How to search the experimental conditions that improve a Partial Least Squares calibration model.: Application to a flow system with electrochemical detection for the determination of sulfonamides in milk

被引:9
作者
Diez, R. [1 ]
Sarabia, L. A. [2 ]
Sanchez, M. S. [2 ]
Ortiz, M. C. [1 ]
机构
[1] Univ Burgos, Dept Chem, Fac Sci, Burgos 09001, Spain
[2] Univ Burgos, Dept Math & Computat, Fac Sci, Burgos 09001, Spain
关键词
partial least squares; PLS; decision limit; CC alpha; capability of detection; CC beta; flow analysis; sulfonamides; experimental design; pareto-optimal front;
D O I
10.1016/j.chemolab.2007.12.006
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper deals with the selection of experimental conditions and how the signals obtained in these conditions influence the fitted Partial Least Squares calibration model. The multivariate signals come from a flow analysis system with amperometric detection when determining sulfadiazine, sulfamerazine and sulfamethazine in milk. The solution (carrier plus analyte) was pumped through the system to provide a continuous supply of analyte to the cell. The detector was programmed for a scan mode operation being the multivariate signal the hydrodynamic voltammogram. To obtain an analytical signal of enough analytical quality, the Net Analyte Signal and its standard deviation have been optimised by using an experimental design. The conflicting behaviour of the two responses has been solved by estimating the Pareto-optimal front. The multivariate signals recorded in the optimal conditions found have been calibrated by Partial Least Squares regression and their figures of merit validated according to the criteria established in European Decision 2002/657/EC. In relation to the permitted limit, 100 mu g l(-1) in milk, for the total content of sulronamides established in the Commission Regulation EC no. 281/96 the proposed method has a decision limit of 109.1 mu g l(-1) and the capability of detection is 117.9 mu g l(-1) for both probability of false noncompliance and of false compliance equal to 5%. A recovery of 86.5% +/- 2.4% (n=5) has been obtained. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:71 / 82
页数:12
相关论文
共 32 条
  • [1] [Anonymous], 1997, 11843 ISO 1
  • [2] [Anonymous], 1994, Accuracy trueness and precision of measurement methods and results
  • [3] [Anonymous], ROBUST REGRESSION OU
  • [4] [Anonymous], 2000, 11843 ISO 2
  • [5] Analysis of chlorinated, sulfochlorinated and sulfonamide derivatives of n-tetradecane by gas chromatography/mass spectrometry
    Assassi, N
    Tazerouti, A
    Canselier, JP
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1071 (1-2) : 71 - 80
  • [6] Suppressing the temperature effect in near infrared spectroscopy by using orthogonal signal correction
    Blanco, Marcelo
    Valdes, Damarih
    [J]. JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2006, 14 (03) : 155 - 160
  • [7] Deb K., 2001, Multi-Objective Optimization using Evolutionary Algorithms
  • [8] Application of N-PLS to gas chromatographic and sensory data of traditional balsamic vinegars of modena
    Durante, Caterina
    Cocchi, Marina
    Grandi, Margherita
    Marchetti, Andrea
    Bro, Rasmus
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2006, 83 (01) : 54 - 65
  • [9] Handling complex effects in slurry-sampling-electrothermal atomic Zn absorption spectrometry by multivariate calibration
    Felipe-Sotelo, M.
    Cal-Prieto, M. J.
    Gomez-Carracedo, M. P.
    Andrade, J. M.
    Carlosena, A.
    Prada, D.
    [J]. ANALYTICA CHIMICA ACTA, 2006, 571 (02) : 315 - 323
  • [10] Gehring TA, 1995, J AOAC INT, V78, P1161