In vitro evaluation of 56 coronary artery stents by 256-slice multi-detector coronary CT

被引:9
|
作者
Steen, Henning [1 ]
Andre, Florian [1 ]
Korosoglou, Grigorios [1 ]
Mueller, Dirk [2 ]
Hosch, Waldemar [3 ]
Kauczor, Hans-Ulrich [3 ]
Giannitsis, Evangelos [1 ]
Katus, Hugo A. [1 ]
机构
[1] Univ Heidelberg, Dept Cardiol, D-69120 Heidelberg, Germany
[2] Philips GmbH, Healthcare Div, D-20099 Hamburg, Germany
[3] Univ Heidelberg, Dept Diagnost & Intervent Radiol, D-69120 Heidelberg, Germany
关键词
Multislice CT; Coronary stents; CT angiography; Artifacts; Coronary artery disease; In-stent; Restenosis; SPIRAL COMPUTED-TOMOGRAPHY; DUAL-SOURCE CT; ROW CT; ANGIOGRAPHY; RESTENOSIS; VISUALIZATION;
D O I
10.1016/j.ejrad.2010.08.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: We sought to investigate stent lumen visibility of 56 coronary stents with the newest 256-multi-slice-CT (256-MDCT) technology for different reconstruction algorithms in an in vitro model. Background: Early identification of in-stent restenosis (ISR) is important to avoid recurrent ischemia and prevent acute myocardial infarction (AMI). Since angiography has the disadvantage of high costs and its invasiveness, MDCT could be a convenient and safe non-invasive alternative for detection of ISR. Material and methods: Percentages of in-stent lumen diameter and in-stent signal attenuation (measured as contrast-to-noise ratio (CNR)) of 56 coronary stents (group A <= 2.5 mm; group B = 2.75-3.0 mm; group C = 3.5-4.0 mm) were evaluated in a coronary vessel in vitro phantom (iodine-filled plastic tubes) employing four different reconstruction algorithms (XCD, CC, CD, XCB) on a novel 256-MDCT (Philips-iCT, collimation = 128 mm x 0.625 mm; rotation time = 270 ms; tube current = 800 mA s with 120 kV). Analysis was conducted with the semi-automatical full-width-at-half-maximum (FWHM) method. P-values <0.05 were regarded statistically significant. Results: In-stent lumen diameter >60% for group C stents was significantly larger and CNR was significantly lower (both p < 0.05) for sharp kernels (CD; XCD) when compared to groups A/B. The FWHM-method showed significantly smaller in-stent lumen diameter (p < 0.05) when compared to the manual method. Conclusion: 256-MDCT could potentially be employed for clinical assessment of stent patency in stents >3.0 mm when analysed with cardio-dedicated sharp kernels, although clinical studies corroborating this claim should be performed. However, stents <= 3.0 mm reconstructed by soft kernels revealed insufficient in-stent lumen visualisation and should not be used in clinical practice. Further improvements in spatial and temporal image resolution as well as reductions of radiation exposure and image noise have to be accomplished for the ambitious goal of characterising both CT coronary artery anatomy and in-stent lumen. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:143 / 150
页数:8
相关论文
共 50 条
  • [31] Evaluation of peripheral artery stent with 64-slice multi-detector row CT angiography: Prospective comparison with digital subtraction angiography
    Li, Xiao-ming
    Li, Yu-hua
    Tian, Jian-ming
    Xiao, Yi
    Lu, Jian-ping
    Jing, Zai-ping
    Sheng, Jing
    Edwin, Angela
    Wu, Fang-hong
    EUROPEAN JOURNAL OF RADIOLOGY, 2010, 75 (01) : 98 - 103
  • [32] Step and shoot coronary CT angiography using 256-slice CT: effect of heart rate and heart rate variability on image quality
    D. Muenzel
    P. B. Noel
    F. Dorn
    M. Dobritz
    E. J. Rummeny
    A. Huber
    European Radiology, 2011, 21 : 2277 - 2284
  • [33] Imaging of coronary artery stents using multislice computed tomography: in vitro evaluation
    Maintz, D
    Juergens, KU
    Wichter, T
    Grude, M
    Heindel, W
    Fischbach, R
    EUROPEAN RADIOLOGY, 2003, 13 (04) : 830 - 835
  • [34] Assessment of coronary artery calcification in hemodialysis patients using multi-detector spiral CT scan
    Nitta, K
    Akiba, T
    Suzuki, K
    Uchida, K
    Ogawa, T
    Majima, K
    Watanabe, RI
    Aoki, T
    Nihei, H
    HYPERTENSION RESEARCH, 2004, 27 (08) : 527 - 533
  • [35] Evaluation of epicardial adipose tissue by coronary multi-detector computed tomography: an independent predictor of obstructive coronary artery disease
    Maryam Moradi
    Vahid Talebi
    Egyptian Journal of Radiology and Nuclear Medicine, 54
  • [36] Radiation dose to patients and image quality evaluation from coronary 256-slice computed tomographic angiography
    Chen, Liang-Kuang
    Wu, Tung-Hsin
    Yang, Ching-Ching
    Tsai, Chia-Jung
    Lee, Jason J. S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) : 368 - 371
  • [37] Evaluation of epicardial adipose tissue by coronary multi-detector computed tomography: an independent predictor of obstructive coronary artery disease
    Moradi, Maryam
    Talebi, Vahid
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2023, 54 (01)
  • [38] Assessment of the Relationship Between the Coronary Venous and Arterial Systems Using 256-Slice Computed Tomography
    Bai, Wei
    Xu, Xiao
    Ma, Heng
    Zhong, Lin
    Xie, Haizhu
    Sun, Chunjuan
    Shi, Yinghong
    Che, Kaili
    Liu, Meijie
    Yang, Jun
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2020, 44 (01) : 1 - 6
  • [39] Coronary artery vasculitis: assessment with cardiac multi-detector computed tomography
    Jeon, Chang Ho
    Kim, Yeo Koon
    Chun, Eun Ju
    Kim, Jeong A.
    Yong, Hwan Seok
    Doo, Kyung Won
    Choi, Sang Il
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2015, 31 : 59 - 67
  • [40] Combining Coronary with Carotid and Cerebrovascular Angiography Using Prospective ECG Gating and Iterative Reconstruction with 256-slice CT
    Zhang, Jin-Ling
    Liu, Bai-Lu
    Zhao, Yan-Ming
    Liang, Hong-Wei
    Wang, Guo-Kun
    Wan, Yong
    Huang, Ya-Hua
    Shen, Bao-Zhong
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2015, 32 (08): : 1291 - 1298