Painful paresthesiae are infrequent during brachial plexus localization using low-current peripheral nerve stimulation

被引:22
|
作者
Karaca, P
Hadzic, A
Yufa, M
Vloka, JD
Brown, AR
Visan, A
Sanborn, K
Santos, AC
机构
[1] Columbia Univ, Coll Phys & Surg, St Lukes Roosevelt Hosp Ctr, Dept Anesthesiol, New York, NY USA
[2] Columbia Univ, Coll Phys & Surg, Columbia Presbyterian Med Ctr, Dept Anesthesiol, New York, NY USA
关键词
regional anesthesia; nerve block; nerve stimulator;
D O I
10.1016/j.rapm.2003.08.004
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Considerable controversy exists over the relationship of paresthesia to nerve stimulation. The purpose of this study was to determine the frequency with which patients report paresthesia at the point that an acceptable motor response is obtained to low-intensity current electrical stimulation. Methods: Low-intensity current nerve stimulation (0.6 mA, 200 microseconds, 2 Hz) was used to identify the brachial plexus in 64 consecutive patients having shoulder or arm surgery with an interscalene block. During nerve localization and while maintaining a motor response (0.20 mA-0.40 mA), the patients were queried regarding any radiating sensation or pain (paresthesia) in the shoulder or extremity on the side of the blockade. Sensory distribution of the block, motor strength of the arm muscles, and adequacy of anesthesia were used to assess the extent of blockade. Results: Ninety-five percent of patients had satisfactory surgical anesthesia. None of the patients spontaneously reported having a paresthesia during nerve stimulation. However, on careful questioning, half of the patients (55%) reported electrical paresthesia, defined as dull tingling sensation traveling down to their hands and coinciding with the motor response. in addition, most patients (71%) spontaneously reported having a mild, radiating paresthesia on initial injection of local anesthetic. Conclusions: Painful paresthesiae should be infrequent when a low-stimulating current is used to identify the neural components of the brachial plexus and when the block needle is advanced slowly. Low-current intensity nerve stimulation can be used to achieve successful interscalene block with minimal discomfort to the patient.
引用
收藏
页码:380 / 383
页数:4
相关论文
共 6 条
  • [1] Anatomical and experimental studies of brachial plexus, sciatic, and femoral nerve-location using peripheral nerve stimulation in the dog
    Mahler, Stephan P.
    Adogwa, Andrew O.
    VETERINARY ANAESTHESIA AND ANALGESIA, 2008, 35 (01) : 80 - 89
  • [2] Sonographic Location of the Radial Nerve Confirmed by Nerve Stimulation During Axillary Brachial Plexus Blockade
    Wong, Daniel M.
    Gledhill, Sam
    Thomas, Rowan
    Barrington, Michael J.
    REGIONAL ANESTHESIA AND PAIN MEDICINE, 2009, 34 (05) : 503 - 507
  • [3] Current-Distance Relationships for Peripheral Nerve Stimulation Localization
    Li, Jianming
    Kong, Xuan
    Gozani, Shai N.
    Shi, Riyi
    Borgens, Richard B.
    ANESTHESIA AND ANALGESIA, 2011, 112 (01) : 236 - 241
  • [4] Real-time visualisation of peripheral nerve trauma during subepineural injection in pig brachial plexus using micro-ultrasound
    Chandra, Anu
    Soenjaya, Yohannes
    Yan, Judy
    Felts, Paul
    McLeod, Graeme
    Demore, Christine
    BRITISH JOURNAL OF ANAESTHESIA, 2021, 127 (01) : 153 - 163
  • [5] Brachial plexus block in a tamandua-mirim (Tamandua tetradactyla) using a peripheral nerve stimulator: Case report
    Silva, A. M.
    Silva, K. F.
    Santos, L. M.
    Azuaga, L. B. S.
    Jardim, P. H. A.
    Albuquerque, V. B.
    Frazilio, F. O.
    ARQUIVO BRASILEIRO DE MEDICINA VETERINARIA E ZOOTECNIA, 2018, 70 (02) : 479 - 485
  • [6] A Retrospective Study Evaluating the Effect of Low Doses of Perineural Dexamethasone on Ropivacaine Brachial Plexus Peripheral Nerve Block Analgesic Duration
    Schnepper, Gregory D.
    Kightlinger, Benjamin, I
    Jiang, Yunyun
    Wolf, Bethany J.
    Bolin, Eric D.
    Wilson, Sylvia H.
    PAIN MEDICINE, 2018, 19 (07) : 1485 - 1493