Novel fast-arrayed CCD X-ray detector using interline transfer-type CCD for time-resolved X-ray diffraction measurements

被引:3
作者
Ito, Kazuki
Fujisawa, Tetsuro
Iwata, Tadahisa
机构
[1] RIKEN, Harima Inst, SPring 8 Ctr, Sayo, Hyogo 6795148, Japan
[2] Gifu Univ, Grad Sch Engn, Dept Biomol Sci, Gifu 5011193, Japan
[3] RIKEN, Wako Inst, Bioengn Lab & Ecomol Sci Res 2, Wako, Saitama 3510198, Japan
[4] Univ Tokyo, Grad Sch Agr & Life Sci, Bunkyo Ku, Tokyo 1138657, Japan
关键词
fiber-optic taper (FOT); charge-coupled device (CCD); X-ray detector; calibration; time-resolved measurement; wide-angle X-ray diffraction (WAXD);
D O I
10.1016/j.nima.2007.08.173
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel fast-arrayed fiber-optic taper (FOT) coupled with a charge-coupled device (CCD) X-ray detector was developed. The interline transfer-type CCDs are used as image sensor. The detector was developed for the simultaneous time-resolved small- and wide-angle X-ray diffraction measurement at SPring-8 BL45XU. The detector consists of an array of two identical units, each comprising a phosphor, a demagnifying FOT, and a large format interline transfer-type CCD as an image sensor. The image stored in the CCD can be read out through the dual readout channels with 12 bits ADC within 220 ms (4.5 fps) without binning and within 65 ms (15 fps) with 8 x 8 binning. Moreover, this detector does not need to stop the incoming X-rays during the CCD readout and can simultaneously execute both exposure and readout on the CCD with a small dead time of similar to 1 mu s so that a duty-cycle ratio of almost 100% is achieved. Therefore, it allows the continuous rotation method. The basic performance characteristics of the detector are described. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:673 / 682
页数:10
相关论文
共 14 条
[1]  
ADACHI S, 2001, NUCL INSTRUM METH A, V10, P1209
[2]   DEVELOPMENT OF NON-INTENSIFIED CHARGE-COUPLED-DEVICE AREA X-RAY-DETECTORS [J].
ALLINSON, NM .
JOURNAL OF SYNCHROTRON RADIATION, 1994, 1 :54-62
[3]   HIGH-SPATIAL-RESOLUTION SCINTILLATOR SCREENS COUPLED TO CCD DETECTORS FOR X-RAY-IMAGING APPLICATIONS [J].
CASTELLI, CM ;
ALLINSON, NM ;
MOON, KJ ;
WATSON, DL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3) :649-653
[4]   Evaluation of imaging performance of a taper optics CCD 'FReLoN' camera designed for medical imaging [J].
Coan, P ;
Peterzol, A ;
Fiedler, S ;
Ponchut, C ;
Labiche, JC ;
Bravin, A .
JOURNAL OF SYNCHROTRON RADIATION, 2006, 13 :260-270
[5]  
DOI Y, 1990, MICROBIOL POLYESTERS
[6]   Small-angle X-ray scattering station at the SPring-8 RIKEN beamline [J].
Fujisawa, T ;
Inoue, K ;
Oka, T ;
Iwamoto, H ;
Uruga, T ;
Kumasaka, T ;
Inoko, Y ;
Yagi, N ;
Yamamoto, M ;
Ueki, T .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2000, 33 (01) :797-800
[7]   Charge-coupled device area x-ray detectors [J].
Gruner, SM ;
Tate, MW ;
Eikenberry, EF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (08) :2815-2842
[8]   Correction method and software for image distortion and nonuniform response in charge-coupled device-based X-ray detectors utilizing X-ray image intensifier [J].
Ito, K ;
Kamikubo, H ;
Yagi, N ;
Amemiya, Y .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2005, 44 (12) :8684-8691
[9]  
Ito K, 2007, AIP CONF PROC, V879, P1160
[10]   A 3 x 6 arrayed CCD X-ray detector for continuous rotation method in macromolecular crystallography [J].
Ito, Kazuki ;
Gaponov, Yurii ;
Sakabe, Noriyoshi ;
Amemiya, Yoshiyuki .
JOURNAL OF SYNCHROTRON RADIATION, 2007, 14 :144-150