Review of measurement methods and clinical validation studies of noninvasive blood pressure monitors: accuracy requirements and protocol considerations for devices that require patient-specific calibration by a secondary method or device before use

被引:9
作者
Ng, Kim-Gau [1 ]
机构
[1] Upper Iowa Univ, Singapore, Singapore
关键词
clinical accuracy; noninvasive blood pressure monitor; patient-specific calibration; validation protocol; HYPERTENSION-INTERNATIONAL-PROTOCOL; VOLUME-COMPENSATION METHOD; FINGER ARTERIAL-PRESSURE; FUZZY-LOGIC CONTROLLER; PULSE TRANSIT-TIME; CORTRONIC APM-770; SURGICAL-PATIENTS; EUROPEAN-SOCIETY; RECORDING FORCE; ARM POSITION;
D O I
10.1097/MBP.0b013e32834e3c22
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Existing standardized protocols for clinical validation of noninvasive blood pressure (BP) monitors do not have specific provisions for monitors that require patient-specific calibration by a secondary method or device before use. This article seeks to identify accuracy requirements and protocol considerations for such monitors. Measurement methods that require patient-specific calibration were reviewed to identify their clinical accuracy requirements. Validation studies of monitors that use these methods were reviewed to identify limitations in their protocols. For a monitor that requires patient-specific calibration, inadequate adaptation of existing protocols can fail to validate the accuracy of the monitor for its intended use. A protocol for such a monitor must have provisions to assess the monitor's accuracy in tracking intrapatient BP changes, from the calibrated level, after a calibration or between calibrations. Performing a patient-specific calibration with a patient at rest and immediately evaluating the monitor against a reference method with the patient also at rest will not assess this accuracy, because changes in BP at rest and over a short time are generally small compared with those that occur over 24 h. For validation purposes, intrapatient BP changes can be achieved by validating the monitor over an extended period or induced by means of physical maneuvers or pharmacological interventions. The secondary method or device used for the calibration must be accurate. The protocol must also have provisions to assess the monitor's ability to correct for changes in hydrostatic pressure, reject or correct for motion artefacts, and correct for other factors that affect measurement accuracy during use. There is a need to establish protocol provisions to ensure that noninvasive BP monitors that require patient-specific calibration are properly validated for their intended use before they are placed on the market or introduced into clinical use. Blood Press Monit 16: 291-303 (C) 2011 Wolters Kluwer Health | Lippincott Williams & Wilkins. Blood Pressure Monitoring 2011, 16: 291-303
引用
收藏
页码:291 / 303
页数:13
相关论文
共 103 条
[1]   Continuous Non-Invasive Blood Pressure Monitoring, a Validation Study of Nexfin in a Pregnant Population [J].
Akkermans, J. ;
Diepeveen, M. ;
Ganzevoort, W. ;
van Montfrans, G. A. ;
Westerhof, B. E. ;
Wolf, H. .
HYPERTENSION IN PREGNANCY, 2009, 28 (02) :230-242
[2]   Modelling the relationship between peripheral blood pressure and blood volume pulses using linear and neural network system identification techniques [J].
Allen, J ;
Murray, A .
PHYSIOLOGICAL MEASUREMENT, 1999, 20 (03) :287-301
[3]  
American National Standard ANSI/AAMI, SP101987 ANSIAAMI
[4]  
American National Standards/ Association for the Advancement of Medical Instrumentation, 8106022009 ANSIAAMII
[5]  
[Anonymous], [No title captured], Patent No. [RE37,852, 37852]
[6]  
[Anonymous], 106042004 EN
[7]  
[Anonymous], SP102002 ANSIAAMI
[8]  
[Anonymous], SP101992 ANSIAAMI
[9]  
[Anonymous], 2011, [No title captured], Patent No. [2011/0054328, 20110054328]
[10]  
Australian Standard, 36551989 AS