A Vector Flow Imaging Method for Portable Ultrasound Using Synthetic Aperture Sequential Beamforming

被引:16
|
作者
Di Ianni, Tommaso [1 ]
Hoyos, Carlos Armando Villagomez [1 ]
Ewertsen, Caroline [2 ]
Kjeldsen, Thomas Kim [3 ]
Mosegaard, Jesper [3 ]
Nielsen, Michael Bachmann [2 ]
Jensen, Jorgen Arendt [1 ]
机构
[1] Tech Univ Denmark, Ctr Fast Ultrasound Imaging, Dept Elect Engn, DK-2800 Lyngby, Denmark
[2] Copenhagen Univ Hosp, Rigshosp, Dept Radiol, DK-2100 Copenhagen, Denmark
[3] Alexandra Inst, Visual Comp Lab, DK-8200 Aarhus, Denmark
关键词
Portable ultrasound; synthetic aperture (SA); transverse oscillation (TO); vector flow imaging (VFI); VELOCITY ESTIMATION; BLOOD-FLOW; TRANSVERSE OSCILLATION; DOPPLER; AUTOCORRELATION; VELOCIMETRY; SYSTEM;
D O I
10.1109/TUFFC.2017.2742599
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper presents a vector flow imaging method for the integration of quantitative blood flow imaging in portable ultrasound systems. The method combines directional transverse oscillation (TO) and synthetic aperture sequential beamforming to yield continuous velocity estimation in the whole imaging region. Six focused emissions are used to create a high-resolution image (HRI), and a dual-stage beamforming approach is used to lower the data throughput between the probe and the processing unit. The transmit/receive focal points are laterally separated to obtain a TO in the HRI that allows for the velocity estimation along the lateral and axial directions using a phase-shift estimator. The performance of the method was investigated with constant flow measurements in a flow rig system using the SARUS scanner and a 4.1-MHz linear array. A sequence was designed with interleaved B-mode and flow emissions to obtain continuous data acquisition. A parametric study was carried out to evaluate the effect of critical parameters. The vessel was placed at depths from 20 to 40 mm, with beam-to-flow angles of 65 degrees, 75 degrees, and 90 degrees. For the lateral velocities at 20 mm, a bias between -5% and -6.2% was obtained, and the standard deviation (SD) was between 6% and 9.6%. The axial bias was lower than 1% with an SD around 2%. The mean estimated angles were 66.70 degrees +/- 2.86 degrees, 72.65 degrees +/- 2.48 degrees, and 89.13 degrees +/- 0.79 degrees for the three cases. A proof-of-concept demonstration of the real-time processing and wireless transmission was tested in a commercial tablet obtaining a frame rate of 27 frames/s and a data rate of 14 MB/s. An in vivo measurement of a common carotid artery of a healthy volunteer was finally performed to show the potential of the method in a realistic setting. The relative SD averaged over a cardiac cycle was 4.33%.
引用
收藏
页码:1655 / 1665
页数:11
相关论文
共 50 条
  • [1] Vector Velocity Estimation for Portable Ultrasound using Directional Transverse Oscillation and Synthetic Aperture Sequential Beamforming
    Di Ianni, Tommaso
    Hemmsen, Martin Christian
    Jensen, Jorgen Arendt
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [2] Sequential beamforming for synthetic aperture imaging
    Kortbek, Jacob
    Jensen, Jorgen Arendt
    Gammelmark, Kim Lokke
    ULTRASONICS, 2013, 53 (01) : 1 - 16
  • [3] Motion compensated beamforming in synthetic aperture vector flow imaging
    Oddershede, Niels
    Jensen, Jorgen Arendt
    2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 2027 - 2031
  • [4] Second harmonic imaging using synthetic aperture sequential beamforming
    Du, Yigang
    Rasmussen, Joachim
    Jensen, Henrik
    Jensen, Jorgen Arendt
    2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 2261 - 2264
  • [5] Low Complexity 3D Ultrasound Imaging Using Synthetic Aperture Sequential Beamforming
    Zhou, Jian
    Wei, Siyuan
    Sampson, Richard
    Yang, Ming
    Jintamethasawat, Rungroj
    Kripfgans, Oliver D.
    Fowlkes, J. Brian
    Wenisch, Thomas F.
    Chakrabarti, Chaitali
    2016 IEEE INTERNATIONAL WORKSHOP ON SIGNAL PROCESSING SYSTEMS (SIPS), 2016, : 33 - 38
  • [6] PORTABLE PARALLEL KERNELS FOR HIGH-SPEED BEAMFORMING IN SYNTHETIC APERTURE ULTRASOUND IMAGING
    Amaro, Joao
    Falcao, Gabriel
    Yiu, Billy Y. S.
    Yu, Alfred C. H.
    2013 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2013, : 2688 - 2692
  • [7] Transverse flow imaging using synthetic aperture directional beamforming
    Jensen, JA
    Nikolov, SI
    2002 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2002, : 1523 - 1527
  • [8] Synthetic Aperture Sequential Beamforming for Phased Array Imaging
    Bera, Deep
    Bosch, Johan G.
    de Jong, Nico
    Vos, Hendrik J.
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [9] Synthetic Aperture Sequential Beamforming
    Kortbek, Jacob
    Jensen, Jorgen Arendt
    Gammelmark, Kim Lokke
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 966 - +
  • [10] High Volume Rate 3D Ultrasound Imaging Based on Synthetic Aperture Sequential Beamforming
    Zhou, Jian
    Wei, Siyuan
    Sampson, Richard
    Jintamethasawat, Rungroj
    Kripfgans, Oliver D.
    Fowlkes, J. Brian
    Wenisch, Thomas F.
    Chakrabarti, Chaitali
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,