Electrochemical immunoassay for amyloid-beta 1-42 peptide in biological fluids interfacing with a gold nanoparticle modified carbon surface

被引:36
作者
Diba, Farhana Sharmin
Kim, Suhee
Lee, Hye Jin [1 ]
机构
[1] Kyungpook Natl Univ, Dept Chem, 80 Daehakro, Daegu City 41566, South Korea
基金
新加坡国家研究基金会;
关键词
Surface sandwich assay; Electrochemical immunosensor; Amyloid-beta; 1-42; peptide; Screen printed carbon electrode; Biological fluids; TERMINUS-SPECIFIC ANTIBODY; PLASMON RESONANCE; SERUM SAMPLES; AGGREGATION; ADSORPTION; REDUCTION; BIOSENSOR; ELECTRODE; PROTEINS; GLYCOL;
D O I
10.1016/j.cattod.2017.02.039
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
An electrochemical immunosensor involving the formation of a surface sandwich complex on a gold nanoparticle (NP) modified screen printed carbon electrode (SPCE) is demonstrated for the femtomolar detection of amyloid-beta 1-42 peptide (A beta) in both serum and plasma. Both bioreceptors forming the assay are highly selective antibodies for A beta, namely antiA beta (12F4) and (1E11) which possess different binding sites for the A beta peptide. In order to improve the sensing performance for complex biological fluidic matrix analysis, different mixed monolayers of thiol modified polyethylene glycol (PEG) and mercaptopropionic acid (MPA) were self-assembled onto the Au NP-SPCE followed by tethering antiAp (12F4) to MPA using a heterobifunctional cross linker. Surface sandwich complexes of antiAp (12F4)/A beta/antiA beta (1E11)-ALP were then formed via sequential adsorption with the latter antiAp (1E11) conjugated to alkaline phosphatase (ALP) enzyme. The reaction of surface immobilized ALP with the substrate, 4-amino phenyl phosphate (APP), generated voltammetric detection signals that linearly increased as a function of A beta concentration. Differential pulse voltammetry was applied to establish a lowest detectable concentration of 100 fM of A beta with a linear response range from 100 fM to 25 pM. Following optimization, the immunoassay platform was applied in diluted human serum and plasma samples to determine the native concentration of A beta and the results were validated using a commercially available ELISA test. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 39 条
[1]   Serum amyloid-beta levels are increased in patients with obstructive sleep apnea syndrome [J].
Bu, Xian-Le ;
Liu, Yu-Hui ;
Wang, Qing-Hua ;
Jiao, Shu-Sheng ;
Zeng, Fan ;
Yao, Xiu-Qing ;
Gao, Dong ;
Chen, Ji-Chuan ;
Wang, Yan-Jiang .
SCIENTIFIC REPORTS, 2015, 5
[2]  
Cffikae T., 2008, BIOELECTROCHEMISTRY, V4, P118
[3]   Reduction of Non-Specific Protein Adsorption Using Poly(ethylene) Glycol (PEG) Modified Polyacrylate Hydrogels In Immunoassays for Staphylococcal Enterotoxin B Detection [J].
Charles, Paul T. ;
Stubbs, Veronte R. ;
Soto, Carissa M. ;
Martin, Brett D. ;
White, Brandy J. ;
Taitt, Chris R. .
SENSORS, 2009, 9 (01) :645-655
[4]   Nanofluidic biosensing for β-amyloid detection using surface enhanced Raman spectroscopy [J].
Chou, I-Hsien ;
Benford, Melodie ;
Beier, Hope T. ;
Cote, Gerard L. ;
Wang, Miao ;
Jing, Nan ;
Kameoka, Jun ;
Good, Theresa A. .
NANO LETTERS, 2008, 8 (06) :1729-1735
[5]   Simultaneous determination of biomarkers for Alzheimer's disease using sol-gel-derived optical array biosensor [J].
Doong, Ruey-an ;
Lee, Pin-Shang ;
Anitha, Kowthalam .
BIOSENSORS & BIOELECTRONICS, 2010, 25 (11) :2464-2469
[6]  
Friedhoff LT, 2001, INT J NEUROPSYCHOPH, V4, P127, DOI 10.1017/S1461145701002310
[7]  
Gerwert K., ANAL CHEM, P2755
[8]   Self-assembled PEG monolayer based SPR immunosensor for label-free detection of insulin [J].
Gobi, K. Vengatajalabathy ;
Iwasaka, Hiroyuki ;
Miura, Norio .
BIOSENSORS & BIOELECTRONICS, 2007, 22 (07) :1382-1389
[9]  
Gonazolez M. Diaz, 2008, BIOSENS BIOELECTRON, V23, P1340
[10]  
Hasty AH, 1999, J LIPID RES, V40, P1529