Pulse Arrival Time is Not an Adequate Surrogate for Pulse Transit Time in Terms of Tracking Diastolic Pressure

被引:0
|
作者
Zhang, Guanqun [1 ]
Xu, Da [1 ]
Olivier, N. Bari [2 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
[2] Michigan State Univ, Dept Small Anim Clin Sci, E Lansing, MI 48824 USA
来源
2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2011年
关键词
PRE-EJECTION PERIOD; BLOOD-PRESSURE; WAVE VELOCITY; MARKER;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We compared pulse arrival time (PAT) and pulse transit time (PTT) in terms of their ability to track diastolic pressure (DP). We performed the comparison using high fidelity, invasive arterial waveforms recorded from six dogs during multiple interventions. On average, DP ranged from 40 to 106 mmHg and therefore varied widely. PAT and PTT were able to predict DP with average root-mean-squared-errors of 9.8 +/- 5.8 mmHg and 5.7 +/- 2.0 mmHg (p = 0.02). Thus, even though PAT is simpler to measure, we can only recommend using PTT for tracking DP.
引用
收藏
页码:6462 / 6464
页数:3
相关论文
共 50 条
  • [1] Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
    Zhang, Guanqun
    Gao, Mingwu
    Xu, Da
    Olivier, N. Bari
    Mukkamala, Ramakrishna
    JOURNAL OF APPLIED PHYSIOLOGY, 2011, 111 (06) : 1681 - 1686
  • [2] Pulse arrival time as a surrogate of blood pressure
    Finnegan, Eoin
    Davidson, Shaun
    Harford, Mirae
    Jorge, Joao
    Watkinson, Peter
    Young, Duncan
    Tarassenko, Lionel
    Villarroel, Mauricio
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [3] Assessment of Pulse Transit/Arrival Time as Noninvasive Blood Pressure Predictors in Finger and Earlobe Sites
    Selvaraj, Nandakumar
    2016 IEEE HEALTHCARE INNOVATION POINT-OF-CARE TECHNOLOGIES CONFERENCE (HI-POCT), 2016, : 200 - 203
  • [4] Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation
    Joonnyong Lee
    JangJay Sohn
    Jonghyun Park
    SeungMan Yang
    Saram Lee
    Hee Chan Kim
    BioMedical Engineering OnLine, 17
  • [5] Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation
    Lee, Joonnyong
    Sohn, JangJay
    Park, Jonghyun
    Yang, SeungMan
    Lee, Saram
    Kim, Hee Chan
    BIOMEDICAL ENGINEERING ONLINE, 2018, 17
  • [6] Pulse transit time as a surrogate measure of changes in systolic arterial pressure in children during sleep
    Vlahandonis, Anna
    Biggs, Sarah N.
    Nixon, Gillian M.
    Davey, Margot J.
    Walter, Lisa M.
    Horne, Rosemary S. C.
    JOURNAL OF SLEEP RESEARCH, 2014, 23 (04) : 406 - 413
  • [7] An implantable optical blood pressure sensor based on pulse transit time
    Fiala, Jens
    Bingger, Philipp
    Ruh, Dominic
    Foerster, Katharina
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas
    BIOMEDICAL MICRODEVICES, 2013, 15 (01) : 73 - 81
  • [8] Pulse Transit Time and Blood Pressure During Cardiopulmonary Exercise Tests
    Wibmer, T.
    Doering, K.
    Kropf-Sanchen, C.
    Ruediger, S.
    Blanta, I.
    Stoiber, K. M.
    Rottbauer, W.
    Schumann, C.
    PHYSIOLOGICAL RESEARCH, 2014, 63 (03) : 287 - 296
  • [9] A Study on the Correlation between Arterial Pressure, Pulse Transit Time and Other Time-Domain Characteristic Parameters of Radial Pulse Wave
    Su, Yilun
    He, Rui
    Ji, Lianying
    2ND INTERNATIONAL CONFERENCE ON SENSORS, INSTRUMENT AND INFORMATION TECHNOLOGY (ICSIIT 2015), 2015, : 357 - 363
  • [10] Correlation Study Between Blood Pressure And Pulse Transit Time
    Pereira, T.
    Sanches, R.
    Reis, P.
    Pego, J.
    Simoes, R.
    2015 IEEE 4TH PORTUGUESE MEETING ON BIOENGINEERING (ENBENG), 2015,