Preparation of the diphenylcarbazone-functionalized silica gel and its application to on-line selective solid-phase extraction and determination of mercury by flow-injection spectrophotometry

被引:63
作者
Fan, Jing [1 ,2 ]
Qin, Yuxia [2 ]
Ye, Cunling [1 ,2 ]
Peng, Pingan [1 ]
Wu, Chunlai [2 ]
机构
[1] Chinese Acad Sci, Guangzhou Inst Geochem, State Key Lab Organ Geochem, Guangzhou 510640, Peoples R China
[2] Henan Normal Univ, Henan Key Lab Environm Pollut Control, Sch Chem & Environm Sci, Henan 453007, Peoples R China
关键词
diphenylcarbazone-functionalized silica gel; separation; pre-concentration; mercury(II); flow-injection analysis;
D O I
10.1016/j.jhazmat.2007.04.111
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A new solid-phase extractor, diphenylcarbazone-functionalized silica gel has been synthesized and confirmed by IR and Raman spectrometry. The new solid-phase extractor is found to be stable in 1-6 mol L(-1) HCl or H(2)SO(4), and also in common organic solvents. It can be used to separate and enrich Hg(II) selectively from eight metal ions with similar characteristics such as Cd(II), Ni(II), Co(II), Mn(H), Pb(H), Zn(II), Cu(H) and Fe(III). The pre-concentration factor is as high as 500. The results show that this new solid extractor has a good stability and can be reused for many times without decreasing its extraction percentage. The micro-column packed with diphenylcarbazone-functionalized silica gel was used for on-line solid-phase extraction and determination of mercury in real samples by flow-injection spectrophotometry. At the optimal conditions, the linear range and the detection limit for the determination of Hg(II) is found to be 1-1500 and 0.90 ng mL(-1), respectively. The relative standard deviation of I I replicate measurements is less than 3%. The proposed method was demonstrated to be simple, fast, selective, low cost and less pollution. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:343 / 350
页数:8
相关论文
共 30 条