A prototype quantitative film scanner for radiochromic film dosimetry

被引:14
作者
Ranade, Manisha K.
Li, Jonathan G.
Dubose, Ryan S. [1 ]
Kozelka, Jakub [2 ]
Simon, William E. [2 ]
Dempsey, James F.
机构
[1] Spartanburg Reg Hosp Gibbs Canc Ctr, Spartanburg, SC 29303 USA
[2] Sun Nucl Corp, Melbourne, FL 32940 USA
关键词
densitometer; film dosimetry; radiochromic film;
D O I
10.1118/1.2828203
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have developed a high resolution, quantitative, two-dimensional optical film scanner for use with a commercial high sensitivity radiochromic film (RCF) for measuring single fraction externalbeam radiotherapy dose distributions. The film scanner was designed to eliminate artifacts commonly observed in RC,F dosimetry. The scanner employed a stationary light source and detector with a moving antireflective glass film platen attached to a high precision computerized X-Y translation stage. An ultrabright red light emitting diode (LED) with a peak output at 633 nm and full width at half maximum (FWHM) of 16 nm was selected as the scanner light source to match the RCF absorption peak. A dual detector system was created using two silicon photodiode detectors to simultaneously measure incident and transmitted light. The LED light output was focused to a submillimeter (FWHM 0.67 mm) spot size, which was determined from a scanning knife-edge technique for measuring Gaussian optical beams. Data acquisition was performed with a 16-bit A/D card in conjunction with commercial software. The linearity of the measured densities on the scanner was tested using a calibrated neutral-density step filter. Sensitometric curves and three IMRT field scans were acquired with a spatial resolution of I mm for both radiographic film and RCR The results were compared with measurements taken with a commercial diode array under identical delivery conditions. The RCF was rotated by. 90 deg and. rescanned to study orientation effects. Comparison between the RCF and the diode array measurements using percent dose difference and distance-to-agreement criteria produced average passing rates of 99.0% using 3%/3 rum criteria and 96.7% using 2%/2 rum criteria. The same comparison between the radiographic film and diode array measurements resulted in average passing rates 96.6% and 91.6% for the above two criteria, respectively. No measurable light-scatter or interference scanner artifacts were observed. The RCF rotated by 90 deg showed no measurable orientation effect. A scan of a 15 X 15 cm(2) area with I turn resolution required 22 min to acquire. The LED densitometer provides an accurate film dosimetry system with 4 rum or better resolution. The scanner eliminates the orientation dependence of RCF dosimetry that was previously reported with commercial flatbed scanners. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:473 / 479
页数:7
相关论文
共 50 条
  • [31] Extension of radiochromic film dosimetry to three dimensions: A feasibility study
    Adamczyk, S.
    Skorska, M.
    NUCLEAR PHYSICS METHODS AND ACCELERATORS IN BIOLOGY AND MEDICINE, 2007, 958 : 286 - 287
  • [32] Radiochromic film dosimetry for a 62 MeV clinical proton beam
    Fidanzio, A
    Azario, L
    De Angelis, C
    Pacilio, M
    Onori, S
    Piermattei, A
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2002, 18 (01): : 15 - 23
  • [33] Raman scattering for dosimetry using GAFCHROMIC EBT3 radiochromic dosimetry film
    Talarico, Olga S.
    Krylova, Tatyana A.
    Melnik, Nikolay N.
    MEDICAL PHYSICS, 2019, 46 (04) : 1883 - 1887
  • [34] Evaluation of Scanner Image Properties of Radiochromic Film for Measuring Effective Energy
    Gotanda, Tatsuhiro
    Katsuda, Toshizo
    Gotanda, Rumi
    Akagawa, Takuya
    Kuwano, Tadao
    Tanki, Nobuyoshi
    Yabunaka, Koichi
    Tabuchi, Akihiko
    Shimono, Tetsunori
    Kawaji, Yasuyuki
    Takeda, Yoshihiro
    6TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 45 : 220 - 223
  • [35] Scanner uniformity improvements for radiochromic film analysis with matt reflectance backing
    Ethan Butson
    Hani Alnawaf
    Peter K. N. Yu
    Martin Butson
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 401 - 407
  • [36] Scanner uniformity improvements for radiochromic film analysis with matt reflectance backing
    Butson, Ethan
    Alnawaf, Hani
    Yu, Peter K. N.
    Butson, Martin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (03) : 401 - 407
  • [37] Effective energy measurement using radiochromic film: application of a mobile scanner
    Gotanda, Tatsuhiro
    Katsuda, Toshizo
    Gotanda, Rumi
    Kuwano, Tadao
    Akagawa, Takuya
    Tanki, Nobuyoshi
    Tabuchi, Akihiko
    Shimono, Tetsunori
    Kawaji, Yasuyuki
    Takeda, Yoshihiro
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2016, 22 (04): : 85 - 92
  • [38] An inexpensive method of small photon field dosimetry with EBT3 radiochromic film
    Mourya A.
    Aggarwal L.M.
    Mandal A.
    Shahi U.P.
    Ram C.
    Asthana A.K.
    Pradhan S.
    Health and Technology, 2018, 8 (3) : 171 - 177
  • [39] Predicting optical densitometer response as a function of light source characteristics for radiochromic film dosimetry
    Reinstein, LE
    Gluckman, GR
    Amols, HI
    MEDICAL PHYSICS, 1997, 24 (12) : 1935 - 1942
  • [40] Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy
    Palmer, Antony L.
    Bradley, David
    Nisbet, Andrew
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04): : 280 - 296