Application of X-ray fluorescence spectrometry for screening pharmaceutical products for Elemental Impurities according to ICH guideline Q3D

被引:12
作者
Sauer, Benjamin [1 ]
Xiao, Youhong [2 ]
Zoontjes, Michel [2 ]
Kroll, Christian [1 ]
机构
[1] Salutas Pharma GmbH, Otto Von Guericke Allee 1, D-39179 Barleben, Germany
[2] Malvern Panalyt BV, Lelyweg 1, NL-7600 AA Almelo, Netherlands
关键词
X-ray fluorescence; EDXRF; ICH guideline Q3D; ICH Q3D; Elemental impurities; Pharmaceutical products; USP; DIGESTION; MERCURY; CURVE;
D O I
10.1016/j.jpba.2019.113005
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Energy-dispersive X-ray fluorescence spectrometry (EDXRF) is a suitable analytical procedure for screening drug products for Elemental Impurities (EI) according to ICH guideline Q3D. EDXRF represents a cost-efficient, robust and standard-free alternative compared to other methodologies for trace analysis, and therefore utilization of this application should be encouraged. This study demonstrates the capability of EDXRF for EI screening of oral solid dosage drug products (OSD products) within a defined matrix range. Method development and validation focused on class 1 (Cd, Pb, As, Hg) and class 2A (Co, V, Ni) elements, as defined by ICH guideline Q3D. In order to limit validation activities, a novel cluster approach was applied, based on matrix properties. This included comprehensive characterization of method performance parameters for exemplary pharmaceutical matrices and demonstration of LOQ independence from matrix effects by using a set of limit samples representing typical matrix variations of OSD products. The methodology can be used as a limit test for class 1 and class 2A elements and is fully compliant with method validation requirements according to the European Pharmacopeia. The novelty of the present work is the application of EDXRF for a routine screening of OSD products for Elemental Impurities within the pharmaceutical industry beyond previously published feasibility studies for a limited number of pharmaceutical raw materials or products. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页数:8
相关论文
共 23 条
  • [1] Investigation of analyte losses using microwave-assisted sample digestion and closed vessels with venting
    Amaral, Clarice D. B.
    Fialho, Lucimar L.
    Camargo, Felipe P. R.
    Pirola, Camillo
    Nobrega, Joaquim A.
    [J]. TALANTA, 2016, 160 : 354 - 359
  • [2] [Anonymous], 2019, INT COUNCIL HARMONIZ
  • [3] [Anonymous], 2018, 20237 EUR PHARM
  • [4] [Anonymous], 2010, 20408072010 EUR PHAR
  • [5] Rapid Limit Tests for Metal Impurities in Pharmaceutical Materials by X-ray Fluorescence Spectroscopy Using Wavelet Transform Filtering
    Arzhantsev, Sergey
    Li, Xiang
    Kauffman, John F.
    [J]. ANALYTICAL CHEMISTRY, 2011, 83 (03) : 1061 - 1068
  • [6] Total reflection X-ray fluorescence trace mercury determination by trapping complexation:: Application in advanced oxidation technologies
    Custo, Graciela
    Litter, Marta I.
    Rodriguez, Diana
    Vazquez, Cristina
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (10-11) : 1119 - 1123
  • [7] Davis D, 2017, SPECTROSCOPY-US, V32, P12
  • [8] Feasibility of wavelength dispersive X-ray fluorescence spectrometry for the determination of metal impurities in pharmaceutical products and dietary supplements in view of regulatory guidelines
    Figueiredo, Alexandra
    Fernandes, Tania
    Costa, Isabel Margarida
    Goncalves, Luisa
    Brito, Jose
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 122 : 52 - 58
  • [9] The comparative verification of calibration curve and background fundamental parameter methods for impurity analysis in drug materials
    Furukawa, Hiroaki
    Ichimaru, Naoto
    Suzuki, Keijiro
    Nishino, Makoto
    Broughton, Jennifer
    Leinders, Johan
    Ochi, Hirotomo
    [J]. X-RAY SPECTROMETRY, 2017, 46 (05) : 382 - 387
  • [10] Haschke M., 2017, RONTGENFLUORESZENZAN