Use of Nonmydriatic Spectral-Domain Optical Coherence Tomography for Diagnosing Diabetic Macular Edema

被引:21
作者
Lara Medina, Francisco Javier [1 ]
Ispa Callen, Carmen [1 ]
Rebolleda, Gema [2 ]
Munoz-Negrete, Francisco J. [2 ]
Ispa Callen, Maria J. [3 ]
Gonzalez Del Valle, Fernando [1 ]
机构
[1] Hosp Gen Mancha La Ctr, Dept Ophthalmol, Retinal Unit, E-13600 Ciudad Real, Spain
[2] Hosp Ramon & Cajal, Dept Ophthalmol, Madrid, Spain
[3] Univ Zaragoza, Dept Stat, Aragon Govt, Zaragoza, Spain
关键词
RETINAL THICKNESS; PHOTOGRAPHY; INSTRUMENTS; RETINOPATHY; OCT;
D O I
10.1016/j.ajo.2011.08.008
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the accuracy of 3 spectral-domain (SD) optical coherence tomography (OCT) devices (Topcon 3D-1000 [Topcon]; Cirrus HD [Carl Zeiss Meditec, Inc], and Spectralis OCT [Heidelberg Engineering]) before and after mydriasis for the diagnosis of diabetic macular edema. DESIGN: Cross-sectional study. METHODS: Sixty-two eyes of 62 consecutive patients with diabetes without recent loss of vision referred for retinal control were assessed. Two scans were performed for each SD OCT instrument. Central retinal thickness was measured before and after pupil dilation. Pupil dynamic was studied using pMetrics pupillometer (iVIS Technologies), and lens opacity was measured by Pentacam densitometry (Oculus). The diagnostic accuracy of SD OCT devices was assessed by sensitivity, specificity, and area under the receiver operating characteristic curve. Logistic regression analysis was used to assess the effect of pupil size and lens opacity on the reliability of SD OCT in the acquisition of adequate images. RESULTS: The area under the receiver operating characteristic curve for the Topcon 3D OCT device was 0.84, that for the Cirrus HD OCT device was 0.93, and that for the Spectralis OCT device was 0.91. Significant differences in area under the receiver operating characteristic curve before and after pupillary dilatation were not found. Sensitivity and specificity associated with the cutoff value for the best performance were 82% and 74% for the Topcon 3D OCT device, 90% and 87% for the Cirrus HD OCT device, and 90 h and 84% for the Spectralis OCT device, respectively. The Topcon 3D OCT device had an 11.3% segmentation algorithm failure rate for the central millimeter of the fovea, and the nuclear lens density was significantly greater in these eyes than in those without failure (17.1 +/- 1.1 mm vs 10.4 +/- 0.2 mm; P < .05). CONCLUSIONS: SD OCT is a useful tool to detect and to measure diabetic macular edema without the need for pupil dilatation. (Am J Ophthalmol 2012;153:536-543. (C) 2012 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:536 / 543
页数:8
相关论文
共 19 条
[1]  
BRESNICK GH, 1986, OPHTHALMOLOGY, V93, P989
[2]   Comparison of the clinical diagnosis of diabetic macular edema with diagnosis by optical coherence tomography [J].
Browning, DJ ;
McOwen, MD ;
Bowen, RM ;
O'Marah, TL .
OPHTHALMOLOGY, 2004, 111 (04) :712-715
[3]   Heartbeat-Induced Axial Motion Artifacts in Optical Coherence Tomography Measurements of the Retina [J].
de Kinkelder, Roy ;
Kalkman, Jeroen ;
Faber, Dirk J. ;
Schraa, Olaf ;
Kok, Pauline H. B. ;
Verbraak, Frank D. ;
van Leeuwen, Ton G. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (06) :3908-3913
[4]   Reproducibility of Retinal Nerve Fiber Layer and Macular Thickness Measurement with the RTVue-100 Optical Coherence Tomograph [J].
Garas, Anita ;
Vargha, Peter ;
Hollo, Gabor .
OPHTHALMOLOGY, 2010, 117 (04) :738-746
[5]   Comparison of Optical Coherence Tomography in Diabetic Macular Edema, with and without Reading Center Manual Grading from a Clinical Trials Perspective [J].
Glassman, Adam R. ;
Beck, Roy W. ;
Browning, David J. ;
Danis, Ronald P. ;
Kollman, Craig .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (02) :560-566
[6]   Retinal thickness in diabetic retinopathy - A study using optical coherence tomography (OCT) [J].
Goebel, W ;
Kretzchmar-Gross, T .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2002, 22 (06) :759-767
[7]   QUANTITATIVE ASSESSMENT OF MACULAR EDEMA WITH OPTICAL COHERENCE TOMOGRAPHY [J].
HEE, MR ;
PULIAFITO, CA ;
WONG, C ;
DUKER, JS ;
REICHEL, E ;
RUTLEDGE, B ;
SCHUMAN, JS ;
SWANSON, EA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (08) :1019-1029
[8]   OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA [J].
HEE, MR ;
IZATT, JA ;
SWANSON, EA ;
HUANG, D ;
SCHUMAN, JS ;
LIN, CP ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) :325-332
[9]  
KINYOUN J, 1989, OPHTHALMOLOGY, V96, P746
[10]  
Koozekanani D, 2000, INVEST OPHTH VIS SCI, V41, P1486