A computational approach for the lifetime prediction of cardiovascular balloon-expandable stents

被引:25
作者
Auricchio, F. [1 ]
Constantinescu, A. [2 ]
Conti, M. [1 ]
Scalet, G. [1 ]
机构
[1] Univ Pavia, Dipartimento Ingn Civile & Architettura, I-27100 Pavia, Italy
[2] CNRS, Ecole Polytech, Lab Mecan Solides, UMR 7649, F-91128 Palaiseau, France
基金
欧洲研究理事会;
关键词
Cardiovascular stents; 316L stainless steel; Dang Van fatigue criterion; Fatigue-life assessment; HIGH-CYCLE FATIGUE; CRYSTAL PLASTICITY; CORONARY-ARTERY; FRACTURE; DESIGN; SIZE; DEFORMATION; IMPACT; SCALE;
D O I
10.1016/j.ijfatigue.2015.02.002
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
This paper presents a methodology for the numerical fatigue-life assessment of cardiovascular balloon-expandable stents. The methodology is based on a global computational approach composed of a mechanical finite element analysis, followed by a fatigue analysis. The method is applied to a classical 316L stainless steel coronary stent design (i.e., the Medinol/Boston Scientific NIR (TM) stent). Fatigue criteria based on elastic and plastic shakedown concepts for finite and infinite lifetime are used to predict fatigue crack initiation and are calibrated on experimental data related to 316L stainless steel mu m-size components, manufactured as stents. The results from the fatigue analysis allow to discuss several aspects affecting stent lifetime, such as the applied cyclic loading including systolic diastolic pressurization and bending. The generality of the proposed methodology encourages further investigations of such an approach for its application to other materials or small-scale components. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:69 / 79
页数:11
相关论文
共 50 条
  • [1] Fatigue life assessment of cardiovascular balloon-expandable stents: A two-scale plasticity-damage model approach
    Argente dos Santos, H. A. F.
    Auricchio, F.
    Conti, M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 15 : 78 - 92
  • [2] On the numerical investigation of cardiovascular balloon-expandable stent using finite element method
    Azaouzi, M.
    Makradi, A.
    Petit, J.
    Belouettar, S.
    Polit, O.
    COMPUTATIONAL MATERIALS SCIENCE, 2013, 79 : 326 - 335
  • [3] A computational study on vascular damage caused by shape memory alloy self-expandable and balloon-expandable stents in a stenosed artery
    Rouhani, Fatemeh
    Fereidoonnezhad, Behrooz
    Zakerzadeh, Mohammad Reza
    Baghani, Mostafa
    JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 2019, 30 (20) : 3113 - 3123
  • [4] Future Balloon-Expandable Stents: High or Low-Strength Materials?
    Khalilimeybodi, Ali
    Khoei, Amir Alishzadeh
    Sharif-Kashani, Babak
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2020, 11 (02) : 188 - 204
  • [5] Computational Analysis of Self-Expanding and Balloon-Expandable Transcatheter Heart Valves
    Pasta, Salvatore
    Gandolfo, Caterina
    BIOMECHANICS, 2021, 1 (01): : 43 - 52
  • [6] Sequential Structural and Fluid Dynamics Analysis of Balloon-Expandable Coronary Stents: A Multivariable Statistical Analysis
    Martin, David
    Boyle, Fergal
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2015, 6 (03) : 314 - 328
  • [7] Cobalt Chromium or Stainless Steel Balloon-Expandable Bare Metal Stents for Iliac Occlusive Disease?
    Torsello, Giovanni Federico
    Stavroulakis, Konstantinos
    Chlouverakis, Gregory
    Torsello, Giovanni Battista
    JOURNAL OF ENDOVASCULAR THERAPY, 2024,
  • [8] Balloon-expandable vs. self-expanding stents: new insights into a renewed debate
    de la Tone Hernandez, Jose M.
    EUROINTERVENTION, 2015, 11 (08) : 852 - 854
  • [9] Incidence, Anatomical Location, and Clinical Significance of Compressions and Fractures in Infrapopliteal Balloon-Expandable Metal Stents
    Karnabatidis, Dimitris
    Katsanos, Konstantinos
    Spiliiopoulos, Stavros
    Diamantopoulos, Athanasios
    Kagadis, George C.
    Siablis, Dimitris
    JOURNAL OF ENDOVASCULAR THERAPY, 2009, 16 (01) : 15 - 22
  • [10] Sequential Structural and Fluid Dynamics Analysis of Balloon-Expandable Coronary Stents: A Multivariable Statistical Analysis
    David Martin
    Fergal Boyle
    Cardiovascular Engineering and Technology, 2015, 6 : 314 - 328