A Comparison of Different Methods of Temperature Measurements in Sick Newborns

被引:26
|
作者
Uslu, Sinan [1 ,2 ]
Ozdemir, Hamus [2 ]
Bulbul, Ali [1 ]
Comert, Serdar [1 ]
Bolat, Fatih [1 ]
Can, Emrah [1 ]
Nuhoglu, Asiye [1 ]
机构
[1] Sisli Etfal Children Hosp, Dept Pediat, Div Neonatol, TR-34460 Istanbul, Turkey
[2] Diyarbakir Children Hosp, Div Neonatol, Dept Pediat, Diyarbakir, Turkey
关键词
Temperature; newborn; thermometers; axillary; rectal; tympanic; digital; infrared; TYMPANIC THERMOMETER; BODY-TEMPERATURE; AXILLARY; CHILDREN; ACCURACY; MERCURY;
D O I
10.1093/tropej/fmq120
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
We aimed to compare the accuracy of digital axillary thermometer (DAT), rectal glass mercury thermometer (RGMT), infrared tympanic thermometer (ITT) and infrared forehead skin thermometer (IFST) measurements with traditional axillary glass mercury thermometer (AGMT) for intermittent temperature measurement in sick newborns. A prospective, descriptive and comparative study in which five different types of thermometer readings were performed sequentially for 3 days. A total of 1989 measurements were collected from 663 newborns. DAT and ITT measurements correlated most closely to AGMT (r = 0.94). The correlation coefficent for IFST and RGMT were 0.74 and 0.87, respectively. The mean differences for DAT, ITT, RGMT and IFST were +0.02 degrees C, +0.03 degrees C, +0.25 degrees C and +0.55 degrees C, respectively. There were not any clinical differences (defined as a mean difference of 0.2 degrees C) between both mean AGMT&DAT and AGMT&ITT measurements. Our study suggests that tympanic thermometer measurement could be used as an acceptable and practical method for sick newborn in neonatal units.
引用
收藏
页码:418 / 423
页数:6
相关论文
共 50 条
  • [1] A Comparison of Different Methods of Temperature Measurement by Mothers and Physicians in Healthy Newborns
    Oncel, Mehmet Yekta
    Tekgunduz, Kadir Serafettin
    Ozdemir, Ramazan
    Calisici, Erhan
    Karahan, Sevilay
    Erdeve, Omer
    Oguz, Serife Suna
    Dilmen, Ugur
    INDIAN JOURNAL OF PEDIATRICS, 2013, 80 (03): : 190 - 194
  • [2] A Comparison of Different Methods of Temperature Measurement by Mothers and Physicians in Healthy Newborns
    Mehmet Yekta Oncel
    Kadir Serafettin Tekgunduz
    Ramazan Ozdemir
    Erhan Calisici
    Sevilay Karahan
    Omer Erdeve
    Serife Suna Oguz
    Ugur Dilmen
    The Indian Journal of Pediatrics, 2013, 80 : 190 - 194
  • [3] Comparison of rectal and axillary temperature measurements in preterm newborns
    McCarthy, Lisa K.
    O'Donnell, Colm Patrick Finbarr
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2021, 106 (05): : F509 - F513
  • [4] The Comparison of Effect of Two Different Bathing Methods on Heart Rate, Oxygen Saturation and Body Temperature of Newborns
    Ar, I.
    Gozen, D.
    EUROPEAN JOURNAL OF PEDIATRICS, 2016, 175 (11) : 1839 - 1840
  • [5] Comparison of different methods for dead space measurements in ventilated newborns using CO2-volume plot
    Wenzel, U
    Wauer, RR
    Schmalisch, G
    INTENSIVE CARE MEDICINE, 1999, 25 (07) : 705 - 713
  • [6] Comparison of different methods for dead space measurements in ventilated newborns using CO2-volume plot
    U. Wenzel
    R. R. Wauer
    G. Schmalisch
    Intensive Care Medicine, 1999, 25 : 705 - 713
  • [7] Ultrasound measurements of the newborn hip - comparison of two methods in 657 newborns
    Czubak, J
    Kotwicki, T
    ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (05): : 550 - 551
  • [8] Ultrasound measurements of the newborn hip - Comparison of two methods in 657 newborns
    Czubak, J
    Kotwicki, T
    Piontek, T
    Skrzypek, H
    ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (01): : 21 - 24
  • [9] COMPARISON OF COAGULATION AND FIBRINOLYTIC MECHANISMS IN NORMAL AND SICK NEWBORNS
    SELF, EJ
    CORRIGAN, JJ
    SOUTHERN MEDICAL JOURNAL, 1970, 63 (12) : 1485 - &
  • [10] TRANSCUTANEOUS MEASUREMENTS OF ARTERIAL PO2 IN SICK NEWBORNS
    INDYK, L
    JAMES, LS
    PEDIATRIC RESEARCH, 1975, 9 (04) : 366 - 366