Cesium selective polymeric membrane sensor based on p-isopropylcalix[6]arene and its application in environmental samples

被引:51
作者
Ahmadzadeh, Saeid [1 ]
Rezayi, Majid [2 ]
Kassim, Anuar [3 ]
Aghasi, Majid [4 ]
机构
[1] Kerman Univ Med Sci, Inst Neuropharmacol, Pharmaceut Res Ctr, Kerman, Iran
[2] Univ Malaya, Fac Sci, Dept Chem, UM Ctr Ion Liquids, Kuala Lumpur 50603, Malaysia
[3] Univ Putra Malaysia, Fac Sci, Dept Chem, Serdang 43400, Selangor, Malaysia
[4] Kerman Univ Med Sci, Herbal & Tradit Med Res Ctr, Kerman, Iran
关键词
PVC-MEMBRANE; ION; ELECTRODES; CATION; IMMOBILIZATION; COMPLEXATION; METHYLAMINE; IONOPHORES; ISE;
D O I
10.1039/c5ra02799c
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The present work deals with developing a highly selective membrane electrode based on p-isopropylcalix[ 6] arene for accurate determination of trace amounts of cesium cations. The amounts of ionophore (5 mg), PVC (30 mg), dioctyl phthalate (DOP) as a plasticizer (62 mg) and potassium tetrakis(4-chlorophenyl) borate (KTpClPB) as an ionic additive (3 mg) were optimized in the preparation of the membrane. The effects of various parameters such as pH, composition of the membrane, different cation interferences and concentration of internal solution were investigated. The electrode exhibited a linear response with a near Nernstian slope of 57.29 +/- 0.31 (mV per decade of activity) over the pH range from 2 to 11 with a satisfactory concentration range of 1.0 x 10(-6) to 1.0 x 10(-1) M. The value of 1.0 x 10(-6) M was obtained as a lower detection limit for the fabricated cesium-selective electrode. The developed sensor exhibited a good reproducibility over a period of about 12 weeks with a fast response time of less than 10 seconds. The proposed sensor was successfully applied as an indicator electrode for the potentiometric titration of a cesium solution with EDTA. The validation of the developed electrode was done using an inductively coupled plasma technique (ICP) which confirmed the accuracy and precision of the fabricated electrode in the successful determination of cesium trace amounts in the environmental samples.
引用
收藏
页码:39209 / 39217
页数:9
相关论文
共 46 条
  • [1] Ahmadzadeh S, 2011, INT J ELECTROCHEM SC, V6, P4749
  • [2] Thermodynamic Study of the Complexation of p-Isopropylcalix[6]arene with Cs+ Cation in Dimethylsulfoxide-Acetonitrile Binary Media
    Ahmadzadeh, Saeid
    Kassim, Anuar
    Rezayi, Majid
    Rounaghi, Gholam Hossein
    [J]. MOLECULES, 2011, 16 (09) : 8130 - 8142
  • [3] AMMANN D, 1983, ION SEL ELECTRODE R, V5, P3
  • [4] [Anonymous], 2009, MALAYSIAN J CHEM
  • [5] [Anonymous], 2006, Encyclopedia of sensors, potentiometric ion sensors
  • [6] A new cesium ion selective graphite rod electrode based on Cs-molybdophosphate
    Arida, HAM
    Aglan, RF
    El-Reefy, SA
    [J]. ANALYTICAL LETTERS, 2004, 37 (01) : 21 - 33
  • [7] Ion sensors: current limits and new trends
    Bakker, E
    Diamond, D
    Lewenstam, A
    Pretsch, E
    [J]. ANALYTICA CHIMICA ACTA, 1999, 393 (1-3) : 11 - 18
  • [8] Bakker E, 2004, TALANTA, V63, P3, DOI 10.1016/j.talanta.2003.10.006
  • [9] SELECTIVITY OF POLYMER MEMBRANE-BASED ION-SELECTIVE ELECTRODES - SELF-CONSISTENT MODEL DESCRIBING THE POTENTIOMETRIC RESPONSE IN MIXED ION SOLUTIONS OF DIFFERENT CHARGE
    BAKKER, E
    MERUVA, RK
    PRETSCH, E
    MEYERHOFF, ME
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (19) : 3021 - 3030
  • [10] Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics
    Bakker, E
    Buhlmann, P
    Pretsch, E
    [J]. CHEMICAL REVIEWS, 1997, 97 (08) : 3083 - 3132