Effects of dispersive electrode position and surface area on electrical parameters and temperature during radiofrequency catheter ablation

被引:34
作者
Nath, S
DiMarco, JP
Gallop, RG
McRury, ID
Haines, DE
机构
[1] Cardiovascular Division, Department of Internal Medicine, Univ. of Virginia School of Medicine, Charlottesville, VA
[2] Cardiovascular Division, Department of Internal Medicine, Univ. of Virginia Hlth. Sci. Center, Charlottesville
关键词
D O I
10.1016/S0002-9149(97)89215-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Positioning of the dispersive electrode has no significant effect during radiofrequency ablation, Doubling the surface area of the dispersive electrode results in a lower impedance, higher current delivery, and increased tip temperatures, particularly if the baseline impedance is > 100 Omega. These findings may have important implications for optimizing radiofrequency energy delivery using currently available radiofrequency generators.
引用
收藏
页码:765 / &
页数:4
相关论文
共 3 条
  • [1] HINDRICKS G, 1995, RADIOFREQUENCY CATHE, P97
  • [2] CELLULAR ELECTROPHYSIOLOGICAL EFFECTS OF HYPERTHERMIA ON ISOLATED GUINEA-PIG PAPILLARY-MUSCLE - IMPLICATIONS FOR CATHETER ABLATION
    NATH, S
    LYNCH, C
    WHAYNE, JG
    HAINES, DE
    [J]. CIRCULATION, 1993, 88 (04) : 1826 - 1831
  • [3] EFFECTS OF HEATING ON IMPULSE PROPAGATION IN SUPERFUSED CANINE MYOCARDIUM
    SIMMERS, TA
    DEBAKKER, JMT
    WITTKAMPF, FHM
    HAUER, RNW
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (06) : 1457 - 1464