Correlation Between Pixel Values in a Cone-Beam Computed Tomographic Scanner and the Computed Tomographic Values in a Multidetector Row Computed Tomographic Scanner

被引:13
作者
Chindasombatjaroen, Jira [1 ,2 ]
Kakimoto, Naoya [1 ]
Shimamoto, Hiroaki [1 ]
Murakami, Shumei [1 ]
Furukawa, Souhei [1 ]
机构
[1] Osaka Univ, Dept Oral & Maxillofacial Radiol, Grad Sch Dent, Suita, Osaka 5650871, Japan
[2] Mahidol Univ, Fac Dent, Dept Oral & Maxillofacial Radiol, Bangkok 10700, Thailand
关键词
cone-beam computed tomography; multidetector row computed tomography; pixel values; computed tomographic values; correlation; RADIATION-THERAPY; DENTAL USE; CT; DENSITY; CALIBRATION; DOSIMETRY; QUALITY; UNIT;
D O I
10.1097/RCT.0b013e31822d9725
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The study's objectives were to investigate the correlation between pixel values obtained from a cone-beam computed tomographic (CBCT) scanner and the computed tomographic (CT) values from a multidetector row CT scanner and to determine whether they have a linear relationship at various tube voltages and tube currents. Methods: A phantom with different concentrations of contrast medium at the center of a multidetector row CT and a CBCT scanner was scanned at various imaging parameters. Computed tomographic values and pixel values were measured using ImageJ. Regression analysis was performed, as well as correlation tests with a Pearson correlation coefficient. Results: A significant correlation between pixel values and the CT values of the same specimen was observed (P < 0.0001), and a linear relationship was found between the values from the 2 scanners at each parameter. Conclusions: A high correlation and linear relationship between the CT values and pixel values were found at each parameter. Therefore, linear functions can be used to convert a pixel value from the CBCT machine used in this study to the CT values.
引用
收藏
页码:662 / 665
页数:4
相关论文
共 19 条
[1]   Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[2]   Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J].
Cozzi, L ;
Fogliata, A ;
Buffa, F ;
Bieri, S .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :335-338
[3]   Effect of scattered radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N ;
Yoshida, K .
MEDICAL PHYSICS, 2001, 28 (04) :469-474
[4]  
Hashimoto Koji, 2006, J Oral Sci, V48, P27, DOI 10.2334/josnusd.48.27
[5]   Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy [J].
Hatton, Joan ;
McCurdy, Boyd ;
Greer, Peter B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (15) :N329-N346
[6]   Bone quality assessment based on cone beam computed tomography imaging [J].
Hua, Yan ;
Nackaerts, Olivia ;
Duyck, Joke ;
Maes, Frederik ;
Jacobs, Reinhilde .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (08) :767-771
[7]   Effects of image artifacts on gray-value density in limited-volume cone-beam computerized tomography [J].
Katsumata, Akitoshi ;
Hirukawa, Akiko ;
Okumura, Shinji ;
Naitoh, Munetaka ;
Fujishita, Masami ;
Ariji, Eiichiro ;
Langlais, Robert P. .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2007, 104 (06) :829-836
[8]   Comparison of radiation doses between cone beam CT and multi detector CT: TLD measurements [J].
Kim, Sangroh ;
Yoshizumi, Terry T. ;
Toncheva, Greta ;
Yoo, Sua ;
Yin, Fang-Fang .
RADIATION PROTECTION DOSIMETRY, 2008, 132 (03) :339-345
[9]   Effect of object location on the density measurement and Hounsfield conversion in a NewTom 3G cone beam computed tomography unit [J].
Lagravere, M. O. ;
Carey, J. ;
Ben Zvi, M. ;
Packota, G. V. ;
Major, P. W. .
DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (06) :305-308
[10]   Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology [J].
Ludlow, John B. ;
Ivanovic, Marija .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2008, 106 (01) :106-114