Intraoperative vs. weightbearing patellar kinematics in total knee arthroplasty: A cadaveric study

被引:16
作者
Anglin, C. [1 ,2 ]
Brimacombe, J. M. [1 ]
Wilson, D. R. [3 ,4 ,5 ]
Masri, B. A. [5 ]
Greidanus, N. V. [5 ]
Tonetti, J. [6 ]
Hodason, A. J. [1 ]
机构
[1] Univ British Columbia, Dept Mech Engn, Vancouver, BC V6T 1W5, Canada
[2] Univ Calgary, Ctr Bioengn Res & Educ, Calgary, AB, Canada
[3] Univ British Columbia, Div Orthopaed Engn Res, Vancouver, WA USA
[4] Vancouver Coastal Hlth Res Inst, Vancouver, WA USA
[5] Univ British Columbia, Dept Orthopaed, Vancouver, BC, Canada
[6] Hop Michallon, Orthopaed Trauma Dept, Grenoble, France
基金
加拿大自然科学与工程研究理事会;
关键词
total knee replacement; patella; kinematics; knee mechanics; computer-assisted surgery;
D O I
10.1016/j.clinbiomech.2007.08.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background. During knee replacement surgery, surgeons optimize intraoperative patellar tracking with the aim of optimizing postoperative tracking. This link has not been investigated to date. Our research questions were: (1) How well do patellar kinematics correlate between passive and weightbearing flexion across numerous changes in component placement? (2) How do the kinematics differ between the two loading configurations? Methods. Eight cadaveric knee joints with modified knee components that allowed 11 different femoral, tibial and patellar placements were tested in two experimental rigs simulating intraoperative and weightbearing dynamic flexion. Baseline placement had all components in neutral position. Pearson correlation coefficients were calculated for absolute baseline kinematics and for relative kinematics due to changes in component position (i.e., the 10 altered positions vs. baseline). Findings. Correlations between intraoperative and weightbearing rigs for absolute baseline kinematics were unpredictable, ranging from poor to excellent (mean 0.56 for tilt and mean 0.50 for shift). Correlations between rigs for changes in tilt and shift, i.e. relative kinematics, were strong (>0.8) or very strong (>0.9), with the exception of shift in early flexion (0.54). Differences in relative kinematics, which averaged 2.2 degrees in tilt (standard deviation 1.8 degrees) and 1.6 mm in shift (standard deviation 1.7 mm), were notably smaller and less variable than differences in absolute kinematics, which averaged 4.2 degrees in tilt (standard deviation 3.6 degrees) and 4.3 mm in shift (standard deviation 3.9 mm). Interpretation. The results of this study suggest that, while absolute kinematics may differ between conditions, if a surgeon adjusts a component position to improve patellar kinematics intraoperatively, the effects of such a geometric change will likely carry through to the postoperative joint. (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:60 / 70
页数:11
相关论文
共 29 条
[1]   Correlation of patellar tracking pattern with trochlear and retropatellar surface topographies [J].
Ahmed, AM ;
Duncan, NA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2000, 122 (06) :652-660
[2]  
ANGLIN C, 2007, T 53 ORTH RES SOC SA
[3]  
ANOUCHI YS, 1993, CLIN ORTHOP RELAT R, P170
[4]   Patellar position after total knee arthroplasty - Influence of femoral component malposition [J].
Armstrong, AD ;
Brien, HJC ;
Dunning, CE ;
King, GJW ;
Johnson, JA ;
Chess, DG .
JOURNAL OF ARTHROPLASTY, 2003, 18 (04) :458-465
[5]   A new patellofemoral scoring system for total knee arthroplasty [J].
Baldini, Andrea ;
Anderson, John A. ;
Zampetti, Piergiuseppe ;
Pavlov, Helene ;
Sculco, Thomas P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (452) :150-154
[6]   Patellofemoral evaluation after total knee arthroplasty - Validation of a new weight-bearing axial radiographic view [J].
Baldini, Andrea ;
Anderson, John A. ;
Cerulli-Mariani, Pierpaolo ;
Kalyvas, James ;
Pavlov, Helene ;
Sculco, Thomas P. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (08) :1810-1817
[7]   Patellar tilt and subluxation following subvastus and parapatellar approach in total knee arthroplasty - Implication for surgical technique [J].
Bindelglass, DF ;
Vince, KG .
JOURNAL OF ARTHROPLASTY, 1996, 11 (05) :507-511
[8]  
BRIMACOMBE JM, 2006, THESIS U BRIT COLUMB
[9]  
Chew JTH, 1997, CLIN ORTHOP RELAT R, P87
[10]   APPLICATION OF THE JOINT COORDINATE SYSTEM TO 3-DIMENSIONAL JOINT ATTITUDE AND MOVEMENT REPRESENTATION - A STANDARDIZATION PROPOSAL [J].
COLE, GK ;
NIGG, BM ;
RONSKY, JL ;
YEADON, MR .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :344-349