Improving safety in the electrophysiology laboratory using a simple radiation dose reduction strategy: a study of 1007 radiofrequency ablation procedures

被引:32
作者
Rogers, Dominic P. S. [1 ]
England, Fiona [2 ]
Lozhkin, Konstantin [3 ]
Lowe, Martin D. [1 ]
Lambiase, Pier D. [1 ]
Chow, Anthony W. C. [1 ]
机构
[1] UCLH, Heart Hosp, Dept Cardiac Electrophysiol, London W1G 8PH, England
[2] UCLH, Heart Hosp, Dept Radiol, London, England
[3] UCLH, Heart Hosp, Dept Med Phys & Bioengn, London, England
关键词
X-RAY EXAMINATIONS; CATHETER ABLATION; ATRIAL-FIBRILLATION; CARDIAC ELECTROPHYSIOLOGY; AREA PRODUCT; EXPOSURE; PATIENT; RISK; FLUOROSCOPY; TACHYCARDIA;
D O I
10.1136/hrt.2010.204222
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The use of fluoroscopic screening involves exposure to ionising radiation for both patients and operators. Objective To assess the effects of radiation dose reduction manoeuvres (DRM) during radiofrequency ablation (RFA) procedures. Design Prospective study of DRM. Setting Tertiary cardiac centre. Interventions Two DRM were combined: removal of the secondary radiation grid and programming an ultra-low pulsed fluoroscopy rate. These methods were assessed using an anthropomorphic phantom model to measure skin entrance dose rates. Procedures were classified as complex (ablation of atrial fibrillation, ventricular tachycardia or complex congenital heart disease arrhythmias) or simple (all other RFA). Main outcome measures Dose area product and screening times were compared for ablations performed before and after DRM. Equivalent doses to organs and malignancy risk were determined by computer modelling. Results Over a 39-month period, 1007 ablation procedures were performed (631 simple, 376 complex). Radiation dose was significantly reduced after DRM for both simple (20.4 +/- 26.9 Gycm(2) vs 8.0 +/- 10.3 Gycm(2), p<0.00001) and complex ablations (63.3 +/- 50.1 Gycm(2) vs 32.8 +/- 31.7 Gycm(2), p<0.00001) with no difference in screening times. The mean lifetime risk of fatal cancer attributable to radiation exposure per million procedures was reduced from 182 to 68 for simple ablations and from 440 to 155 for complex ablations. Conclusions Significant reductions in radiation exposure during RFA were achieved using simple DRM, corresponding to a two-thirds reduction of the risk of excess fatal malignancy.
引用
收藏
页码:366 / 370
页数:5
相关论文
共 30 条
  • [1] [Anonymous], 2007, RECOMMENDAT IN PRESS
  • [2] RADIATION EXPOSURE DURING RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR CONNECTIONS
    CALKINS, H
    NIKLASON, L
    SOUSA, J
    ELATASSI, R
    LANGBERG, J
    MORADY, F
    [J]. CIRCULATION, 1991, 84 (06) : 2376 - 2382
  • [3] Catheter ablation of accessory pathways, atrioventricular nodal reentrant tachycardia, and the atrioventricular junction - Final results of a prospective, multicenter clinical trial
    Calkins, H
    Yong, P
    Miller, JM
    Olshansky, B
    Carlson, M
    Saul, JP
    Huang, SKS
    Liem, LB
    Klein, LS
    Moser, SA
    Bloch, DA
    Gillette, P
    Prystowsky, E
    [J]. CIRCULATION, 1999, 99 (02) : 262 - 270
  • [4] Worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation
    Cappato, R
    Calkins, H
    Chen, SA
    Davies, W
    Iesaka, Y
    Kalman, J
    Kim, YH
    Klein, G
    Packer, D
    Skanes, A
    [J]. CIRCULATION, 2005, 111 (09) : 1100 - 1105
  • [5] Long-term safety and efficacy of circumferential ablation with pulmonary vein isolation
    Cheema, Aamir
    Dong, Jun
    Dalal, Darshan
    Vasamreddy, Chandrasekhar R.
    Marine, Joseph E.
    Henrikson, Charles A.
    Spragg, David
    Cheng, Alan
    Nazarian, Saman
    Sinha, Sunil
    Halperin, Henry
    Berger, Ronald
    Calkins, Hugh
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (10) : 1080 - 1085
  • [6] Conceptus radiation dose and risk from cardiac catheter ablation procedures
    Damilakis, J
    Theocharopoulos, N
    Perisinakis, K
    Manios, E
    Dimitriou, P
    Vardas, P
    Gourtsoyiannis, N
    [J]. CIRCULATION, 2001, 104 (08) : 893 - 897
  • [7] X-ray dose reduction in fluoroscopically guided electrophysiology procedures
    Davies, AG
    Cowen, AR
    Kengyelics, SM
    Moore, J
    Pepper, C
    Cowan, C
    Sivanathan, MU
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (03): : 262 - 271
  • [8] den BA, 1994, CIRCULATION, V89, P2710
  • [9] Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis
    Efstathopoulos, Efstathios P.
    Katritsis, Demosthenes G.
    Kottou, Sofia
    Kalivas, Nectarios
    Tzanalaridou, Efthalia
    Giazitzoglou, Eleftherios
    Korovesis, Socrates
    Faulkner, Keith
    [J]. EUROPACE, 2006, 8 (06): : 443 - 448
  • [10] Food and Drug Administration, 1995, IMP INF PHYS OTH HEA