Conditions for Worm-Robot Locomotion in a Flexible Environment: Theory and Experiments

被引:39
作者
Zarrouk, David [1 ]
Sharf, Inna [2 ]
Shoham, Moshe [1 ]
机构
[1] Technion Israel Inst Technol, Fac Mech Engn, IL-32000 Haifa, Israel
[2] McGill Univ, Dept Mech Engn, Montreal, PQ H3A 3R1, Canada
关键词
Biological tissue; compliance; efficiency; experimental validation; flexible surface; intestine; locomotion; medical robots; prototype; worm robots; DEVICES;
D O I
10.1109/TBME.2011.2182612
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Biological vessels are characterized by their substantial compliance and low friction that present a major challenge for crawling robots for minimally invasive medical procedures. Quite a number of studies considered the design and construction of crawling robots; however, very few focused on the interaction between the robots and the flexible environment. In a previous study, we derived the analytical efficiency of worm locomotion as a function of the number of cells, friction coefficients, normal forces, and local (contact) tangential compliance. In this paper, we introduce the structural effects of environment compliance, generalize our previous analysis to include dynamic and static coefficients of friction, determine the conditions of locomotion as function of the external resisting forces, and experimentally validate our previous and newly obtained theoretical results. Our experimental setup consists of worm robot prototypes, flexible interfaces with known compliance and a Vicon motion capture system to measure the robot positioning. Separate experiments were conducted to measure the tangential compliance of the contact interface that is required for computing the analytical efficiency. The validation experiments were performed for both types of compliant conditions, local and structural, and the results are shown to be in clear match with the theoretical predictions. Specifically, the convergence of the tangential deflections to an arithmetic series and the partial and overall loss of locomotion verify the theoretical predictions.
引用
收藏
页码:1057 / 1067
页数:11
相关论文
共 18 条
[1]   From wired to wireless: A miniature robot for intestinal inspection [J].
Chi, Dongxiang ;
Yan, Guozheng .
Journal of Medical Engineering and Technology, 2003, 27 (02) :71-76
[2]   Modeling and experimental validation of the locomotion of endoscopic robots in the colon [J].
Dario, P ;
Ciarletta, P ;
Menciassi, A ;
Kim, B .
INTERNATIONAL JOURNAL OF ROBOTICS RESEARCH, 2004, 23 (4-5) :549-556
[3]   A Self-Propelled Inflatable Earthworm-Like Endoscope Actuated by Single Supply Line [J].
Glozman, Daniel ;
Hassidov, Noam ;
Senesh, Merav ;
Shoham, Moshe .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (06) :1264-1272
[4]   Locomotion techniques for robotic colonoscopy [J].
Kassim, Irwan ;
Phee, Louis ;
Ng, Wan S. ;
Gong, Feng ;
Dario, Paolo ;
Mosse, Charles A. .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2006, 25 (03) :49-56
[5]   Design and fabrication of a locomotive mechanism for capsule-type endoscopes using shape memory alloys (SMAs) [J].
Kim, B ;
Lee, S ;
Park, JH ;
Park, JO .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2005, 10 (01) :77-86
[6]   Evaluation of the critical stroke of an earthworm-like robot for capsule endoscopes [J].
Kwon, Jiwoon ;
Park, Sukho ;
Park, Jongoh ;
Kim, Byungkyu .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H4) :397-405
[7]  
Li W. D., 2006, 2006 9th International Conference on Control, Automation, Robotics and Vision (IEEE Cat. No.06EX1361)
[8]  
Menciassi A, 2006, AUTON ROBOT, V21, P155, DOI 10.1007/s10514-006-9846-9
[9]  
Nakamura T, 2008, IEEE INT CONF ROBOT, P238
[10]  
Phee L, 2003, SPRINGER TRAC ADV RO, V6, P467