Determination of Pu isotopes in vegetation using a new on-line FI-ICP-DRC-MS protocol after microwave digestion

被引:29
作者
Epov, VN [1 ]
Benkhedda, K
Evans, RD
机构
[1] Trent Univ, Peterborough, ON K9J 7B8, Canada
[2] RAS, SB, Vinogradov Inst Geochem, Irkutsk 664033, Russia
关键词
D O I
10.1039/b504569j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new protocol for the determination of Pu isotopes in leaves has been developed. The method consists of microwave sample digestion, flow injection on-line preconcentration/ matrix separation, sample desolvation and introduction using an APEX unit and elemental determination using ICP-MS with a dynamic reaction cell. The preconcentration and matrix separation was performed using an anion-exchange resin (AG MP-1M). Detection limits as low as 2.9 fg g(-1) and 0.46 fg g(-1) were obtained for Pu-239 and other Pu isotopes (Pu-240, Pu-241, Pu-242 and Pu-244), respectively, for 10 g of NIST-SRM 1515, Apple Leaves. The yield recovery of Pu was also determined for different leaf samples collected locally and spiked with a known concentration of Pu-239 and Pu-242.
引用
收藏
页码:990 / 992
页数:3
相关论文
共 13 条
[1]   Sector field inductively coupled plasma mass spectrometry, another tool for plutonium isotopes and plutonium isotope ratios determination in environmental matrices [J].
Agarande, M ;
Benzoubir, S ;
Neiva-Marques, AM ;
Bouisset, P .
JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2004, 72 (1-2) :169-176
[2]   Rapid determination of plutonium in urine using flow injection on-line preconcentration and inductively coupled plasma mass spectrometry [J].
Epov, VN ;
Benkhedda, K ;
Cornett, RJ ;
Evans, RD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (05) :424-430
[3]   A rapid and accurate method for the determination of plutonium in food using magnetic sector ICP-MS with an ultra-sonic nebuliser and ion chromatography [J].
Evans, P ;
Elahi, S ;
Lee, K ;
Fairman, B .
JOURNAL OF ENVIRONMENTAL MONITORING, 2003, 5 (01) :175-179
[4]   Rapid dating of recent aquatic sediments using Pu activities and 240Pu/239Pu as determined by quadrupole inductively coupled plasma mass spectrometry [J].
Ketterer, ME ;
Watson, BR ;
Matisoff, G ;
Wilson, CG .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2002, 36 (06) :1307-1311
[5]   Rapid determination of Pu isotopes and atom ratios in small amounts of environmental samples by an on-line sample pre-treatment system and isotope dilution high resolution inductively coupled plasma mass spectrometry [J].
Kim, CS ;
Kim, CK ;
Lee, JI ;
Lee, KJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (03) :247-255
[6]   Development of rapid bioassay method for plutonium [J].
Kuwabara, J ;
Noguchi, H .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2002, 252 (02) :273-276
[7]  
McLendon H.R., 1976, TRANSURANIUM NUCLIDE, P347
[8]   Rapid analysis for plutonium-239 in 1 ml of urine by magnetic sector inductively coupled plasma mass spectrometry with a desolvating introduction system [J].
Pappas, RS ;
Ting, BG ;
Paschal, DC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (06) :762-766
[9]   Man-made plutonium in environment - Possible serious hazard for living species [J].
Perelygin, VP ;
Chuburkov, YT .
RADIATION MEASUREMENTS, 1997, 28 (1-6) :385-392
[10]   Preventing uranium hydride formation in standard uranium samples for determination of 239Pu by ICP-MS [J].
Vais, V ;
Li, CS ;
Cornett, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (09) :1281-1283