Analysis and improvement of positioning reliability and accuracy of theta pipette configuration for scanning ion conductance microscopy

被引:3
作者
Jiao, Yangbohan [1 ,2 ]
Zhuang, Jian [1 ,2 ]
Li, Fei [3 ,4 ]
Zheng, Qiangqiang [1 ,2 ]
Gao, Zijun [1 ,2 ,5 ]
机构
[1] Xi An Jiao Tong Univ, Key Lab, Educ Minist Modern Design Rotor Bearing Syst, Xian 710049, Peoples R China
[2] Xi An Jiao Tong Univ, Sch Mech Engn, Xian 710049, Peoples R China
[3] Xi An Jiao Tong Univ, Sch Life Sci & Technol, Key Lab Biomed Informat Engn, Minist Educ, Xian 710049, Peoples R China
[4] Xi An Jiao Tong Univ, Bioinspired Engn & Biomech Ctr BEBC, Xian 710049, Peoples R China
[5] Xi An Jiao Tong Univ, Honghui Hosp, Dept Anesthesiol, Xian 710068, Peoples R China
基金
中国国家自然科学基金;
关键词
Scanning ion conductance microscopy (SICM); Theta pipette; Finite element method (FEM); Imaging reliability; Imaging accuracy; TOPOGRAPHY; CELLS; VISUALIZATION; DELIVERY;
D O I
10.1016/j.ultramic.2021.113240
中图分类号
TH742 [显微镜];
学科分类号
摘要
Scanning ion conductance microscopy (SICM) as an emerging non-contact scanning probe microscopy technique and featuring its strong in-situ detectability for soft and viscous samples, is increasingly used in biomedical and materials related studies. In SICM measurements, employing theta pipette as SICM probe to scan sample is an effective method to extend the applications of SICM for multi-parameter measurement. There are two crucial but still unclear issues that influence the reliability and accuracy of the usage of theta pipette in the SICM measurements, which are the safe feedback threshold and the horizontal measurement offset. In this work, aiming at the theta pipette configuration of SICM, we systematically investigated the two issues of the theta pipette by both finite element method (FEM) simulation and SICM experiments. The FEM analysis results show that the safe feedback threshold of the one side barrel of the theta pipette is above 99.5%, and the horizontal measurement offset is similar to 0.53 times of the inner radius of the probe tip. Based on this, we proposed an improved scanning method used by the theta pipette to solve the reliability and accuracy problems caused by the feedback threshold too close to the reference current (100%) and the measurement offset error at the tip radius level. Then through testing the polydimethylsiloxane (PDMS) samples with different embossed patterns with the improved method of SICM, we can conclude that the improved method can enhance the scanning reliability by adding the double barrels approaching process and increase the positioning accuracy by compensating an offset distance. The theoretical analysis and the improved scanning method in this work demonstrate more property and usage details of the theta pipette, and further improve the reliability and accuracy of the diversified multifunctional applications of the theta pipette for SICM to meet the increasingly complex and precise research needs.
引用
收藏
页数:10
相关论文
共 27 条
[1]   Multiscale Visualization and Quantitative Analysis of L-Cystine Crystal Dissolution [J].
Adobes-Vidal, Maria ;
Shtukenberg, Alexander G. ;
Ward, Michael D. ;
Unwin, Patrick R. .
CRYSTAL GROWTH & DESIGN, 2017, 17 (04) :1766-1774
[2]   Cyclic nucleotide imaging and cardiovascular disease [J].
Berisha, Filip ;
Nikolaev, Viacheslav O. .
PHARMACOLOGY & THERAPEUTICS, 2017, 175 :107-115
[3]   Scanning Ion Conductance Microscopy: a Model for Experimentally Realistic Conditions and Image Interpretation [J].
Edwards, Martin A. ;
Williams, Cara G. ;
Whitworth, Anna L. ;
Unwin, Patrick R. .
ANALYTICAL CHEMISTRY, 2009, 81 (11) :4482-4492
[4]   Mechanical properties of bovine erythrocytes derived from ion current measurements using micropipettes [J].
Gangotra, Ankita ;
Willmott, Geoff R. .
BIOELECTROCHEMISTRY, 2019, 128 :204-210
[5]   Correlative Stimulated Emission Depletion and Scanning Ion Conductance Microscopy [J].
Hagemann, Philipp ;
Gesper, Astrid ;
Happel, Patrick .
ACS NANO, 2018, 12 (06) :5807-5815
[6]   THE SCANNING ION-CONDUCTANCE MICROSCOPE [J].
HANSMA, PK ;
DRAKE, B ;
MARTI, O ;
GOULD, SAC ;
PRATER, CB .
SCIENCE, 1989, 243 (4891) :641-643
[7]   Mechanical properties of paraformaldehyde-treated individual cells investigated by atomic force microscopy and scanning ion conductance microscopy [J].
Kim, Seong-Oh ;
Kim, Joonhui ;
Okajima, Takaharu ;
Cho, Nam-Joon .
NANO CONVERGENCE, 2017, 4
[8]   Nanoscale In Situ Characterization of Li-ion Battery Electrochemistry Via Scanning Ion Conductance Microscopy [J].
Lipson, Albert L. ;
Ginder, Ryan S. ;
Hersam, Mark C. .
ADVANCED MATERIALS, 2011, 23 (47) :5613-+
[9]   Bioinspired Protein Channel-Based Scanning Ion Conductance Microscopy (Bio-SICM) for Simultaneous Conductance and Specific Molecular Imaging [J].
Macazo, Florika C. ;
White, Ryan J. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2016, 138 (08) :2793-2801
[10]   Nanoscale Surface Charge Visualization of Human Hair [J].
Maddar, Faduma M. ;
Perry, David ;
Brooks, Rhiannon ;
Page, Ashley ;
Unwin, Patrick R. .
ANALYTICAL CHEMISTRY, 2019, 91 (07) :4632-4639