Gas chromatography with flame ionization and flameless sulfur chemiluminescence detectors in series for dual channel detection of sulfur compounds

被引:13
作者
Chen, YC [1 ]
Lo, JG [1 ]
机构
[1] NATL TSING HUA UNIV,DEPT NUCL SCI,HSINCHU 30043,TAIWAN
关键词
gas chromatography; sulfur chemiluminescence detector; flame ionisation; dual-channel detection; optimization;
D O I
10.1007/BF02293004
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The dual-channel detection of samples with detectors coupled in series is possible only when the first detector is a non-destructive one. However, the configuration where a destructive detector, such as the flame ionization detector (FID), is follow cd by the nameless sulfur chemiluminescence detector (nameless SCD) can be used to detect simultaneously carbon and sulfur. In this work, the FID and nameless SCD are coupled in series for dual-channel detection of sulfur compounds in three gasoline samples. Optimum conditions for flameless SCD were evaluated by altering the flow rates of hydrogen and air and the oxygen pressure to the ozone generator. Samples are identified by retention indices where the homologous n-alkylthiols are used as standards. The increment in retention index per carbon number of the n-alkylthiols is 105.54 in the Quadrex 007 series, 0.32 mm i.d., 4 mu m film, column.
引用
收藏
页码:522 / 526
页数:5
相关论文
共 14 条
[1]   UNIVERSAL SULFUR DETECTION BY CHEMI-LUMINESCENCE [J].
BENNER, RL ;
STEDMAN, DH .
ANALYTICAL CHEMISTRY, 1989, 61 (11) :1268-1271
[2]   IMPROVED GAS-CHROMATOGRAPHIC OZONE CHEMILUMINESCENCE DETECTOR [J].
BRUENING, W ;
CONCHA, FJM .
JOURNAL OF CHROMATOGRAPHY, 1977, 142 (NOV) :191-201
[3]   DETERMINATION OF SULFUR COMPONENTS IN LIGHT PETROLEUM STREAMS BY HIGH-RESOLUTION GAS-CHROMATOGRAPHY WITH CHEMILUMINESCENCE DETECTION [J].
CHAWLA, B ;
DISANZO, F .
JOURNAL OF CHROMATOGRAPHY, 1992, 589 (1-2) :271-279
[4]   INSTRUMENTAL CONTRIBUTIONS TO BAND BROADENING IN GAS-CHROMATOGRAPHY .2. EXPERIMENTAL EVALUATION [J].
CRAM, SP ;
GLENN, TH .
JOURNAL OF CHROMATOGRAPHY, 1976, 119 (APR28) :55-70
[5]  
DISANZO FP, 1994, HRC-J HIGH RES CHROM, V17, P255
[6]  
GAINES KK, 1990, HRC-J HIGH RES CHROM, V13, P489
[7]   FLUORINE INDUCED CHEMI-LUMINESCENCE DETECTOR FOR REDUCED SULFUR-COMPOUNDS [J].
NELSON, JK ;
GETTY, RH ;
BIRKS, JW .
ANALYTICAL CHEMISTRY, 1983, 55 (11) :1767-1770
[8]   REDOX CHEMI-LUMINESCENCE DETECTOR - APPLICATION TO GAS-CHROMATOGRAPHY [J].
NYARADY, SA ;
BARKLEY, RM ;
SIEVERS, RE .
ANALYTICAL CHEMISTRY, 1985, 57 (11) :2074-2079
[9]   SELECTIVE CHEMILUMINESCENCE DETECTION OF SULFUR-CONTAINING-COMPOUNDS COUPLED WITH NITROGEN-PHOSPHORUS DETECTION FOR GAS-CHROMATOGRAPHY [J].
RYERSON, TB ;
BARKLEY, RM ;
SIEVERS, RE .
JOURNAL OF CHROMATOGRAPHY A, 1994, 670 (1-2) :117-126
[10]   MEANING OF GC DETECTOR CHARACTERISTICS [J].
SEVCIK, J ;
LIPS, JE .
CHROMATOGRAPHIA, 1979, 12 (10) :693-703