Screw insertion torque as parameter to judge the fixation. Assessment of torque and pull-out strength in different bone densities and screw-pitches

被引:30
作者
Addevico, Francesco [1 ]
Morandi , Massimo [2 ]
Scaglione, Michelangelo [1 ]
Solitro, Giovanni F. [3 ]
机构
[1] Pisa Univ Hosp, Orthopead Clin, Pisa, Italy
[2] Louisiana State Univ Hlth, Sci Ctr Shreveport, Dept Orthopaed Surg, Div Orthopaed Trauma, Shreveport, LA USA
[3] Louisiana State Univ Hlth, Sci Ctr Shreveport, Dept Orthopaed Surg, Biomech Lab, Shreveport, LA USA
关键词
INTERFRAGMENTARY COMPRESSION; SYNTHETIC FOAM; CORTICAL BONE; AUGMENTATION; BIOMECHANICS; MODELS; PLATE;
D O I
10.1016/j.clinbiomech.2019.12.004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Pull-out strength is a critical parameter to judge screw fixation in orthopaedic implants. However, the insertion torque is the main feeling in the hand of a surgeon relating to the strength of synthesis. The correlation between pull-out strength and torque is not completely understood. This creates uncertainty about the key-question: Should the torque be considered a valid parameter to judge the quality of fixation? Methods: Using the ASTM F543 as reference, three screws differing only in pitch (1.5, 2.1, 2.8 mm pitches) were tested in three foam-block densities (10, 15, and 20 pcf). The correlation was investigated by assessing the role of density and screw geometry. Findings: Torque was related to pull-out strength in all configurations (R = 0.979, P = 0.000). No difference in pull-out strength was found when screws were tightened to a range of 71.6%, SD = 7.6, of torque to fail (P > 0.05). Torque and pull-out strength were stratified according to density that influenced the two parameters up to 524% (P < 0.000). Pitch determined pull-out strength up to 33% (P < 0.000) while the 2.1 mm screw pitch showed the highest pull-out strength and torque in all configurations. Interpretation: Insertion torque was demonstrated to be a valid parameter to judge the quality of bone under fixation and therefore, the strength of the synthesis. Surgeons should not tighten the screws to values approaching torque to fail to obtain the highest pull-out strength. Density was the main factor influencing pull-out strength and torque. Pitch is another parameter deciding screw holding capacity.
引用
收藏
页码:130 / 135
页数:6
相关论文
共 33 条
[1]   Stability and Spine Pedicle Screws Fixation Strength - A Comparative Study of Bone Density and Insertion Angle [J].
Amirouche F. ;
Solitro G.F. ;
Magnan B.P. .
Spine Deformity, 2016, 4 (4) :261-267
[2]   Use of a synthetic bone void filler to augment screws in osteopenic ankle fracture fixation [J].
Andreassen, GS ;
Hoiness, PR ;
Skraamm, I ;
Granlund, O ;
Engebretsen, L .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2004, 124 (03) :161-165
[3]  
ARO HT, 1993, CLIN ORTHOP RELAT R, P8
[4]   The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone [J].
Battula, Suneel ;
Schoenfeld, Andrew J. ;
Sahai, Vivek ;
Vrabec, Gregory A. ;
Tank, Jason ;
Njus, Glen O. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04) :990-995
[5]   Mechanical analysis of the bone to plate interface of the LC-DCP and of the PC-FIX on human femora [J].
Borgeaud, M ;
Cordey, J ;
Leyvraz, PF ;
Perren, SM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :29-36
[6]   Biomechanical analysis of different types of pedicle screw augmentation: A cadaveric and synthetic bone sample study of instrumented vertebral specimens [J].
Chao, Kuo-Hua ;
Lai, Yu-Shu ;
Chen, Wen-Chuan ;
Chang, Chia-Ming ;
McClean, Colin J. ;
Fan, Chang-Yuan ;
Chang, Chia-Hao ;
Lin, Leou-Chyr ;
Cheng, Cheng-Kung .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (10) :1506-1512
[7]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[8]   Effect of screw torque level on cortical bone pullout strength [J].
Cleek, Tammy M. ;
Reynolds, Karen J. ;
Hearn, Trevor C. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (02) :117-123
[9]   Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and bone [J].
Cordey, J ;
Borgeaud, M ;
Perren, SM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :21-28
[10]   Inadvertent Screw Stripping During Ankle Fracture Fixation in Elderly Bone [J].
Dinah, A. Feroz ;
Mears, Simon C. ;
Knight, Trevor A. ;
Soin, Sandeep P. ;
Campbell, John T. ;
Belkoff, Stephen M. .
GERIATRIC ORTHOPAEDIC SURGERY & REHABILITATION, 2011, 2 (03) :86-89