Use of Nonmydriatic Spectral-Domain Optical Coherence Tomography for Diagnosing Diabetic Macular Edema

被引:21
|
作者
Lara Medina, Francisco Javier [1 ]
Ispa Callen, Carmen [1 ]
Rebolleda, Gema [2 ]
Munoz-Negrete, Francisco J. [2 ]
Ispa Callen, Maria J. [3 ]
Gonzalez Del Valle, Fernando [1 ]
机构
[1] Hosp Gen Mancha La Ctr, Dept Ophthalmol, Retinal Unit, E-13600 Ciudad Real, Spain
[2] Hosp Ramon & Cajal, Dept Ophthalmol, Madrid, Spain
[3] Univ Zaragoza, Dept Stat, Aragon Govt, Zaragoza, Spain
关键词
RETINAL THICKNESS; PHOTOGRAPHY; INSTRUMENTS; RETINOPATHY; OCT;
D O I
10.1016/j.ajo.2011.08.008
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the accuracy of 3 spectral-domain (SD) optical coherence tomography (OCT) devices (Topcon 3D-1000 [Topcon]; Cirrus HD [Carl Zeiss Meditec, Inc], and Spectralis OCT [Heidelberg Engineering]) before and after mydriasis for the diagnosis of diabetic macular edema. DESIGN: Cross-sectional study. METHODS: Sixty-two eyes of 62 consecutive patients with diabetes without recent loss of vision referred for retinal control were assessed. Two scans were performed for each SD OCT instrument. Central retinal thickness was measured before and after pupil dilation. Pupil dynamic was studied using pMetrics pupillometer (iVIS Technologies), and lens opacity was measured by Pentacam densitometry (Oculus). The diagnostic accuracy of SD OCT devices was assessed by sensitivity, specificity, and area under the receiver operating characteristic curve. Logistic regression analysis was used to assess the effect of pupil size and lens opacity on the reliability of SD OCT in the acquisition of adequate images. RESULTS: The area under the receiver operating characteristic curve for the Topcon 3D OCT device was 0.84, that for the Cirrus HD OCT device was 0.93, and that for the Spectralis OCT device was 0.91. Significant differences in area under the receiver operating characteristic curve before and after pupillary dilatation were not found. Sensitivity and specificity associated with the cutoff value for the best performance were 82% and 74% for the Topcon 3D OCT device, 90% and 87% for the Cirrus HD OCT device, and 90 h and 84% for the Spectralis OCT device, respectively. The Topcon 3D OCT device had an 11.3% segmentation algorithm failure rate for the central millimeter of the fovea, and the nuclear lens density was significantly greater in these eyes than in those without failure (17.1 +/- 1.1 mm vs 10.4 +/- 0.2 mm; P < .05). CONCLUSIONS: SD OCT is a useful tool to detect and to measure diabetic macular edema without the need for pupil dilatation. (Am J Ophthalmol 2012;153:536-543. (C) 2012 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:536 / 543
页数:8
相关论文
共 50 条
  • [1] Assessment of serous macular detachment in eyes with diabetic macular edema by use of spectral-domain optical coherence tomography
    Koleva-Georgieva, Desislava
    Sivkova, Nelly
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 247 (11) : 1461 - 1469
  • [2] Comparison of Time- and Spectral-Domain Optical Coherence Tomography in Management of Diabetic Macular Edema
    Liu, Melissa M.
    Wolfson, Yulia
    Bressler, Susan B.
    Do, Diana V.
    Ying, Howard S.
    Bressler, Neil M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (03) : 1370 - 1377
  • [3] Interobserver Agreement in Detecting Spectral-Domain Optical Coherence Tomography Features of Diabetic Macular Edema
    Heng, Ling Zhi
    Pefianaki, Maria
    Hykin, Philip
    Patel, Praveen J.
    PLOS ONE, 2015, 10 (05):
  • [4] Spectral-domain Optical Coherence Tomography Measurement of Macular Volume for Diagnosing Glaucoma
    Mori, Satoshi
    Hangai, Masanori
    Sakamoto, Atsushi
    Yoshimura, Nagahisa
    JOURNAL OF GLAUCOMA, 2010, 19 (08) : 528 - 534
  • [5] Segmentational Diurnal Analysis of Retinal Thickness in Diabetic Macular Edema with Spectral-domain Optical Coherence Tomography
    Sagong, Min
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [6] En Face Spectral-Domain Optical Coherence Tomography Imaging of Outer Retinal Hard Exudates in Diabetic Macular Edema Based on Optical Coherence Tomography Patterns
    Kim, Yonguk
    Yu, Seung-Young
    Kwak, Hyung Woo
    OPHTHALMIC SURGERY LASERS & IMAGING RETINA, 2016, 47 (04) : 313 - 321
  • [7] Optical coherence tomography in diabetic macular edema
    Baskin, Darrell E.
    CURRENT OPINION IN OPHTHALMOLOGY, 2010, 21 (03) : 172 - 177
  • [8] Comparing central retinal thickness in diabetic macular edema measured by two different spectral-domain optical coherence tomography devices
    Suzuma, Kiyoshi
    Yamada, Yoshihisa
    Liu, Michi
    Tsuiki, Eiko
    Fujikawa, Azusa
    Kitaoka, Takashi
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2011, 55 (06) : 620 - 624
  • [9] Fundus autofluorescence and spectral domain optical coherence tomography in uveitic macular edema
    Roesel, Martin
    Henschel, Andreas
    Heinz, Carsten
    Dietzel, Martha
    Spital, Georg
    Heiligenhaus, Arnd
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 247 (12) : 1685 - 1689
  • [10] Reproducibility of Spectral-Domain Optical Coherence Tomography Retinal Thickness Measurements and Conversion to Equivalent Time-Domain Metrics in Diabetic Macular Edema
    Bressler, Susan B.
    Edwards, Allison R.
    Chalam, Kakarla V.
    Bressler, Neil M.
    Glassman, Adam R.
    Jaffe, Glenn J.
    Melia, Michele
    Saggau, David D.
    Plous, Oren Z.
    JAMA OPHTHALMOLOGY, 2014, 132 (09) : 1113 - 1122