Maximizing Data Transmission Rate for Implantable Devices Over a Single Inductive Link: Methodological Review

被引:43
作者
Trigui, Aref [1 ]
Hached, Sami [1 ]
Ammari, Ahmed Chiheb [2 ]
Savaria, Yvon [1 ]
Sawan, Mohamad [1 ]
机构
[1] Polytech Montreal, Dept Elect Engn, Montreal, PQ H3T 1J4, Canada
[2] Sultan Qaboos Univ, Coll Engn, Elect & Comp Engn Dept, Muscat 123, Oman
基金
加拿大自然科学与工程研究理事会;
关键词
Data transmission; downlink; implantable medical device (IMD); inductive link; uplink; wireless power transfer (WPT); WIRELESS POWER TRANSMISSION; BPSK DEMODULATOR; MEDICAL DEVICES; BACK TELEMETRY; COUPLED POWER; EFFICIENT; DESIGN; CMOS; REGULATIONS; TRANSCEIVER;
D O I
10.1109/RBME.2018.2873817
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Due to the constantly growing geriatric population and the projected increase of the prevalence of chronic diseases that are refractory to drugs, implantable medical devices (IMDs) such as neurostimulators, endoscopic capsules, artificial retinal prostheses, and brain-machine interfaces are being developed. According to many business forecast firms, the IMD market is expected to grow and they are subject to much research aiming to overcome the numerous challenges of their development. One of these challenges consists of designing a wireless power and data transmission system that has high power efficiency, high data rates, low power consumption, and high robustness against noise. This is in addition to minimal design and implementation complexity. This manuscript concerns a comprehensive survey of the latest techniques used to power up and communicate between an external base station and an IMD. Patient safety considerations related to biological, physical, electromagnetic, and electromagnetic interference concerns for wireless IMDs are also explored. The simultaneous powering and data communication techniques using a single inductive link for both power transfer and bidirectional data communication, including the various data modulation/demodulation techniques, are also reviewed. This review will hopefully contribute to the persistent efforts to implement compact reliable IMDs while lowering their cost and upsurging their benefits.
引用
收藏
页码:72 / 87
页数:16
相关论文
共 129 条
[1]  
Abdelhedi Riadh, 2016, 2016 7th International Conference on Sciences of Electronics, Technologies of Information and Telecommunications (SETIT), P97, DOI 10.1109/SETIT.2016.7939849
[2]  
Agarwal Kush, 2017, IEEE Rev Biomed Eng, V10, P136, DOI 10.1109/RBME.2017.2683520
[3]  
Ammari A., 2015, WILEY ENCY ELECT ELE, P1, DOI DOI 10.1002/047134608X.W6605.PUB2
[4]  
[Anonymous], 2009, INDUCTIVE POWERING B, DOI DOI 10.1007/978-90-481-2412-1_4
[5]  
[Anonymous], 2016, CONTEGO 325 MAH BATT
[6]  
Aqueveque P., 2017, Physical Disabilities-Therapeutic Implications, P95
[7]   A Thermal Energy Harvesting Power Supply With an Internal Startup Circuit for Pacemakers [J].
Ashraf, Mohammadreza ;
Masoumi, Nasser .
IEEE TRANSACTIONS ON VERY LARGE SCALE INTEGRATION (VLSI) SYSTEMS, 2016, 24 (01) :26-37
[8]  
Atluri S, 2005, I IEEE EMBS C NEUR E, P533
[9]   Application of Wireless Power Transmission Systems in Wireless Capsule Endoscopy: An Overview [J].
Basar, Md Rubel ;
Ahmad, Mohd Yazed ;
Cho, Jongman ;
Ibrahim, Fatimah .
SENSORS, 2014, 14 (06) :10929-10951
[10]   Active High Power Conversion Efficiency Rectifier With Built-In Dual-Mode Back Telemetry in Standard CMOS Technology [J].
Bawa, Gaurav ;
Ghovanloo, Maysam .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2008, 2 (03) :184-192