray CT machine based on a Geant4 Monte Carlo simulation

被引:1
作者
Wadeson, Nicola [1 ]
Morton, Edward [2 ]
Lionheart, William [1 ]
机构
[1] Univ Manchester, Sch Math, Manchester, Lancs, England
[2] Rapiscan Syst Ltd, Surrey, England
来源
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING | 2010年 / 7622卷
基金
英国工程与自然科学研究理事会;
关键词
Compton scatter; Rayleigh scatter; Geant4; Monte-Carlo simulation; x-ray CT; RTT; DIFFRACTION;
D O I
10.1117/12.843981
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A high-speed motionless-gantry x-ray CT machine has been designed to allow for 3D images to be collected in real time. By using multiple, switched x-ray sources and fixed detector rings, the time consuming mechanical rotation of conventional CT machines can be removed. However, the nature of this design limits the possibility of detector collimation since each detector must now be able to record the energy of x-ray beams from a number of different directions. The lack of collimation has implications in the reconstructed image due to an increase in the number of scattered photons recorded. A Monte Carlo computer simulation of the x-ray machine has been developed, using the Geant4 software toolkit, to analyse the behaviour of both Rayleigh and Compton scattered photons when considering airport baggage and medical applications. Four different scattering objects were analysed based on 50kVp, 100kVp and 150kVp spectra for a tungsten target. Two suitcase objects, a body and a brain phantom were chosen as objects typical of airport baggage and medical CT. The results indicate that the level of scatter is negligible for a typical airport baggage application, since the majority of space in a suitcase consists of clothing, which has a low density. Scatter contributes to less than 1% of the image in all instances. However, due to the large amounts of water found in the human body, the level of scatter in the medical instances are significantly higher, reaching 37% when the body phantom is analysed at 50kVp.
引用
收藏
页数:8
相关论文
共 13 条
[1]  
AGNOSTINELLI S, 2006, NUCL INSTRUM METH A, V506, P250
[2]   Energy dispersive X-ray diffraction as a means to identify illicit materials: A preliminary optimisation study [J].
Cook, Emily ;
Fong, Ruby ;
Horrocks, Julie ;
Wilkinson, David ;
Speller, Robert .
APPLIED RADIATION AND ISOTOPES, 2007, 65 (08) :959-967
[3]  
Coursey J., 2005, ATOMIC WEIGHTS ISOTO
[4]  
CRANLEY K, 78 I PHYS ENG MED
[5]  
Cullen D E, 1997, UCRL50400, V6
[6]   COMPTON SCATTER EFFECTS IN CT RECONSTRUCTIONS [J].
GLOVER, GH .
MEDICAL PHYSICS, 1982, 9 (06) :860-867
[7]   SCATTERED RADIATION IN FAN BEAM IMAGING-SYSTEMS [J].
JOHNS, PC ;
YAFFE, M .
MEDICAL PHYSICS, 1982, 9 (02) :231-239
[8]  
JOSEPH PH, 2001, MED PHYS, V28
[9]   The GEANT low energy Compton scattering (GLECS) package for use in simulating advanced Compton telescopes [J].
Kippen, RM .
NEW ASTRONOMY REVIEWS, 2004, 48 (1-4) :221-225
[10]   A CdZnTe array for the detection of explosives in baggage by energy-dispersive X-ray diffraction signatures at multiple scatter angles [J].
Malden, CH ;
Speller, RD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 449 (1-2) :408-415