Development of a high-resolution two-dimensional detector-based dose verification system for tumor-tracking irradiation in the CyberKnife system

被引:5
作者
Kawabata, Fumitaka [1 ]
Kamomae, Takeshi [2 ]
Okudaira, Kuniyasu [1 ]
Komori, Masataka [3 ]
Oguchi, Hiroshi [3 ]
Sasaki, Motoharu [4 ]
Mori, Masaki [1 ]
Kawamura, Mariko [2 ]
Abe, Shinji [1 ]
Ishihara, Shunichi [2 ]
Naganawa, Shinji [2 ]
机构
[1] Nagoya Univ Hosp, Dept Radiol Technol, Nagoya, Aichi, Japan
[2] Nagoya Univ, Dept Radiol, Grad Sch Med, Nagoya, Aichi, Japan
[3] Nagoya Univ, Dept Integrated Hlth Sci, Grad Sch Med, Nagoya, Aichi, Japan
[4] Tokushima Univ, Inst Biomed Sci, Dept Therapeut Radiol, Grad Sch, Tokushima, Japan
关键词
basic characterization; CyberKnife; SRS MapCHECK; tumor-tracking; STEREOTACTIC BODY RADIOTHERAPY; STEP-AND-SHOOT; RADIATION-THERAPY; QUALITY-ASSURANCE; PHOTON-BEAM; LUNG; RADIOSURGERY; DOSIMETRY; MOTION; FILM;
D O I
10.1002/acm2.13645
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We aim to evaluate the basic characteristics of SRS MapCHECK (SRSMC) for CyberKnife (CK) and establish a dose verification system using SRSMC for the tumor-tracking irradiation for CK. The field size and angular dependence of SRSMC were evaluated for basic characterization. The output factors (OPFs) and absolute doses measured by SRSMC were compared with those measured using microDiamond and microchamber detectors and those calculated by the treatment planning system (TPS). The angular dependence was evaluated by comparing the SRSMC with a microchamber. The tumor-tracking dose verification system consists of SRSMC and a moving platform. The doses measured using SRSMC were compared with the doses measured using a microchamber and radiochromic film. The OPFs and absolute doses of SRSMC were within +/- 3.0% error for almost all field sizes, and the angular dependence was within +/- 2.0% for all incidence angles. The absolute dose errors between SRSMC and TPS tended to increase when the field size was smaller than 10 mm. The absolute doses of the tumor-tracking irradiation measured using SRSMC and those measured using a microchamber agreed within 1.0%, and the gamma pass rates of SRSMC in comparison with those of the radiochromic film were greater than 95%. The basic characteristics of SRSMC for CK presented acceptable results for clinical use. The results of the tumor-tracking dose verification system realized using SRSMC were equivalent to those of conventional methods, and this system is expected to contribute toward improving the efficiency of quality control in many facilities.
引用
收藏
页数:13
相关论文
共 40 条
[1]  
Accuray Incorporated, 2019, PED INF INC HUM PRES
[2]   Comprehensive evaluation of the high-resolution diode array for SRS dosimetry [J].
Ahmed, Saeed ;
Zhang, Geoffrey ;
Moros, Eduardo G. ;
Feygelman, Vladimir .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10) :13-23
[3]   Impacts of respiratory phase shifts on motion-tracking accuracy of the CyberKnife Synchrony™ Respiratory Tracking System [J].
Akino, Yuichi ;
Shiomi, Hiroya ;
Sumida, Iori ;
Isohashi, Fumiaki ;
Seo, Yuji ;
Suzuki, Osamu ;
Tamari, Keisuke ;
Otani, Keisuke ;
Higashinaka, Naokazu ;
Hayashida, Miori ;
Mabuchi, Nobuhisa ;
Ogawa, Kazuhiko .
MEDICAL PHYSICS, 2019, 46 (09) :3757-3766
[4]   Evaluation of the accuracy of the CyberKnife Synchrony Respiratory Tracking System using a plastic scintillator [J].
Akino, Yuichi ;
Sumida, Iori ;
Shiomi, Hiroya ;
Higashinaka, Naokazu ;
Murashima, Yoshiichi ;
Hayashida, Miori ;
Mabuchi, Nobuhisa ;
Ogawa, Kazuhiko .
MEDICAL PHYSICS, 2018, 45 (08) :3506-3515
[5]   STEREOTACTIC BODY RADIOTHERAPY FOR PRIMARY HEPATOCELLULAR CARCINOMA [J].
Andolino, David L. ;
Johnson, Cynthia S. ;
Maluccio, Mary ;
Kwo, Paul ;
Tector, A. Joseph ;
Zook, Jennifer ;
Johnstone, Peter A. S. ;
Cardenes, Higinia R. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (04) :E447-E453
[6]  
Andreo P., 2000, IAEA TECHNICAL REPOR
[7]   Stereotactic body radiation therapy: The report of AAPM Task Group 101 [J].
Benedict, Stanley H. ;
Yenice, Kamil M. ;
Followill, David ;
Galvin, James M. ;
Hinson, William ;
Kavanagh, Brian ;
Keall, Paul ;
Lovelock, Michael ;
Meeks, Sanford ;
Papiez, Lech ;
Purdie, Thomas ;
Sadagopan, Ramaswamy ;
Schell, Michael C. ;
Salter, Bill ;
Schlesinger, David J. ;
Shiu, Almon S. ;
Solberg, Timothy ;
Song, Danny Y. ;
Stieber, Volker ;
Timmerman, Robert ;
Tome, Wolfgang A. ;
Verellen, Dirk ;
Wang, Lu ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2010, 37 (08) :4078-4101
[8]   Sequential Phase I and II Trials of Stereotactic Body Radiotherapy for Locally Advanced Hepatocellular Carcinoma [J].
Bujold, Alexis ;
Massey, Christine A. ;
Kim, John J. ;
Brierley, James ;
Cho, Charles ;
Wong, Rebecca K. S. ;
Dinniwell, Rob E. ;
Kassam, Zahra ;
Ringash, Jolie ;
Cummings, Bernard ;
Sykes, Jenna ;
Sherman, Morris ;
Knox, Jennifer J. ;
Dawson, Laura A. .
JOURNAL OF CLINICAL ONCOLOGY, 2013, 31 (13) :1631-+
[9]   Rapid radiographic film calibration for IMRT verification using automated MLC fields [J].
Childress, NL ;
Dong, L ;
Rosen, II .
MEDICAL PHYSICS, 2002, 29 (10) :2384-2390
[10]   Comparison of Dosimetric Performance among Commercial Quality Assurance Systems for Verifying Pretreatment Plans of Stereotactic Body Radiotherapy Using Flattening-Filter-Free Beams [J].
Chung, Jin-Beom ;
Kang, Sang-Won ;
Eom, Keun-Yong ;
Song, Changhoon ;
Choi, Kyoung-Sik ;
Suh, Tae Suk .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2016, 31 (11) :1742-1748