Posterior capsular opacification comparison between morphology and objective visual function

被引:24
作者
Lu, Chengzhe [1 ]
Yu, Shasha [1 ]
Song, Hui [1 ]
Zhao, Yun [1 ]
Xie, Shiyong [1 ]
Tang, Xin [1 ,2 ,3 ,4 ]
Yuan, Xiaoyong [1 ,2 ,3 ,4 ]
机构
[1] Tianjin Eye Hosp, 4 Gansu Rd, Tianjin 300020, Peoples R China
[2] Tianjin Eye Inst, Tianjin, Peoples R China
[3] Tianjin Key Lab Ophthalmol & Visual Sci, Tianjin, Peoples R China
[4] Tianjin Med Univ, Clin Coll Ophthalmol, Tianjin 300020, Peoples R China
关键词
PCO morphology; Severity; Objective visual function; Objective scatter index (OSI); Optical quality analysis system values (OVs); PHAKIC INTRAOCULAR-LENS; OPTICAL-QUALITY; STRAYLIGHT MEASUREMENT; CONTRAST SENSITIVITY; LIGHT-SCATTER; SEVERITY;
D O I
10.1186/s12886-019-1051-z
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
BackgroundTo compare the influence of posterior capsule opacification (PCO) morphology and severity on intraocular stray light and visual function with different levels of contrast.MethodsForty-five patients diagnosed with PCO were included in this prospective consecutive case series. The Optical Quality Analysis System II (OQAS II) was adopted to assess the objective visual function including objective scatter index (OSI) and optical quality analysis system values (OVs) with 100, 20, and 9% contrast. RTVue-100 OCT was used to evaluate the PCO morphology and severity. Comparisons among visual function, morphology, and severity between pear type and fibrosis PCO were performed. The correlations among the PCO morphology, severity, OSI, and OVs were also determined.ResultsThere was a significant correlation between increased OSI and decreased visual acuity in PCO patients before laser capsulotomy. The changes of OSI were also correlated with the PCO area for the 3mm IOL optic region (r=0.43, p=0.02). The OSI was significantly higher in pear type PCO when compared with fibrosis PCO (Z=-4.06, p0.001). In addition, the increased OSI in pear type PCO was significantly correlated with the 100% OVs and the 20% OVs but not with the 9% OVs. In fibrosis PCO, OSI was only correlated with the 100% OVs and the 20% OVs pre-YAG.ConclusionsOSI and OVs could objectively indicate the visual function impairment in PCO patients. Effects of PCO on light scattering and on objective visual function might be explained by the variations of morphology and severity.
引用
收藏
页数:8
相关论文
共 30 条
  • [1] Posterior capsule morphology determinants of visual function
    Aslam, TM
    Aspinall, P
    Dhillon, B
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2003, 241 (03) : 208 - 212
  • [2] Evaluation of visual quality of spherical and aspherical intraocular lenses by Optical Quality Analysis System
    Chen, Yan
    Wang, Xue
    Zhou, Chuan-Di
    Wu, Qiang
    [J]. INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2017, 10 (06) : 914 - 918
  • [3] Visual acuity and contrast sensitivity in different types of posterior capsule opacification
    Cheng, CY
    Yen, MY
    Chen, SJ
    Kao, SC
    Hsu, WM
    Liu, JH
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (07) : 1055 - 1060
  • [4] Changes in Intraocular Straylight and Visual Acuity with Age in Cataracts of Different Morphologies
    Gholami, Sonia
    Reus, Nicolaas J.
    van den Berg, Thomas J. T. P.
    [J]. JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [5] Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method
    Guber, Ivo
    Bachmann, Lucas M.
    Guber, Josef
    Bochmann, Frank
    Lange, Alex P.
    Thiel, Michael A.
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 249 (09) : 1367 - 1371
  • [6] Comparison of methods to quantify posterior capsule opacification using forward and backward light scattering
    Hirnschall, Nino
    Crnej, Alja
    Gangwani, Vinod
    Findl, Oliver
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (05) : 728 - 735
  • [7] Assessment of subjective intraocular forward scattering and quality of vision after posterior chamber phakic intraocular lens with a central hole (Hole ICL) implantation
    Iijima, Ayaka
    Shimizu, Kimiya
    Yamagishi, Mayumi
    Kobashi, Hidenaga
    Igarashi, Akihito
    Kamiya, Kazutaka
    [J]. ACTA OPHTHALMOLOGICA, 2016, 94 (08) : E716 - E720
  • [8] Iliescu Ioana Madalina, 2017, Rom J Ophthalmol, V61, P267
  • [9] Correlation between the measurement of posterior capsule opacification severity and visual function testing
    Jose, RMJ
    Bender, LF
    Boyce, JF
    Heatley, C
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (03) : 534 - 542
  • [10] Clinical Evaluation of Optical Quality and Intraocular Scattering after Posterior Chamber Phakic Intraocular Lens Implantation
    Kamiya, Kazutaka
    Shimizu, Kimiya
    Igarashi, Akihito
    Kobashi, Hidenaga
    Ishii, Rie
    Sato, Nobuyuki
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2012, 53 (06) : 3161 - 3166