Flow Field Study Comparing Design Iterations of a 50 cc Left Ventricular Assist Device

被引:10
作者
Nanna, Jason C. [1 ]
Wivholm, Jennifer A. [1 ]
Deutsch, Steven [1 ]
Manning, Keefe B. [1 ]
机构
[1] Penn State Univ, Dept Bioengn, University Pk, PA 16802 USA
关键词
MECHANICAL CIRCULATORY SUPPORT; STATE ARTIFICIAL-HEART; PULSATILE;
D O I
10.1097/MAT.0b013e318224e20b
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Randomized Evaluation of Mechanical Assistance for the Treatment of Congestive Heart Failure (REMATCH) study shows that implanted ventricular assist devices improve survival time and quality of life when used as a permanent therapy in patients who do not qualify for heart transplant. The success of the pulsatile 70 cc stroke volume left ventricular assist device (LVAD) developed by Penn State has led to the development of a 50 cc stroke volume pump for use in patients with smaller chest cavities to benefit a larger patient population. The initial 50 cc pump shows regions of in vivo thrombus formation, which correlate to low wall shear rates within the device. In an in vitro evaluation of three new designs (V-2, V-3, and V-4) of the 50 cc LVAD, identical except for the location and orientation of their outlet ports, particle image velocimetry (PIV) is used to capture planar flow field data within the pumps. V-2 has an outlet port that is located parallel to the inlet. In V-3, the outlet port is rotated away from the inlet port, with the intention of minimizing the amount of fluid turning needed to exit the device. With V-4 the outlet port is moved to the center of the pump to prolong the desirable rotational flow. PIV data were taken at six planar locations within the pump. Although the modifications to the outlet port locations serve their intended purpose, they also introduce unwanted changes in the flow. Poorer wall washing and weaker rotational flow are observed with V-3 and V-4. Although the differences between the devices are subtle, the device that has the most desirable flow characteristics is V-2. ASAIO Journal 2011; 57:349-357.
引用
收藏
页码:349 / 357
页数:9
相关论文
共 23 条
[1]   Peripheral vascular reactivity in patients with pulsatile vs axial flow left ventricular assist device support [J].
Amir, O ;
Radovancevic, B ;
Delgado, RM ;
Kar, B ;
Radovancevic, R ;
Henderson, M ;
Cohn, WE ;
Smart, FW .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (04) :391-394
[2]  
[Anonymous], 2009, UNOS ANN REP US SCI
[3]   Differences in pulsatile and non-pulsatile mechanical circulatory support in long-term use [J].
Drews, Thorsten ;
Jurmann, Michael ;
Michael, Dandel ;
Miralem, Pasic ;
Weng, Yuguo ;
Hetzer, Roland .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (10) :1096-1101
[4]   Mechanical circulatory support in patients of advanced age [J].
Drews, Thorsten ;
Stepanenko, Alexander ;
Dandel, Michael ;
Buz, Semih ;
Lehmkuhl, Hans Brendan ;
Hetzer, Roland .
EUROPEAN JOURNAL OF HEART FAILURE, 2010, 12 (09) :990-994
[5]   Correlation of in vivo clot deposition with the flow characteristics in the 50 cc Penn State Artificial Heart:: A preliminary study [J].
Hochareon, P ;
Manning, KB ;
Fontaine, AA ;
Tarbell, JM ;
Deutsch, S .
ASAIO JOURNAL, 2004, 50 (06) :537-542
[6]   Fluid dynamic analysis of the 50 cc Penn State artificial heart under physiological operating conditions using particle image velocimetry [J].
Hochareon, P ;
Manning, KB ;
Fontaine, AA ;
Tarbell, JM ;
Deutsch, S .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (05) :585-593
[7]   Wall shear-rate estimation within the 50cc Penn State artificial heart using particle image velocimetry [J].
Hochareon, P ;
Manning, KB ;
Fontaine, AA ;
Tarbell, JM ;
Deutsch, S .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (04) :430-437
[8]   Diaphragm motion affects flow patterns in an artificial heart [J].
Hochareon, P ;
Manning, KB ;
Fontaine, AA ;
Deutsch, S ;
Tarbell, JM .
ARTIFICIAL ORGANS, 2003, 27 (12) :1102-1109
[9]   VISUALIZATION AND ANALYSIS OF MURAL THROMBOGENESIS ON COLLAGEN, POLYURETHANE AND NYLON [J].
HUBBELL, JA ;
MCINTIRE, LV .
BIOMATERIALS, 1986, 7 (05) :354-&
[10]   Single-centre experience with the Thoratec® paracorporeal ventricular assist device for patients with primary cardiac failure [J].
Kirsch, Matthias ;
Vermes, Emmanuelle ;
Damy, Thibaud ;
Nakashima, Kuniki ;
Senechal, Melanie ;
Boval, Bernadette ;
Drouet, Ludovic ;
Loisance, Daniel .
ARCHIVES OF CARDIOVASCULAR DISEASES, 2009, 102 (6-7) :509-518