Effect of Nerve Cuff Electrode Geometry on Onset Response Firing in High-Frequency Nerve Conduction Block

被引:38
作者
Ackermann, D. Michael, Jr. [1 ,2 ,3 ]
Bhadra, Niloy [1 ]
Foldes, Emily L. [1 ]
Wang, Xiao-Feng [4 ]
Kilgore, Kevin L. [1 ,2 ,5 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
[2] MetroHlth Med Ctr, Dept Orthoped, Cleveland, OH 44109 USA
[3] Cleveland FES Ctr, Cleveland, OH 44109 USA
[4] Cleveland Clin Fdn, Dept Quantitat Hlth Sci, Cleveland, OH 44195 USA
[5] Louis Stokes Vet Affairs Med Ctr, Cleveland, OH 44106 USA
关键词
High-frequency alternating current; nerve conduction block; nerve cuff electrode; onset response; peripheral nerve; ELECTRICAL-STIMULATION; POTASSIUM ACCUMULATION; DESIGN;
D O I
10.1109/TNSRE.2010.2071882
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The delivery of high-frequency alternating currents has been shown to produce a focal and reversible conduction block in whole nerve and is a potential therapeutic option for various diseases and disorders involving pathological or undesired neurological activity. However, delivery of high-frequency alternating current to a nerve produces a finite burst of neuronal firing, called the onset response, before the nerve is blocked. Reduction or elimination of the onset response is very important to moving this type of nerve block into clinical applications since the onset response is likely to result in undesired muscle contraction and pain. This paper describes a study of the effect of nerve cuff electrode geometry (specifically, bipolar contact separation distance), and waveform amplitude on the magnitude and duration of the onset response. Electrode geometry and waveform amplitude were both found to affect these measures. The magnitude and duration of the onset response showed a monotonic relationship with bipolar separation distance and amplitude. The duration of the onset response varied by as much as 820% on average for combinations of different electrode geometries and waveform amplitudes. Bipolar electrodes with a contact separation distance of 0.5 mm resulted in the briefest onset response on average. Furthermore, the data presented in this study provide some insight into a biophysical explanation for the onset response. These data suggest that the onset response consists of two different phases: one phase which is responsive to experimental variables such as electrode geometry and waveform amplitude, and one which is not and appears to be inherent to the transition to the blocked state. This study has implications for nerve block electrode and stimulation parameter selection for clinical therapy systems and basic neurophysiology studies.
引用
收藏
页码:658 / 665
页数:8
相关论文
共 32 条
  • [1] ACKERMANN D, 2009, MUSCLE NERVE
  • [2] ACKERMANN D, 2001, MED BIOL EN IN PRESS
  • [3] ACKERMANN D, 2001, J NEUROSCI METH, V193, P72
  • [4] Electrode Design for High Frequency Block: Effect of Bipolar Separation on Block Thresholds and the Onset Response
    Ackermann, D. Michael, Jr.
    Foldes, Emily L.
    Bhadra, Niloy
    Kilgore, Kevin L.
    [J]. 2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 654 - +
  • [5] Effect of Bipolar Cuff Electrode Design on Block Thresholds in High-Frequency Electrical Neural Conduction Block
    Ackermann, D. Michael, Jr.
    Foldes, Emily L.
    Bhadra, Niloy
    Kilgore, Kevin L.
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2009, 17 (05) : 469 - 477
  • [6] Submyelin potassium accumulation may functionally block subsets of local axons during deep brain stimulation: a modeling study
    Bellinger, S. C.
    Miyazawa, G.
    Steinmetz, P. N.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2008, 5 (03) : 263 - 274
  • [7] High-frequency electrical conduction block of mammalian peripheral motor nerve
    Bhadra, N
    Kilgore, KL
    [J]. MUSCLE & NERVE, 2005, 32 (06) : 782 - 790
  • [8] High frequency electrical conduction block of the pudendal nerve
    Bhadra, Narendra
    Bhadra, Niloy
    Kilgore, Kevin
    Gustafson, Kenneth J.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2006, 3 (02) : 180 - 186
  • [9] Bhadra N, 2007, J COMPUT NEUROSCI, V22, P313, DOI [10.1007/s10827-006-0015-5, 10.1007/S10827-006-0015-5]
  • [10] BOGER A, 2008, BLADDER, V27, P1