An improved flow-based procedure is proposed for turbidimetric sulphate determination in waters. The flow system was designed with solenoid micro-pumps in order to improve mixing conditions and minimize reagent consumption as well as waste generation. Stable baselines were observed in view of the pulsed flow characteristic of the systems designed with solenoid micro-pumps, thus making the use of washing solutions unnecessary. The nucleation process was improved by stopping the flow prior to the measurement, thus avoiding the need of sulphate addition. When a 1-cm optical path flow cell was employed, linear response was achieved within 20-200 mg L-1, described by the equation S = -0.0767 + 0.00438C (mg L-1), r = 0.999. The detection limit was estimated as 3 mg L-1 at the 99.7% confidence level and the coefficient of variation was 2.4% (n = 20). The sampling rate was estimated as 33 determinations per hour. A long pathlength (100-cm) flow cell based on a liquid core waveguide was exploited to increase sensitivity in turbidimetry. Baseline drifts were avoided by a periodical washing step with EDTA in alkaline medium. Linear response was observed within 7-16 mg L-1, described by the equation S = -0.865 + 0.132C (mg L-1), r = 0.999. The detection limit was estimated as 150 mu g L-1 at the 99.7% confidence level and the coefficient of variation was 3.0% (n = 20). The sampling rate was estimated as 25 determinations per hour. The results obtained for freshwater and rain water samples were in agreement with those achieved by batch turbidimetry at the 95% confidence level. (C) 2008 Elsevier B.V All rights reserved.
引用
收藏
页码:56 / 62
页数:7
相关论文
共 37 条
[1]   Raman performance characteristics of Teflon®-AF 2400 liquid-core optical-fiber sample cells [J].
Altkorn, R ;
Koev, I ;
Pelletier, MJ .
APPLIED SPECTROSCOPY, 1999, 53 (10) :1169-1176
[2]  
[Anonymous], 1987, ANALYST, V112, P199
[3]  
[Anonymous], J FLOW INJECTION ANA
[4]  
[Anonymous], 1995, Standard methods for the examination of water and wastewater, V19th
[5]   FLOW-INJECTION SPECTROFLUOROMETRIC DETERMINATION OF SULFATE USING CALCEIN AND ZIRCONIUM [J].
CHIMPALEE, N ;
CHIMPALEE, D ;
SUPARUKNARI, S ;
BOONYANITCHAYAKUL, B ;
BURNS, DT .
ANALYTICA CHIMICA ACTA, 1994, 298 (03) :401-404
[6]   Use of reagent in suspension in flow injection analysis. Spectrophotometric determination of sulphate in natural water. [J].
Crnkovic, PM ;
Jacintho, AO .
QUIMICA NOVA, 2002, 25 (02) :254-258
[7]   Light at the end of the tunnel: recent analytical applications of liquid-core waveguides [J].
Dallas, T ;
Dasgupta, PK .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2004, 23 (05) :385-392
[8]   High-sensitivity gas sensors based on gas-permeable liquid core waveguides and long-path absorbance detection [J].
Dasgupta, PK ;
Genfa, Z ;
Poruthoor, SK ;
Caldwell, S ;
Dong, S ;
Liu, SY .
ANALYTICAL CHEMISTRY, 1998, 70 (22) :4661-4669
[9]   Luminescence detection with a liquid care waveguide [J].
Dasgupta, PK ;
Zhang, GF ;
Li, JZ ;
Boring, CB ;
Jambunathan, S ;
Al-Horr, R .
ANALYTICAL CHEMISTRY, 1999, 71 (07) :1400-1407
[10]   THE MINERALOGY OF CASTLEGUARD CAVE, COLUMBIA ICEFIELDS, ALBERTA, CANADA [J].
HARMON, RS ;
ATKINSON, TC ;
ATKINSON, JL .
ARCTIC AND ALPINE RESEARCH, 1983, 15 (04) :503-516