摘要
Background-Distal embolization of debris during percutaneous carotid artery stenting may result in neurological deficit. Filter devices for cerebral protection potentially reduce the risk of embolization. Methods and Results-Elective carotid stent implantation using 3 different types of distal filter protection devices was attempted in 88 consecutive lesions (84 patients) in the internal carotid artery that had > 70% diameter stenosis (mean, 78.7 +/- 10.7%). Procedures were performed in 3 different centers. The mean age of the patients was 69 +/- 8 years, 75% were men, and 35.7% had neurological symptoms. In 86 lesions, a stent was successfully implanted (97.7%). In 83 of these 86 procedures (96.5%), it was possible to position a filter device. In 53% of filters, there was macroscopic evidence of debris. Collected material consisted of lipid-rich macrophages, fibrin material, and cholesterol clefts, Neurological complications during the procedure, in the hospital, and at 30 days of clinical follow-up occurred in only one patient (1.2%). This patient suffered a minor stroke that resolved within I week. Two major adverse cardiac events (2.3%) occurred during the 30 days of follow-up. Conclusions-Filter protection during carotid artery stenting seems feasible and safe. In the present series, thr incidence of neurological complications was low.
引用
收藏
页码:12 / 15
页数:4
相关论文
共 16 条
  • [1] Association of intraoperative transcranial Doppler monitoring variables with stroke from carotid endarterectomy
    Ackerstaff, RGA
    Moons, KGM
    van de Vlasakker, CJW
    Moll, FL
    Vermeulen, FEE
    Algra, A
    Spencer, MP
    [J]. STROKE, 2000, 31 (08) : 1817 - 1823
  • [2] Balloon-protected carotid angioplasty
    Albuquerque, FC
    Teitelbaum, GP
    Lavine, SD
    Larsen, DW
    Giannotta, SL
    [J]. NEUROSURGERY, 2000, 46 (04) : 918 - 921
  • [3] CLINICAL RELEVANCE OF INTRAOPERATIVE EMBOLIZATION DETECTED BY TRANSCRANIAL DOPPLER ULTRASONOGRAPHY DURING CAROTID ENDARTERECTOMY - A PROSPECTIVE-STUDY OF 100 PATIENTS
    GAUNT, ME
    MARTIN, PJ
    SMITH, JL
    RIMMER, T
    CHERRYMAN, G
    RATLIFF, DA
    BELL, PRF
    NAYLOR, AR
    [J]. BRITISH JOURNAL OF SURGERY, 1994, 81 (10) : 1435 - 1439
  • [4] Henry M, 1999, J ENDOVASC SURG, V6, P321, DOI 10.1583/1074-6218(1999)006<0321:CSWCPF>2.0.CO
  • [5] 2
  • [6] Microemboli detected by transcranial Doppler monitoring in patients during carotid angioplasty versus carotid endarterectomy
    Jordan, WD
    Voellinger, DC
    Doblar, DD
    Plyushcheva, NP
    Fisher, WS
    McDowell, HA
    [J]. CARDIOVASCULAR SURGERY, 1999, 7 (01): : 33 - 38
  • [7] SITE AND PATHOGENESIS OF INFARCTS ASSOCIATED WITH CAROTID ENDARTERECTOMY
    KRUL, JMJ
    VANGIJN, J
    ACKERSTAFF, RGA
    EIKELBOOM, BC
    THEODORIDES, T
    VERMEULEN, FEE
    [J]. STROKE, 1989, 20 (03) : 324 - 328
  • [8] CAROTID ANGIOPLASTY - DETECTION OF EMBOLIC SIGNALS DURING AND AFTER THE PROCEDURE
    MARKUS, HS
    CLIFTON, A
    BUCKENHAM, T
    BROWN, MM
    [J]. STROKE, 1994, 25 (12) : 2403 - 2406
  • [9] Predictors of stroke complicating carotid artery stenting
    Mathur, A
    Roubin, GS
    Iyer, SS
    Piamsonboon, C
    Liu, MW
    Gomez, CR
    Yadav, JS
    Chastain, HD
    Fox, LM
    Dean, LS
    Vitek, JJ
    [J]. CIRCULATION, 1998, 97 (13) : 1239 - 1245
  • [10] Efficacy of a filter device in the prevention of embolic events during carotid angioplasty and stenting: An ex vivo analysis
    Ohki, T
    Roubin, GS
    Veith, FJ
    Iyer, SS
    Brady, E
    [J]. JOURNAL OF VASCULAR SURGERY, 1999, 30 (06) : 1034 - 1042