Novel wearable and contactless heart rate, respiratory rate, and oxygen saturation monitoring devices: a systematic review and meta-analysis

被引:18
作者
Chan, P. Y. [1 ]
Ryan, N. P. [1 ]
Chen, D. [1 ]
McNeil, J. [2 ]
Hopper, I [2 ]
机构
[1] Eastern Hlth, Dept Intens Care Med, Melbourne, Vic, Australia
[2] Monash Univ, Sch Publ Hlth & Prevent Med, Melbourne, Vic, Australia
基金
英国医学研究理事会;
关键词
clinical monitoring; contactless devices; infrared; wearable devices; EMERGENCY-DEPARTMENT; ATRIAL-FIBRILLATION; ACOUSTIC METHOD; ACCURACY; VALIDATION; PHOTOPLETHYSMOGRAPHY; CAPNOGRAPHY; ANESTHESIA; AGREEMENT; PRESSURE;
D O I
10.1111/anae.15834
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We performed a systematic review and meta-analysis to identify, classify and evaluate the body of evidence on novel wearable and contactless devices that measure heart rate, respiratory rate and oxygen saturations in the clinical setting. We included any studies of hospital inpatients, including sleep study clinics. Eighty-four studies were included in the final review. There were 56 studies of wearable devices and 29 of contactless devices. One study assessed both types of device. A high risk of patient selection and rater bias was present in proportionally more studies assessing contactless devices compared with studies assessing wearable devices (p = 0.023 and p < 0.0001, respectively). There was high but equivalent likelihood of blinding bias between the two types of studies (p = 0.076). Wearable device studies were commercially available devices validated in acute clinical settings by clinical staff and had more real-time data analysis (p = 0.04). Contactless devices were more experimental, and data were analysed post-hoc. Pooled estimates of mean (95%CI) heart rate and respiratory rate bias in wearable devices were 1.25 (-0.31-2.82) beats.min(-1) (pooled 95% limits of agreement -9.36-10.08) and 0.68 (0.05-1.32) breaths.min(-1) (pooled 95% limits of agreement -5.65-6.85). The pooled estimate for mean (95%CI) heart rate and respiratory rate bias in contactless devices was 2.18 (3.31-7.66) beats.min(-1) (pooled limits of agreement -6.71-10.88) and 0.30 (-0.26-0.87) breaths.min(-1) (pooled 95% limits of agreement -3.94-4.29). Only two studies of wearable devices measured SpO2; these reported mean measurement biases of 3.54% (limits of agreement -5.65-11.45%) and 2.9% (-7.4-1.7%). Heterogeneity was observed across studies, but absent when devices were grouped by measurement modality and reference standard. We conclude that, while studies of wearable devices were of slightly better quality than contactless devices, in general all studies of novel devices were of low quality, with small (< 100) patient datasets, typically not blinded and often using inappropriate statistical techniques. Both types of devices were statistically equivalent in accuracy and precision, but wearable devices demonstrated less measurement bias and more precision at extreme vital signs. The statistical variability in precision and accuracy between studies is partially explained by differences in reference standards.
引用
收藏
页码:1268 / 1280
页数:13
相关论文
共 50 条
[41]   Impact of wearable physical activity monitoring devices with exercise prescription or advice in the maintenance phase of cardiac rehabilitation: systematic review and meta-analysis [J].
Amanda L. Hannan ;
Michael P. Harders ;
Wayne Hing ;
Mike Climstein ;
Jeff S. Coombes ;
James Furness .
BMC Sports Science, Medicine and Rehabilitation, 11
[42]   Diagnostic accuracy of smart gadgets/wearable devices in detecting atrial fibrillation: A systematic review and meta-analysis [J].
Prasitlumkum, Narut ;
Cheungpasitporn, Wisit ;
Chokesuwattanaskul, Anthipa ;
Thangjui, Sittinun ;
Thongprayoon, Charat ;
Bathini, Tarun ;
Vallabhajosyula, Saraschandra ;
Kanitsoraphan, Chanavuth ;
Leesutipornchai, Thiratest ;
Chokesuwattanaskul, Ronpichai .
ARCHIVES OF CARDIOVASCULAR DISEASES, 2021, 114 (01) :4-16
[43]   Q-PPG: Energy-Efficient PPG-Based Heart Rate Monitoring on Wearable Devices [J].
Burrello, Alessio ;
Pagliari, Daniele Jahier ;
Risso, Matteo ;
Benatti, Simone ;
Macii, Enrico ;
Benini, Luca ;
Poncino, Massimo .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2021, 15 (06) :1196-1209
[44]   Adequacy of basal metabolic rate prediction equations in individuals with severe obesity: A systematic review and meta-analysis [J].
Campos, Tatiana Almeida de Moraes ;
Mariz, Virginia Gaissionok ;
Mulder, Alessandra Pinheiro ;
Curioni, Cintia Chaves ;
Bezerra, Flavia Fioruci .
OBESITY REVIEWS, 2024, 25 (06)
[45]   Uncertainties in the Analysis of Heart Rate Variability: A Systematic Review [J].
Lu, Lei ;
Zhu, Tingting ;
Morelli, Davide ;
Creagh, Andrew ;
Liu, Zhangdaihong ;
Yang, Jenny ;
Liu, Fenglin ;
Zhang, Yuan-Ting ;
Clifton, David A. .
IEEE REVIEWS IN BIOMEDICAL ENGINEERING, 2024, 17 :180-196
[46]   Changes in Heart Rate Variability Parameters Following Radiofrequency Ablation in Patients with Atrial Fibrillation: A Systematic Review and Meta-Analysis [J].
Mansourian, Mahsa ;
Teimouri-jervekani, Zahra ;
Soleimani, Azam ;
Nouri, Rasool ;
Marateb, Hamidreza ;
Mansourian, Marjan .
CARDIOVASCULAR DRUGS AND THERAPY, 2024,
[47]   Prognostic value of heart rate variability in atrial fibrillation recurrence following catheter ablation: A systematic review and meta-analysis [J].
Zhang, Enyuan ;
Liang, Shuo ;
Sun, Tianhong ;
Xu, Jing ;
Lu, Fengmin ;
Wu, Dongyan ;
Zhang, Jingkun ;
He, Le ;
Zhang, Fan ;
Fan, Shaobo ;
Ma, Wei .
FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 9
[48]   Wearable Devices Suitable for Monitoring Twenty Four Hour Heart Rate Variability in Military Populations [J].
Hinde, Katrina ;
White, Graham ;
Armstrong, Nicola .
SENSORS, 2021, 21 (04) :1-20
[49]   Review Fast and accurate newborn heart rate monitoring at birth: A systematic review* [J].
Kapadia, Vishal S. ;
Kawakami, Mandira D. ;
Strand, Marya L. ;
Hurst, Cameron Paul ;
Spencer, Angela ;
Schmolzer, Georg M. ;
Rabi, Yacov ;
Wyllie, Jonathan ;
Weiner, Gary ;
Liley, Helen G. ;
Wyckoff, Myra H. .
RESUSCITATION PLUS, 2024, 19
[50]   Accuracy of Resting Metabolic Rate Prediction Equations in Athletes: A Systematic Review with Meta-analysis [J].
O'Neill, Jack Eoin Rua ;
Corish, Clare A. ;
Horner, Katy .
SPORTS MEDICINE, 2023, 53 (12) :2373-2398