Multiple-scattering-free optical glucose monitoring based on femtosecond pulse interferometry

被引:6
作者
Hori, Y [1 ]
Yasui, T [1 ]
Araki, T [1 ]
机构
[1] Osaka Univ, Grad Sch Engn Sci, Dept Mech Sci & Bioengn, Osaka 5608531, Japan
关键词
glucose; fermosecond pulse light; multiple scattering; coherence gate; interferometry; time-of-flight; group refractive index;
D O I
10.1007/s10043-005-0202-z
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have designed a novel multiple-scattering-free optical glucose monitoring system based on femtosecond pulse interferometry. The glucose concentration is determined by a time-of-flight method, while multiple-scattering light is rejected by a coherence gate. The resultant accuracy and precision were 94.7 and 24.5 mg/dl, respectively, for physiological-equivalent glucose concentrations in a scattering medium of 1.0% intralipid solution, while the accuracy and precision for a reference sample containing no scattering medium were 65.3 and 23.8mg/dI, respectively. Comparing these results, we conclude that the proposed method is effective for glucose measurement in a scattering medium. (c) 2005 The Optical Society of Japan.
引用
收藏
页码:202 / 206
页数:5
相关论文
共 13 条
[1]  
Chou C, 1998, APPL OPTICS, V37, P3553, DOI 10.1364/AO.37.003553
[2]   Amplitude sensitive optical heterodyne and phase lock-in technique on small optical rotation angle detection of chiral liquid [J].
Chou, C ;
Huang, YC ;
Feng, CM ;
Chang, M .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1997, 36 (1A) :356-359
[3]  
Cote G L, 1997, J Biomed Opt, V2, P275, DOI 10.1117/12.275221
[4]   Noninvasive monitoring of glucose concentration with optical coherence tomography [J].
Esenaliev, RO ;
Larin, KV ;
Larina, IV ;
Motamedi, M .
OPTICS LETTERS, 2001, 26 (13) :992-994
[5]  
HORI Y, 2002, P SOC PHOTO-OPT INS, V4829, P539
[6]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[7]   Optical glucose sensing in biological fluids:: an overview [J].
McNichols, RJ ;
Coté, GL .
JOURNAL OF BIOMEDICAL OPTICS, 2000, 5 (01) :5-16
[8]   Precise measurement of the refractive index and optical rotatory power of a suspension by a delayed optical heterodyne technique [J].
Mitsui, T ;
Sakurai, K .
APPLIED OPTICS, 1996, 35 (13) :2253-2258
[9]  
Tarr R.V., 1991, THESIS GEORGIA I TEC
[10]   BALLISTIC 2-D IMAGING THROUGH SCATTERING WALLS USING AN ULTRAFAST OPTICAL KERR GATE [J].
WANG, L ;
HO, PP ;
LIU, C ;
ZHANG, G ;
ALFANO, RR .
SCIENCE, 1991, 253 (5021) :769-771