Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography

被引:57
作者
Lavanya, Raghavan
Teo, Livia
Friedman, David S.
Aung, Han T.
Baskaran, Mani
Gao, Hong
Alfred, Tamuno
Seah, Steve K.
Kashiwagi, Kenji
Foster, Paul J.
Aung, Tin
机构
[1] Singapore Natl Eye Ctr, Glaucoma Dept, Singapore 168751, Singapore
[2] Singapore Eye Res Inst, Singapore, Singapore
[3] Wilmer Eye Inst, Baltimore, MD 21205 USA
[4] Johns Hopkins Bloomberg Sch Publ Hlth, Baltimore, MD USA
[5] Med & Vis Res Fdn, Madras, Tamil Nadu, India
[6] Clin Trials & Epidemiol Res Univ, Singapore, Singapore
[7] Univ Yamanashi, Yamanashi, Japan
[8] Inst Ophthalmol, London EC1, England
[9] Moorfields Eye Hosp, London EC1, England
[10] Natl Univ Singapore, Singapore 117548, Singapore
关键词
D O I
10.1136/bjo.2006.113761
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To compare anterior chamber depth measurements by three non-contact devices - the IOLMaster, scanning peripheral anterior chamber depth analyser (SPAC), and Visante anterior segment optical coherence tomography (AS-OCT) Methods: Prospective, cross sectional study of 497 phakic subjects over 50 years of age attending a community clinic in Singapore. Anterior chamber depth of the right eye was measured using all three techniques by the same investigator. Depth measurements were made from the corneal epithelium to the anterior lens surface. The values obtained were compared using Bland - Altman analysis. Results: 232 men and 265 women were examined (mean (SD) age, 63.4 (7.9) years). Mean anterior chamber depth was 3.08 (0.36) mm with IOLMaster, 3.10 (0.44) mm with SPAC, and 3.14 (0.34) mm with AS-OCT. A significant difference was present between the anterior chamber depth measurements recorded by the three devices (p<0.0001). Mean differences between the measurements were: AS-OCT v IOLMaster, 0.062 (0.007) mm (95% limits of agreement, 20.37 to 0.25 mm) (p<0.0001); AS-OCT v SPAC, 0.035 (0.011) mm (-0.44 to 0.51 mm) (p = 0.0001); SPAC v IOLMaster, 0.027 (0.012) mm (-0.57 to 0.50 mm) (p = 0.027). Conclusions: AS-OCT gave systematically deeper anterior chamber measurements than SPAC and IOL-Master. However, as the differences found were small they are unlikely to be clinically important.
引用
收藏
页码:1023 / 1026
页数:4
相关论文
共 22 条
[1]   Phakic anterior chamber lenses for the correction of myopia -: A 7-year cumulative analysis in complications in 283 cases [J].
Alió, JL ;
de la Hoz, F ;
Pérez-Santonja, JJ ;
Ruiz-Moreno, JM ;
Quesada, JA .
OPHTHALMOLOGY, 1999, 106 (03) :458-466
[2]  
ALSBIRK PH, 1976, ACTA OPHTHALMOL S, V54, P5
[3]   Measurement of the internal diameter and depth of the anterior chamber: IOLMaster versus anterior chamber optical coherence tomographer [J].
Baikoff, G ;
Jodai, HJ ;
Bourgeon, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1722-1728
[4]   Static and dynamic analysis of the anterior segment with optical coherence tomography [J].
Baikoff, G ;
Lutun, E ;
Ferraz, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (09) :1843-1850
[5]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]  
Devereux JG, 2000, ARCH OPHTHALMOL-CHIC, V118, P257
[8]   Partial coherence interferometry: A novel approach to biometry in cataract surgery [J].
Drexler, W ;
Findl, O ;
Menapace, R ;
Rainer, G ;
Vass, C ;
Hitzenberger, CK ;
Fercher, AF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 126 (04) :524-534
[9]   High precision biometry of pseudophakic eyes using partial coherence interferometry [J].
Findl, O ;
Drexler, W ;
Menapace, R ;
Hitzenberger, CK ;
Fercher, AF .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (08) :1087-1093
[10]   Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations [J].
Holladay, JT .