Nafion coating the ferrocenylalkanethiol and encapsulated glucose oxidase electrode for amperometric glucose detection

被引:15
|
作者
Peng, Yong [1 ]
Wei, Chuan-Wan [1 ]
Liu, You-Nian [1 ]
Li, Juan [1 ]
机构
[1] Cent S Univ, Coll Chem & Chem Engn, Changsha 410083, Hunan, Peoples R China
基金
中国国家自然科学基金;
关键词
HYDROGEN-PEROXIDE; IN-VITRO; BIOSENSOR; ENZYME; COMPOSITE; SENSORS; IMMOBILIZATION; NANOPARTICLES; MONOLAYERS; MEMBRANES;
D O I
10.1039/c1an15292k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes the fabrication and application of a complex electrode - Nafion film coating ferrocenylalkanethiol (FcC(11)SH) and encapsulated glucose oxidase (GOD) on a gold electrode. FcC(11)SH is employed as a mediator enabling the electron transfer between GOD and the electrode, GOD is encapsulated in polyacrylamide gel to improve the stability of the enzyme, and the Nafion film is coated on the modified electrode to eliminate interferents such as ascorbic acid, uric acid and acetaminophen in amperometric glucose detection. It is noticed that such a complex electrode exhibits excellent catalytic activity for glucose oxidation, and preserves the native structure of GOD and therefore its enzymatic activity. The encapsulated GOD retains more than 80% of its original biocatalytic activity even after 24 days, much longer than that of naked GOD molecules attached directly to the electrode. The oxidation peak current at the modified electrode shows a linear relationship with the glucose concentration in the range from 0.05 to 20 mM with a detection limit of 2.4 mu M. In addition, the electrode displays a rapid response and good reproducibility for glucose detection, and has been successfully employed for glucose detection in blood plasma samples.
引用
收藏
页码:4003 / 4007
页数:5
相关论文
共 50 条
  • [1] Amperometric glucose biosensor based on glucose oxidase encapsulated in carbon nanotube-titania-nafion composite film on platinized glassy carbon electrode
    Choi, Han Nim
    Han, Jee Hoon
    Park, Ji Ae
    Lee, Joong Min
    Lee, Won-Yong
    ELECTROANALYSIS, 2007, 19 (17) : 1757 - 1763
  • [2] AMPEROMETRIC GLUCOSE SENSORS BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN NAFION
    RISHPON, J
    GOTTESFELD, S
    CAMPBELL, C
    DAVEY, J
    ZAWODZINSKI, TA
    ELECTROANALYSIS, 1994, 6 (01) : 17 - 21
  • [3] EVALUATION OF NAFION AS MEDIA FOR GLUCOSE-OXIDASE IMMOBILIZATION FOR THE DEVELOPMENT OF AN AMPEROMETRIC GLUCOSE BIOSENSOR
    FORTIER, G
    VAILLANCOURT, M
    BELANGER, D
    ELECTROANALYSIS, 1992, 4 (03) : 275 - 283
  • [4] AMPEROMETRIC BIOSENSOR FOR GLUCOSE BASED ON NAFION LOADED WITH REDOX COUPLES AND GLUCOSE-OXIDASE
    BELANGER, D
    VAILLANCOURT, M
    FORTIER, G
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1990, 199 : 1 - ANYL
  • [5] Amperometric detection of β-D-glucose using a glucose oxidase-modified rotating disk electrode
    Wang, Delin
    Ronkainen, Niina J.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 231
  • [6] AMPEROMETRIC PROCESSES WITH GLUCOSE-OXIDASE EMBEDDED IN THE ELECTRODE
    SAKURA, S
    BUCK, RP
    BIOELECTROCHEMISTRY AND BIOENERGETICS, 1992, 28 (03): : 387 - 400
  • [7] AMPEROMETRIC DETERMINATION OF GLUCOSE WITH A FERROCENE-MEDIATED GLUCOSE-OXIDASE ELECTRODE
    SHA, XZ
    HUO, JW
    WANG, XZ
    SENSORS AND ACTUATORS B-CHEMICAL, 1993, 12 (01) : 33 - 36
  • [8] AMPEROMETRIC ENZYME ELECTRODE WITH FAST RESPONSE TO GLUCOSE USING A LAYER OF LIPID-MODIFIED GLUCOSE-OXIDASE AND NAFION ANIONIC POLYMER
    MIZUTANI, F
    YABUKI, S
    KATSURA, T
    ANALYTICA CHIMICA ACTA, 1993, 274 (02) : 201 - 207
  • [9] Amperometric detection of glucose with glucose oxidase immobilized in layered double hydroxides
    Shan, Dan
    Yao, Wenjuan
    Xue, Huaiguo
    ELECTROANALYSIS, 2006, 18 (15) : 1485 - 1491
  • [10] An Amperometric Glucose Biosensor Composed of Prussian Blue, Nafion, and Glucose Oxidase Studied by Flow Injection Analysis
    Drew, Steven M.
    Belzer, Tristan
    JOURNAL OF CHEMICAL EDUCATION, 2023, 100 (02) : 760 - 766