Automated development of the contrast-detail curve based on statistical low-contrast detectability in CT images

被引:16
作者
Anam, Choirul [1 ]
Naufal, Ariij [1 ]
Fujibuchi, Toshioh [2 ]
Matsubara, Kosuke [3 ]
Dougherty, Geoff [4 ]
机构
[1] Diponegoro Univ, Fac Sci & Math, Dept Phys, Jl Prof Soedarto SH, Semarang 50275, Central Java, Indonesia
[2] Kyushu Univ, Fac Med Sci, Dept Hlth Sci, Fukuoka, Japan
[3] Kanazawa Univ, Fac Hlth Sci, Inst Med Pharmaceut & Hlth Sci, Dept Quantum Med Technol, Kanazawa, Ishikawa, Japan
[4] Calif State Univ Channel Isl, Dept Appl Phys & Med Imaging, Camarillo, CA USA
关键词
image quality; low-contrast detectability; minimum detectable contrast; RADIATION-DOSE REDUCTION; ITERATIVE RECONSTRUCTION; SPATIAL-RESOLUTION; PHANTOM; DEGRADATION; ALGORITHM; NOISE; SIZE;
D O I
10.1002/acm2.13719
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We have developed a software to automatically find the contrast-detail (C-D) curve based on the statistical low-contrast detectability (LCD) in images of computed tomography (CT) phantoms at multiple cell sizes and to generate minimum detectable contrast (MDC) characteristics. Methods A simple graphical user interface was developed to set the initial parameters needed to create multiple grid region of interest of various cell sizes with a 2-pixel increment. For each cell in the grid, the average CT number was calculated to obtain the standard deviation (SD). Detectability was then calculated by multiplying the SD of the mean CT numbers by 3.29. This process was automatically repeated as many times as the cell size was set at initialization. Based on the obtained LCD, the C-D curve was obtained and the target size at an MDC of 0.6% (i.e., 6-HU difference) was determined. We subsequently investigated the consistency of the target sizes for a 0.6% MDC at four locations within the homogeneous image. We applied the software to images with six noise levels, images of two modules of the American College of Radiology CT phantom, images of four different phantoms, and images of four different CT scanners. We compared the target sizes at a 0.6% MDC based on the statistical LCD and the results from a human observer. Results The developed system was able to measure C-D curves from different phantoms and scanners. We found that the C-D curves follow a power-law fit. We found that higher noise levels resulted in a higher MDC for a target of the same size. The low-contrast module image had a slightly higher MDC than the distance module image. The minimum size of an object detected by visual observation was slightly larger than the size using statistical LCD. Conclusions The statistical LCD measurement method can generate a C-D curve automatically, quickly, and objectively.
引用
收藏
页数:14
相关论文
共 36 条
[1]  
Allert KD, 2007, J X-RAY SCI TECHNOL, V15, P9
[2]  
American College of Radiology, 2020, PHANT OV CT REV 11 1
[3]   Assessment of patient dose and noise level of clinical CT images: automated measurements [J].
Anam, Choirul ;
Budi, Wahyu Setia ;
Adi, Kusworo ;
Sutanto, Heri ;
Haryanto, Freddy ;
Ali, Mohd Hanafi ;
Fujibuchi, Toshioh ;
Dougherty, Geoff .
JOURNAL OF RADIOLOGICAL PROTECTION, 2019, 39 (03) :783-793
[4]   Quantitation of clinical feedback on image quality differences between two CT scanner models [J].
Bache, Steven T. ;
Stauduhar, Paul J. ;
Liu, Xinming ;
Loyer, Evelyne M. ;
Rong, X. John .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02) :163-169
[5]   A simple method for low-contrast detectability, image quality and dose optimisation with CT iterative reconstruction algorithms and model observers [J].
Bellesi L. ;
Wyttenbach R. ;
Gaudino D. ;
Colleoni P. ;
Pupillo F. ;
Carrara M. ;
Braghetti A. ;
Puligheddu C. ;
Presilla S. .
European Radiology Experimental, 1 (1)
[6]  
Burger W., 2009, PRINCIPLES DIGITAL I
[7]  
C Anam, 2020, J Biomed Phys Eng, V10, P623, DOI 10.31661/jbpe.v0i0.2002-1072
[8]   A Software-Based Statistical Method to Compute Low Contrast Detectability in Computed Tomography Images [J].
Chacko, M. ;
Aldoohan, S. .
MEDICAL PHYSICS, 2016, 43 (06) :3396-3396
[9]  
Chao EH, 2000, RADIOLOGY, V217, P162
[10]   A new image quality measure in CT: Feasibility of a contrast-detail measurement method [J].
Davidson, R. ;
Alsleem, H. ;
Floor, M. ;
van der Burght, R. .
RADIOGRAPHY, 2016, 22 (04) :274-281